Teledyne Analytical Instruments 9110E Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Quick Links

NITROGEN OXIDES ANALYZER
© Teledyne Analytical Instruments
Copyright 2001 TAI Inc.
INSTRUCTION MANUAL
MODEL 9110E
(TAI)
16830 Chestnut St.
City of Industry, CA. 91748
USA
Phone:
626-934-1673
Phone:
626-934-9221
626-961-2538
858-657-9816
Website:
http://www.teledyne-ai.com/
044100102
Rev. B
2-June, 2003

Advertisement

Table of Contents

Troubleshooting

loading
Need help?

Need help?

Do you have a question about the 9110E and is the answer not in the manual?

Questions and answers

Summary of Contents for Teledyne Analytical Instruments 9110E

  • Page 1 INSTRUCTION MANUAL MODEL 9110E NITROGEN OXIDES ANALYZER © Teledyne Analytical Instruments (TAI) 16830 Chestnut St. City of Industry, CA. 91748 Phone: 626-934-1673 Phone: 626-934-9221 626-961-2538 858-657-9816 Website: http://www.teledyne-ai.com/ 044100102 Rev. B Copyright 2001 TAI Inc. 2-June, 2003...
  • Page 2: Safety Messages

    M9110E Documentation Model 9110E Instruction Manual SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol.
  • Page 3: Table Of Contents

    Model 9110E Instruction Manual M9110E Documentation TABLE OF CONTENTS SAFETY MESSAGES .......................II TABLE OF CONTENTS ......................III LIST OF FIGURES......................... IX LIST OF TABLES ........................XI LIST OF APPENDICES ....................... XIII 1. M9110E DOCUMENTATION....................1 1.1. Available Documentation ....................1 1.2. Manual Structure......................1 1.3.
  • Page 4 M9110E Documentation Model 9110E Instruction Manual 5.10. Additional Manual (Option 70) ..................43 5.11. Manual on CD (Option 70A)..........Error! Bookmark not defined. 5.12. Extended Warranty (Options 92 & 93) ................43 6. OPERATING INSTRUCTIONS ................... 45 6.1. Overview of Operating Modes ..................45 6.2.
  • Page 5 Model 9110E Instruction Manual M9110E Documentation 6.10.2.3. Editing iDAS Data Channels ................95 6.10.2.4. Trigger Events....................97 6.10.2.5. Editing iDAS Parameters................. 98 6.10.2.6. Sample Period and Report Period..............100 6.10.2.7. Number of Records ..................101 6.10.2.8. RS-232 Report Function ................102 6.10.2.9.
  • Page 6 M9110E Documentation Model 9110E Instruction Manual 8.7.2. Precision Check....................142 8.7.3. Precision Check Procedure ..................142 8.8. Certification of Working Standards ................142 8.8.1. Certification Procedures of Working Standards ............143 8.8.1.1. Other Methods of Establishing Traceability ............143 8.9. References......................... 143 9.
  • Page 7 Model 9110E Instruction Manual M9110E Documentation 10.3.4. Relay Board....................... 182 10.3.4.1. Heater Control .................... 182 10.3.4.2. Valve Control ....................182 10.3.4.3. Status LEDs ....................182 10.3.5. Motherboard...................... 183 10.3.5.1. A to D Conversion..................183 10.3.5.2. Sensor Inputs ..................... 183 10.3.5.3. Thermistor Interface ..................184 10.3.5.4.
  • Page 8 M9110E Documentation Model 9110E Instruction Manual 11.5.7. Keyboard / Display Interface................211 11.5.8. Relay Board ....................... 212 11.5.9. Motherboard ...................... 212 11.5.9.1. A/D functions ....................212 11.5.9.2. Analog Output Voltages ................212 11.5.9.3. Status Outputs .................... 213 11.5.9.4. Control Inputs ..................... 213 11.5.10.
  • Page 9: List Of Figures

