Page 1
Instructions BX43 System microscope To ensure safety, obtain optimum performance and to familiarize yourself fully with the use of this product, we recommend that you study this manual thoroughly before operating this Optical Microscope and Accessory product, and always keep this manual reachable when operating this product.
Page 2
In accordance with European Directive on Waste Electrical and Electronic Equipment, this symbol indicates that the product must not be disposed of as unsorted municipal waste, but should be collected separately. Refer to your local our distributor in EU for return and/or collection systems available in your country.
BX43 Contents Correct assembly and adjustments are critical for the microscope to exhibit its full performance. If you are going to assemble the microscope yourself, please read section “10 Assembly” (Page 34) carefully. Introduction ............................... 1 1 Module nomenclature ........................4 2 Controls .................................
Page 4
Rotating the stage ..................................21 Adjusting the stage height ..............................21 5-4 Observation tube .............................. 22 Adjusting the diopter ................................22 Using the eye shades ................................22 Mounting the eyepiece micrometer disk ......................23 Selecting the light path of the trinocular tube ................23 Adjusting the tilt ...................................24 Using eyepieces incorporating a micrometer ................25 5-5 Condenser ................................
BX43 Introduction Safety precautions (Fig.1) 1. If potentially infectious samples may be observed, use protective gloves or other protective means to prevent the skin from contacting with samples directly. After observation, be sure to clean the portion contacted with samples.
Restrictions in Use When all combinations (@ to ³) below are satisfied at the same time, the microscope may fall if tilted larger than a certain degree, comparing to other combinations. Please mount the camera port of the intermediate observation tube toward diagonally backward 45°...
BX43 Maintenance and storage 1. To clean the lenses and other glass components, simply blow dirt away using a commercially available blower and wipe gently using a piece of cleaning paper (or clean gauze). If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute alcohol.
BX43 Controls If you have not yet assembled the microscope, see “10 Assembly” on page 34. Condenser height adjustment Interpupillary distance adjustment knob (Page 12) scale (Page 11) Allen screwdriver holder Diopter adjustment ring (Page 22) Slide holder clamping lever (Page 19)
Page 10
Main switch (Page 10) : ON : OFF Pre-focusing lever (Page 16) LIM setting switch (Page 13) Fine adjustment knob (Page 16) LED brightness adjustment knob (Page 13) LIM indicator ON : Lit (Green) OFF : Extinguished LIM ON-OFF switch (Page 13)
Page 11
BX43 Swing-out condenser Low magnification condenser U-SC3 U-LC Top lens Top lens swing-out knob Auxiliary lens mounting tool (for BX53) Aperture iris diaphragm ring (Page 12) Auxiliary lens for U-LC Condenser mount dovetail * The auxiliary lens is mounted on the LED lamp housing (U-LHLEDC) or LS30 adapter (U-LS30ADP) when the U-LC is used, see page 35.
Flow of observation · When the LED lamp is used and the LIM is set, the LED brightness adjustment knob is defeated. · When the halogen bulb is used, set the LBD filter in the filter mount on the base (or in the U-FC filter cassette). (Controls used) (Page) LED lamp...
Page 13
BX43 (trinocular tube only) Make a photocopy of the observation procedure pages and post it near your microscope.
Simplified observation procedure 4-1 Basic operation (until observation of specimen) This section describes the basic operation of the microscope until the start of observation of a specimen. For the detailed operating procedure of each control, please read the description page specified below. Press the main switch of the microscope frame to “...
BX43 Rotate the revolving nosepiece to engage the 10X objective in the light path. Rotate the coarse and fine adjustment knobs to bring the specimen in focus. (Details: Page 16) Rotate the stage knob to adjust the observation position. Now you can observe the magnified image of the specimen.
Adjusting the centering Place the specimen. Rotate the revolving nosepiece to select the 10X objective. Rotate the knobs to bring the specimen into focus. Rotate the knob to raise the condenser to its upper limit. Rotate the field iris diaphragm ring in the direction of the arrow so that the diaphragm image comes inside the field of view.
BX43 Using the controls 5-1 Base Adjusting the brightness ( Figs.10 & 11) When the LED lamp is used for observation, turn the LED brightness adjustment knob a clockwise to make illumination brighter. When the halogen bulb is used for observation, adjust the brightness adjustment knob b of the TL4 power supply unit.
