OWNER'S MANUAL
APM Bariatric
PressureGuard
®
Span-America Medical Systems, Inc.
P17159 R0 ECO#2277

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PressureGuard APM Bariatric and is the answer not in the manual?

Questions and answers

Summary of Contents for SPAN PressureGuard APM Bariatric

  • Page 1 OWNER’S MANUAL APM Bariatric PressureGuard ® Span-America Medical Systems, Inc. P17159 R0 ECO#2277...
  • Page 2: Table Of Contents

    Intended Use Indications for Use Modes of Operation FEATURES ............................4 DIRECTIONS FOR SET-UP ......................6 DIRECTIONS FOR USE ........................8 7510 CONTROL UNIT with PRESSUREGUARD APM BARIATRIC ........9 GENERAL DIRECTIONS FOR USE ..................... 11 Linen s Bed Rails Storage...
  • Page 3: Document Symbols

    This manual contains different typefaces and symbols to make the content easier to read and understand: DOCUMENT SYMBOLS • Standard text – used for regular information. • Boldface text – stresses a word or phrase. • NOTE: - sets apart special information or important instruction clarification. Symbol Protection against the ingress of fingers or Attention, Please Read Instructions...
  • Page 4: Introduction

    INTRODUCTION PRESSUREGUARD ® Alternating Pressure Mattresses DESCRIPTION: The system consists of a foam shell with a high-density foam topper serving as the support surface underneath the patient. The foam shell also includes contoured foam bolsters at the sides and ends of the mattress, providing added patient stability and positioning.
  • Page 5 7. N'utilisez jamais ce produit s'il a un cordon ou une fiche endommagé, s'il ne fonctionne pas correctement, s'il est tombé ou endommagé, ou s'il est tombé dans l'eau. Contactez Span-America Medical Systems, Inc., pour le retour de l'unité de contrôle pour examen et réparation.
  • Page 6: Features

    FEATURES ① Cover ② Foam Topper with Geo-Matt design ④ Air Cylinders ③ Safety Edge bolster ⑤ Model 7510 Bariatric P17159 R0 ECO#2277...
  • Page 7: Common Features

    Bariatric Model The cycles and inflation levels are designed to provide and maintain low interface pressures throughout the mattress, and to redistribute peak interface pressure points during the alternating cycle. ⑤ Model 7510 – PressureGuard APM Bariatric Solutions Control Units: P17159 R0 ECO#2277...
  • Page 8: Directions For Set-Up

    DIRECTIONS FOR SET-UP 1. Place the PressureGuard mattress on the bed frame with the air line connectors at the foot end of the bed. Hospital-grade models have a gray vinyl side that should be down on the bed frame, a green fabric side that should be face up toward the patient. 2.
  • Page 9 Always plug the power cable securely into the wall outlet. Make sure the wall-mounted outlet will accommodate a heavy duty or hospital-grade plug and that the outlet is in good working order. The plug of the power cord should fit tightly into the wall outlet. The plug body, the wall outlet, and the wall plate should not be cracked or chipped.
  • Page 10: Directions For Use

    If no obvious connection problem can be corrected, contact local dealer, or call Span-America customer service at (800) 888-6752). AVERTISSEMENT : si le voyant orange de basse pression indiquant une condition de basse pression ne s'éteint pas, vérifiez que toutes les connexions de conduite...
  • Page 11: Mode Selection

    7510 CONTROL UNIT FUNCTIONS Allows selection of air cylinder firmness within a Comfort Level Selection: relatively small range. Press “Softer” or “Firmer” button to achieve desired setting. Begin in softest setting, then adjust for comfort as desired. Mode Selection: Press button to select “Alternate” or “Float” indicated by green indicator light. “Alternate”...
  • Page 12 “Cycle Time” Selection mode: Allows the user to determine the cycle time of the A-B-C inflation and deflation cycle. The control will cycle from A down B up C up  A up B down C upA up B up C down within the time illuminated on the display.
  • Page 13: General Directions For Use

