4. Installation and operation
4.1 Mounting bore preparation
4.1.1 Dimensions
Fig. 5: Mounting situation for
Type 6217AA02000
(10-32 connector)
Page 8
The bore depth D must be sufficient to allow thread engagement over the
full length L (see table below). If the testing vessel has sufficient material
strength, a thread engagement of 0.8 L is acceptable.
L
Direct mounting
Mounting with thermal protection shield
Type 6563A or diaphragm protector
Type 6564
The wall thickness must exceed the bore depth by a web thickness of min.
4 mm. For thicker testing vessels, the top part of the bore must be drilled to
at least ø12.5 mm (BNC ø15 mm, TNC side access necessary), so that the
sensor can be mounted using a tubular socket wrench Type 1373A1.
Quartz high-pressure sensor Type 6217A...
D (1 L)
18 mm
20.5 mm
Fig. 6: Mounting situation for
Type 6217AB02000
(BNC connector)
D (0.8 L)
14.5 mm
16.5 mm
002-689e-05.21