Using the filters (halogen bulb operation only) (Figs.13 to 17) You can place a filter in the light path with either method. · Place a filter on the filter mount on the base and engage in the light path. (Fig.13) ·...
Page 19
BX43 Using the filter cassette (Fig.17) Usable Filters Applications 45-LBD-IF* Color balancing filter 45-ND6, 45-ND25 Neutral density filter 45G-530, 43IF550-W45* Green B&W contrast filters 45O-560 Orange Up to three of the above filters can be loaded in the filter cassette.
5-2 Focusing block Focusing controls (Figs.18 & 19) The focusing block consists of the controls listed in the following table, which also describes the functions of the controls. Name Function Coarse Moves the focus position widely. adjustment knob Fine adjustment Fine adjusts the focus position.
BX43 Replacing the fine adjustment knob (Fig.21) CAUTION The fine adjustment knob has been attached on the right side at the factory. The fine adjustment knob is designed detachable to prevent interference with hand during manipulation of the X-and Y-axis knobs.
5-3 Stage Placing the specimen (Figs.24 & 25-A & 25-B) · The dimensions of the slide glass should be 26 x 76 mm with thickness of 0.9 to 1.2 mm, and the cover glass should have thickness of 0.17 mm. ·...
BX43 When using the specimen hold plate (CX3-SHP) For attaching procedures, see "Attaching specimen hold plate" (page 37). Rotate the coarse focusing knob a to lower the stage sufficiently. Place the specimen on the resin sheet b . CAUTION · As the outside of the aperture c of the resin sheet is not illuminated by the light from the condenser, the specimen Fig.25-B...
Adjusting the X- and Y-Axis knob tension (Fig.27) Hold the X-axis knob a and slide up the Y-axis knob b up to expose the adjustment knobs. Turning the X-axis adjustment knob c or Y-axis adjustment knob d clockwise (in the direction of the arrow) increases the tension and counterclockwise decreases it.
BX43 Rotating the stage (Fig.28) Using the Allen screwdriver, slightly loosen the stage clamping screw The stage can be rotated both clockwise and counterclockwise by the stage clamping screw a . CAUTION A click may be heard and felt during rotation. However, this is due to the construction of the substage and does not indicate a malfunction.
5-4 Observation tube Adjusting the diopter (Fig.31) Set the diopter adjustment rings on both sides to scale “0”. Engage a high-power objective (40X or so) in the light path, look into the right eyepiece with your right eye, and rotate the coarse and fine adjustment knobs to bring the specimen into focus.
BX43 Mounting the eyepiece micrometer disk (Fig.33) When the WHN10X-H eyepieces are used, an eyepiece micrometer disk can be mounted. Use 24 mm dia. x 1.5 mm thick micrometer disks. Following Fig.33, turn the built-in micrometer-mounting frame b to the arrow direction to remove it from the eyepiece and place a micrometer disk a into the mounting frame so that the surface with the model indication faces downward.
Adjusting the tilt (Fig.35 to 38) 5° to 35° Adjust the height and tilt of the observation tube to obtain the most comfortable viewing position. U-TBI-3 5° to 35° U-TBI-CLI 5° to 35° U-ETBI 0° to 25° U-TTBI 0° to 25° U-TTR-2 5°...
BX43 Using eyepieces incorporating a micrometer (Fig.39) When the eyepieces in use incorporate a micrometer, the accuracy of the left-right focusing adjustment (diopter adjustment) can be improved further. Looking into the right eyepiece with your right eye, turn the top of the eyepiece a so that the micrometer in the field of view looks sharpest.
5-6 Immersion objectives CAUTION Be sure to use the provided our Immersion oil. Using immersion objectives (Fig.40) Focus on the specimen with objectives in the order of lower-power to higher-power ones. Before engaging the immersion objective, apply a drop of provided immersion oil onto the specimen at the area to be observed.
BX43 Camera recording · Use a trinocular observation tube such as the U-TTR-2 when recording images with camera. The trinocular tube accepts a camera adapter (certain camera adapters necessitate a camera mount adapter). · Be sure to adjust the parfocality before using a camera adapter. Otherwise, the focusing of the camera image will not match that of the image observed through eyepieces.