    Model 9110E Instruction Manual M9110E Documentation LIST OF FIGURES Figure 3-1: Location of Shipping Screws and Power Configuration Plug ....... 10 Figure 3-2: M9110E Layout ..................11 Figure 3-3: M9110E Rear Panel Layout ................12 Figure 3-4: M9110E Front Panel Layout ................. 12 Figure 3-5: Basic Pneumatic Connections for Units Without Valve Options......
  • Page 10 M9110E Documentation Model 9110E Instruction Manual Figure 10-17: M9110E Electronic Interface Block Diagram ..........187 Figure 10-18: Keyboard and Display Interface Block Diagram..........189 Figure 10-19: Schematic of Basic Software Operation ............190 Figure 11-1: Viewing and Clearing Warning Messages ............194 Figure 11-2: Switching Signal I/O Functions..............196...
  • Page 11 Model 9110E Instruction Manual M9110E Documentation LIST OF TABLES Table 2-1: Model 9110E Basic Unit Specifications ............. 5 Table 3-1: Inlet / Outlet Connector Nomenclature ............13 Table 3-2: Analog Output Pin-Outs................18 Table 3-3: Status Output Signals ................. 18 Table 3-4: Control Input Signals ..................
  • Page 12 M9110E Documentation Model 9110E Instruction Manual Table 11-1: Test Functions - Possible Causes for Out-Of-Range Values ......195 Table 11-2: Relay Board Status LEDs ................199 Table 11-3: DC Power Test Point and Wiring Color Code ..........210 Table 11-4: DC Power Supply Acceptable Levels .............211 Table 11-5: Relay Board Control Devices ...............212...
  • Page 13 Model 9110E Instruction Manual M9110E Documentation LIST OF APPENDICES APPENDIX A: SOFTWARE DOCUMENTATION............229 APPENDIX A-1: M9110E Software Menu Trees and Index, Revision C.8 ......230 APPENDIX A-2: Setup Variables For Serial I/O, Revision C.8..........236 APPENDIX A-3: Warnings and Test Measurements, Revision C.8 ........245 APPENDIX A-4: M9110E Signal I/O Definitions, Revision C.8..........
  • Page 14: User Notes

    M9110E Documentation Model 9110E Instruction Manual USER NOTES: M9110E Rev B...
  • Page 15: M9110E Documentation

    Model 9110E Instruction Manual M9110E Documentation 1. M9110E DOCUMENTATION 1.1. Available Documentation Thank you for purchasing the Model 9110E Nitrogen Oxides Analyzer! The documentation for this instrument is available in different formats: • Printed format, part number 044100100 Additional documentation for the Model 9110E Nitrogen Oxides Analyzer is available from TAI’s website at http://www.teledyne-ai.com/manuals/...
  • Page 16 M9110E Documentation Model 9110E Instruction Manual 2.0 Specifications and Warranty A list of the analyzer’s performance specifications, a description of the conditions and configuration under which EPA equivalency was approved and TAI’s warranty statement. 3.0 Getting Started Concise instructions for setting up, installing and running your analyzer for the first time.
  • Page 17: How To Use This Instruction Manual

    Model 9110E Instruction Manual M9110E Documentation version-specific software menu trees, warning messages, definitions of iDAS & serial I/O variables as well as spare part listings, repair questionnaire, interconnect drawing, detailed pneumatic and electronic schematics. 1.3. How to use this Instruction Manual Throughout this manual, words printed in capital, bold letters, such as SETUP or ENTR represent messages as they appear on the analyzer’s front panel display.
  • Page 18 M9110E Documentation Model 9110E Instruction Manual USER NOTES: M9110E Rev B...
  • Page 19: Specifications, Approvals And Warranty

    Model 9110E Instruction Manual Specifications, Approvals and Warranty 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. M9110E Operating Specifications Table 2-1: Model 9110E Basic Unit Specifications Min/Max Range Min: 0-50 ppb (Physical Analog Output) Max: 0-20 ppm Measurement Units ppb, ppm, µg/m...
  • Page 20: Epa Equivalency Designation

    Specifications, Approvals and Warranty Model 9110E Instruction Manual 2.2. EPA Equivalency Designation The Model 9110E Analyzer is designated as Reference Method Number RFNA-1194-099 (same designation as model M200A) as per 40 CFR Part 53 when operated under the following conditions: •...
  • Page 21: Ce Mark Compliance

    Model 9110E Instruction Manual Specifications, Approvals and Warranty 2.3. CE Mark Compliance The Teledyne Instruments Nitrogen Oxides Analyzers M9110E, M9110EH and M9110EM were tested and found to be fully compliant with: EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) &...
  • Page 22 Specifications, Approvals and Warranty Model 9110E Instruction Manual THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED.
  • Page 23: Getting Started

    Model 9110E Instruction Manual Getting Started 3. GETTING STARTED 3.1. Unpacking and Initial Setup CAUTION The M9110E weighs about 17 kg (40 pounds) without options installed. To avoid personal injury, we recommend to use two persons to lift and carry the analyzer.
  • Page 24: Figure 3-1: Location Of Shipping Screws And Power Configuration Plug