    GENERAL DIRECTIONS FOR USE OF ALL MATTRESS MODELS TROUBLESHOOTING PATIENT COMPLAINTS: Occasionally a patient will complain of feeling as if they are “sinking into a hole”. 1. Sometimes this happens when the head of the bed is elevated and the mattress is in either lateral rotation or alternating pressure.
  • Page 14 For transportation, secure to prevent damage or falls. For shipment, use box and packaging as provided by the manufacturer. BED RAILS: Due to concerns over the possibility of patient entrapment, Span-America recognizes that the use of rails of any length is a matter currently addressed by various regulations from local, national and international government agencies, and by individual facility protocol.
  • Page 15: Use In Wound Care

    USE IN WOUND CARE: Use of PressureGuard APM models is only one element of care in ® the prevention and treatment of pressure injuries. Frequent repositioning, proper care, routine skin assessment, wound treatment and proper nutrition are but a few of the elements required in the prevention and treatment of pressure injuries.
  • Page 16: Cleaning And Disinfection

    CLEANING and DISINFECTION All Models It is important to follow cleaning procedures to avoid cross contamination. Be sure to clean the surface of the control unit in a dry and dust free environment. Wipe down the control unit with a damp cloth pre-soaked with a mild detergent. Avoid contact with dust and proximity to dusty areas.
  • Page 17 disinfectants are not effective against C. difficile and should not be used to disinfect environmental services. For further information relative to this organism and infection control in the healthcare setting, please refer to www.cdc.gov/ncidod/hip. Disinfect mattress covers following contamination with bodily fluids and between patients.
  • Page 18 Span-America Medical Systems, Inc. for replacement. Signalez tout son inhabituel, odeur de brûlé ou tout autre élément inhabituel au personnel de maintenance. Cessez immédiatement d'utiliser le cordon d'alimentation et contactez Span-America Medical Systems, Inc. pour le remplacer.
  • Page 19: Specifications

    SPECIFICATIONS Bacteriostatic, flame resistant, fluidproof, tear resistant COVER: High-density open-cell polyurethane. Conforms to NFPA 101 small scale and Cal TB# 117. FOAM: Urethane and/or Fabric Coated PVC AIR CYLINDERS: 110-120VAC, 60 Hz, 0.17A With respect to electric shock, fire and mechanical hazards ELECTRICAL: only in accordance with UL 60601-1.
  • Page 20: Ordering Information

    ORDERING INFORMATION Product Name Catalog Number Mattress Size PressureGuard APM Bariatric AP8042-29 80”L X 42”W X 7”H with 7510 Control Unit AP8048-29 80”L X 48”W X 7”H AP8054-29 80”L X 54”W X 7”H AP8442-29 84”L X 42”W X 7”H AP8448-29 84”L X 48”W X 7”H...
  • Page 21: Troubleshooting Guide

    Patient not centered on Reposition the patient. mattress. Defective Control Unit Call for Service. Call Span-America for assistance Patient exceeds weight limit. with product selection. Control Unit is not turned on. Turn control unit on. Mattress not inflating or...
  • Page 22: Technical Service

    Problem Possible Cause Solution Mattress is too firm Comfort level too high. Adjust comfort level down to desired level. Control unit vent valve failure. Disconnect control unit, connect airlines to one another to equalize the mattress. Exhaust air from fittings to achieve desired comfort level and reconnect (repeat as needed).
  • Page 23: Appendix A: Emc Information

    Appendix A: EMC Information Guidance and Manufacturer’s Declaration- Electromagnetic Emissions: This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. Emissions Test Compliance Electromagnetic Environment-Guidance Group1 The device uses RF energy only for its internal function.
  • Page 24 3 Vrms 3 Vrms Portable and mobile RF communications equipment 0,15 MHz – 80 MHz 0,15 MHz – 80 MHz should be used no closer to any part of this device, Conducted RF 6 Vrms in ISM bands 6 Vrms in ISM and including cables, than there commended separation IEC 61000-4-6 between 0,15 MHz and...
  • Page 25 Recommended separation distances between portable and mobile RF communications equipment and this device: This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and this device as recommended below, according to the maximum output power of the communications equipment Rated maximum output...