Troubleshooting guide Under certain conditions, performance of the system may be adversely affected by factors other than defects. If problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact us for assistance. Problem Lamp Cause...
Page 33
BX43 Problem Lamp Cause Remedy Page Dirt or dust is visible in the field Clean thoroughly. Common Dirt/dust on the eyepieces. of view. Common Dirt or dust on condenser top lens. Common Dirt/dust on the specimen. Visibility is poor Common...
Page 34
Problem Lamp Cause Remedy Page Image cannot be focused: Common Tension adjustment ring is too Tighten ring. loose. · Objective drifts down by itself. · Fine adjustment knob slips. Coarse adjustment will not go Common Pre-focusing lever is locked at a Unlock pre-focusing lever.
BX43 Optical characteristics (UIS2 series) The following table shows the optical characteristics Magnification of objective Objective series (PL is described as Plan) of combinations of eyepieces and objectives. Numerical aperture (NA) The figure on the right shows the performance data...
BX43 10-2 Detailed assembly procedures Attaching the auxiliary lens for the U-LC condenser (Fig.42) When a condenser other than the U-LC is to be used, remove the auxiliary lens. Attach the auxiliary lens only when the U-LC is to be used.
Page 40
Installing the halogen bulb (Fig.45) Use only the designated bulb 6V30WHAL (PHILIPS 5761). Holding the bulb b a piece of gauze, insert the bulb pins c straight and fully into the pin holes d on the lamp socket. CAUTION To prevent reduced bulb life or cracking, do not touch the bulb with bare hands.
Page 41
BX43 Attaching the specimen holder (Fig.48 to 53) The attaching procedures are different between the regular specimen holder and the specimen hold plate (CX3-SHP). When using the regular specimen holder (Fig.48, 49) Rotate the coarse focusing knob to lower the stage sufficiently.
Place the metal plate c over the resin sheet (one sheet only) b by matching the hole positions of the metal plate with the hole positions of the resin sheet. Match the hole of the overlapped parts with the hole position of the stage plate and place the overlapped parts on the stage plate.
Use only coded 5 position nosepiece for BX43 (BX43-5RES) · Distribute its cable d toward the rear panel by passing it below the arm, and connect the cable to the connector no the rear panel.
Attaching the AC adapter (Fig.58) CAUTION · Cables and cords are vulnerable when bent or twisted. Never subject them to excessive force. · Make sure that the main switch is set to “ ” (OFF) before connecting the power cord. Insert the connector a of the AC adapter firmly into the connector b on the microscope.
BX43 11 11 Preventive inspection sheet for illumination devices · We recommend performing “Preventive Inspections“ periodically (every time you replace lamps and at least once every 6 months). · The table below identifies the check items to be observed. Put (X) if not applicable or ( ) if applicable.
Proper selection of the power supply cord If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to “Specifications” and “Certified Cord” below: Caution : In case you use a non-approved power supply cord for our products, we can no longer warrant the electrical safety of the equipment.
Page 47
BX43 Table 2 HAR flexible cord Approval organizations and cordage harmonization marking methods Alternative marking utilizing Printed or embossed black-red-yellow thread (Length harmonization marking (May be of color section in mm) Approval organization located on jacket or insulation of internal wiring)
Evident. · Copy the Safety Confirmation shown on the next page, fill it out, and submit it to Evident maintenance personnel. · If the Safety Confirmation is not issued, we may not be able to provide maintenance, repairs or other services.
Page 49
Confirmation of Disinfection/sterilization of this product has not been disinfection/sterilization carried out. of product *We do not accept repairs/inspections of products used in the BSL4 environment. Facility name Department Signature Contact details Telephone, E-mail, etc. Evident Corporation...
Page 50
6666 Inatomi, Tatsuno-machi, Kamiina-gun, Nagano 399-0495, Japan Distributed by Evident Europe GmbH Caffamacherreihe 8-10, 20355 Hamburg, Germany Evident Europe GmbH – UK Branch UK Responsible Person Part 2nd Floor Part A, Endeavour House, Coopers End Road, Stansted CM24 1AL, UK Evident Scientific, Inc.
Need help?
Do you have a question about the BX43 and is the answer not in the manual?
Questions and answers