    Getting Started Model 9110E Instruction Manual Figure 3-1: Location of Shipping Screws and Power Configuration Plug NOTE Save these shipping screws and re-install them whenever the unit is shipped. A certain ventilation clearance is required for the operation of the analyzer:...
  • Page 25: M9110E Layout

    Model 9110E Instruction Manual Getting Started 3.2. M9110E Layout Figure 3-2 shows a top-down view of the analyzer. The shown configuration includes the Ethernet board, IZS option, zero-air scrubber and an additional sample dryer. See Chapter 5 for optional equipment. Figure 3-3 shows the rear panel configuration with optional zero-air scrubber mounted to it and two optional fittings for the IZS option.
  • Page 26: Figure 3-3: M9110E Rear Panel Layout

    Getting Started Model 9110E Instruction Manual Figure 3-3: M9110E Rear Panel Layout Figure 3-4: M9110E Front Panel Layout M9110E Rev B...
  • Page 27: Pneumatic Connections

    Sample and calibration gases should only come into contact with PTFE (Teflon), FEP, glass or stainless steel materials. Figure 3-5 and Figure 3-6 illustrate the most common configurations for gas supply and exhaust lines to the Model 9110E Analyzer. Appendix D contains more detailed pneumatic flow diagrams for the analyzer and its various (pneumatically related) options.
  • Page 28: Span Gas Defined

    Getting Started Model 9110E Instruction Manual • Attach the 1/4" exhaust line to the exhaust port of the analyzer and to the inlet port of the pump. • Attach zero air and span gas supply lines as appropriate (Figure 3-5 and Figure 3-6) to the rear panel (Figure 3-3).
  • Page 29: Zero Gas Defined

    Model 9110E Instruction Manual Getting Started For example, if the measurement is NO in ambient air between 0 and 500 ppb, an appropriate span gas would be NO in air at 400 ppb NO concentration (80% of maximum range). Even though NO gas in nitrogen could be used as a span gas, the matrix of the balance gas is different and may cause interference problems or yield incorrect calibrations.
  • Page 30: Figure 3-6: Basic Pneumatic Connections For Units With Valve Options

    Getting Started Model 9110E Instruction Manual Zero/Span Valves – Option 50 Source of MODEL 700 SAMPLE Gas Gas Dilution Calibrator VENT if input is pressurized (with O generator option) Sample VENT MODEL Exhaust Pump 9110E Span External Zero Air Scrubber...
  • Page 31: Electrical Connections

    Model 9110E Instruction Manual Getting Started 3.4. Electrical Connections Refer to Figure 3-3 for the location of the rear panel electrical and pneumatic connections. 3.4.1. Power Connection Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A current at your AC voltage and that it is equipped with a functioning earth ground.
  • Page 32: Figure 3-8: Status Output Connector

    Getting Started Model 9110E Instruction Manual Table 3-2: Analog Output Pin-Outs Analog Output Data Type Voltage Signal Current Signal V Out I Out + Ground I Out - V Out I Out + Ground I Out - V Out I Out +...
  • Page 33: Figure 3-9: Control Input Connector

    Model 9110E Instruction Manual Getting Started If you wish to use the analyzer to remotely activate the zero and span calibration modes, several digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
  • Page 34: Ethernet Connection And Configuration

    Getting Started Model 9110E Instruction Manual speed, 8 data bits, no parity, 1 stop bit and Xon/Xoff flow control. Use the APICOM remote control software that was included with the analyzer to connect to the instrument or use any terminal emulation program. Refer to Section 6.9 of this manual for instructions on configuration and usage of the serial port, Section 6.11 for remote operation of the...
  • Page 35: Warm-Up

    Model 9110E Instruction Manual Getting Started System waits 3 seconds then SELECT DESIRED FUNCTION automatically begins its initialization routine. No action required. START CHECKING FLASH STATUS System is checking the format of the instrument’s flash memory chip. If at this point,...
  • Page 36: Warning Messages

    Getting Started Model 9110E Instruction Manual Name Color Behavior Significance STATUS LEDs Unit is operating in sample mode, front panel Sample Green display is continuously updated. Yellow The instrument’s calibration is not enabled. The analyzer is warming up and out of...
  • Page 37: Functional Check

    Model 9110E Instruction Manual Getting Started Table 3-6: Possible Warning Messages at Start-Up Message Definition ANALOG CAL WARNING The instrument’s A/D circuitry or one of its analog outputs is not calibrated. AZERO WRN XXX.X MV The reading taken during the auto-zero cycle is outside of specified limits.
  • Page 38 Getting Started Model 9110E Instruction Manual SAMPLE RANGE = 500.0 PPB NOX = X.X < TST TST > CAL SETUP RANGE NOX STB SAMP FLW Toggle <TST TST> keys to OZONE FL scroll through list of functions NORM PMT AZERO...
  • Page 39: Figure 3-11: M9110E Pneumatic Diagram In Standard Configuration

    Model 9110E Instruction Manual Getting Started Figure 3-11: M9110E Pneumatic Diagram In Standard Configuration M9110E Rev B...
  • Page 40: Figure 3-12: M9110E Pneumatic Diagram With Options Installed

    Getting Started Model 9110E Instruction Manual Figure 3-12: M9110E Pneumatic Diagram With Options Installed M9110E Rev B...
  • Page 41: First Calibration

    Model 9110E Instruction Manual Getting Started 3.6. First Calibration 3.6.1. Basic Calibration Procedure The initial calibration should be carried out using the 500 ppb range, a span gas with about 400 ppb NO and with the unit set for SINGLE range, which will enable you to compare your calibration results to the factory calibration as listed on the Final Test and Validation Data Sheet .
  • Page 42 Getting Started Model 9110E Instruction Manual Set the expected NO and NOx span gas concentration. If you supply NO gas to the analyzer, the values for expected NO and NO need to be identical. SAMPLE RANGE = 500.0 PPB NOX= X.X <...
  • Page 43 M-P CAL RANGE = 500.0 PPB 03 = EXIT returns to the main SAMPLE display < TST TST > ENTR CONC EXIT The Model 9110E analyzer is now ready for operation. M9110E Rev B...
  • Page 44: Interferences For No Measurements

    ) and carbon dioxide (CO ) but the Model 9110E has been designed to reject most of these interferences. Section 10.1.5 contains more detailed information on interferences. Ammonia is the most common interferent, which is converted to NO in the analyzer’s NO converter and creates a NO signal artifact.
  • Page 45: Frequently Asked Questions & Glossary

    Model 9110E Instruction Manual Frequently Asked Questions & Glossary 4. FREQUENTLY ASKED QUESTIONS & GLOSSARY 4.1. Frequently Asked Questions The following list contains some of the most commonly asked questions relating to the Model 9110E NO Analyzer. Q: Why is the ZERO or SPAN key not displayed during calibration? A: The M9110E disables certain keys whenever the chosen value is out of range for that particular parameter.
  • Page 46: Glossary

    Frequently Asked Questions & Glossary Model 9110E Instruction Manual Q: How often do I need to change the particulate filter? A: Once per week. Table 9-1 contains a maintenance schedule listing the most important, regular maintenance tasks. Highly polluted sample air may require more frequent changes.
  • Page 47 Model 9110E Instruction Manual Frequently Asked Questions & Glossary parts that acts as a computer hard disk drive under DOS with disk drive label “C”. DOC chips come with 8 mb space in the E-series analyzer standard configuration but are available in larger sizes DOS - Disk Operating System , the operating system underlying the M9110E firmware.
  • Page 48: User Notes

    Frequently Asked Questions & Glossary Model 9110E Instruction Manual USER NOTES: M9110E Rev B...
  • Page 49: Optional Hardware And Software

    5. OPTIONAL HARDWARE AND SOFTWARE This section includes a descriptions of the hardware and software options available for the Model 9110E Nitrogen Oxides Analyzer. For assistance with ordering these options please contact the sales department of Teledyne Analytical Instruments at:...
  • Page 50: Current Loop Analog Outputs (Option 41)

    Optional Hardware and Software Model 9110E Instruction Manual accommodate a hand for transport. When pushed in, the handle is flush with the chassis and protrudes out about 9 mm / 3/8”. Installing the strap handle prevents the use of the rack mount slides, although the rack mount brackets, Option 21, can still be used.
  • Page 51: Particulate Filter Kit (Option 42A)

    Figure 3-12 shows the internal, pneumatic connections for a Model 9110E with the zero/span valve option installed. Table 5-1 describes the state of each valve during the analyzer’s various operational modes.
  • Page 52: Internal Zero/Span (Izs) (Option 51)

    Optional Hardware and Software Model 9110E Instruction Manual Table 5-1: IZS or Z/S Valve States Valve Port Connection Mode Valve Condition NO = normally open port NC = normally closed port COM = common port SAMPLE Sample/Cal Open to sample gas inlet...
  • Page 53: Izs Permeation Tubes (Options 53 & 55)

    Figure 3-12 shows the internal pneumatic connections for a Model 9110E with IZS option installed. Table 5-1 shows the operational state of the valves associated with the IZS option during the analyzer’s various operational modes.
  • Page 54: Scrubbers And Expendables

    Optional Hardware and Software Model 9110E Instruction Manual 5.7. Scrubbers and Expendables 5.7.1. Charcoal Scrubber (Option 64A) This kit includes a charcoal scrubber cartridge, which is used to remove NO from an exhaust port. Applications include, but are not limited to, the removal of NO...
  • Page 55: M9110E Expendables Kit (Option 42)

    Model 9110E Instruction Manual Optional Hardware and Software 5.7.5. M9110E Expendables Kit (Option 42) This kit includes a recommended set of expendables for one year of operation of the M9110E. See Appendix B for a detailed listing of the contents.
  • Page 56: Sample Gas Conditioners (Options 86 & 88)

    Optional Hardware and Software Model 9110E Instruction Manual Figure 5-3: M9110E Ethernet Card and Rear Panel With Ethernet Installed 5.9. Sample Gas Conditioners (Options 86 & 88) ® Several permeation devices using Nafion permeation gas exchange tubes are available for applications with high moisture and/or moderate levels of NH in the sample gas.
  • Page 57: Additional Manual (Option 70)

    Model 9110E Instruction Manual Optional Hardware and Software oxygen need to use the separate dryer option 86. The combination conditioner needs to be specified upon ordering the analyzer. 5.10. Additional Manual (Option 70) Additional copies of the printed user’s manual can be purchased from the factory. Please specify the serial number of your analyzer so that we can match the manual version.
  • Page 58 Optional Hardware and Software Model 9110E Instruction Manual USER NOTES: M9110E Rev B...
  • Page 59: Operating Instructions

    Model 9110E Instruction Manual Operating Instructions 6. OPERATING INSTRUCTIONS To assist in navigating the analyzer’s software, a series of menu trees can be found in Appendix A of this manual along with an index of software commands and references to the respective manual sections.
  • Page 60: Sample Mode

    Operating Instructions Model 9110E Instruction Manual or the COM port settings. The SETUP mode is also used for performing various diagnostic tests during troubleshooting. The third operating mode is the CAL mode, which allows calibration of the analyzer in various ways. Because of its importance, this mode is described separately in Chapter 7.
  • Page 61: Table 0-2: Test Functions Defined

    Model 9110E Instruction Manual Operating Instructions Table 0-2: Test Functions Defined Display Parameter Units Description The full scale limit to which the analyzer’s analog outputs are currently set. This is not the physical range of the instrument. If PPB, PPM,...
  • Page 62: Warning Messages

    Operating Instructions Model 9110E Instruction Manual 6.2.2. Warning Messages The most common instrument failures will be reported as a warning on the analyzer’s front panel and through the COM ports. Section 11.1 explains how to use these messages to troubleshoot problems. Section 3.5.3 shows how to view and clear warning messages.
  • Page 63: Calibration Mode

    Model 9110E Instruction Manual Operating Instructions Chapter 7 details all calibration operations, Chapter 1 describes the zero/span valve and IZS options. 6.3. Calibration Mode Owing to its importance, calibration is described separately in Chapter 7, EPA protocol calibration in Chapter 8.
  • Page 64 Operating Instructions Model 9110E Instruction Manual To enable or disable passwords, press the following keystroke sequence: SAMPLE RANGE = 500.0 PPB =XXX.X < TST TST > CAL SETUP SETUP X.X PRIMARY SETUP MENU Exit returns to SAMPLE display CFG DAS RNGE PASS CLK MORE EXIT SETUP X.X...
  • Page 65: Configuration Information (Cfg)

    Model 9110E Instruction Manual Operating Instructions 6.4.2. Configuration Information (CFG) Pressing the CFG key displays the instrument configuration information. This display lists the analyzer model, serial number, firmware revision, software library revision, operating system and other information. Use this information to identify the software and hardware when contacting customer service.
  • Page 66: Setup - Range Configuration (Rnge)

    Operating Instructions Model 9110E Instruction Manual In order to compensate for CPU clocks which run fast or slow, there is a variable to speed up or slow down the clock by a fixed amount every day. To change this variable, press: SAMPLE RANGE = 500.0 PPB...
  • Page 67: Physical And Analog Output Ranges

    <TST> keys of the unit’s SAMPLE mode display. A4 is not available for the current loop option. 6.5.1. Physical and Analog Output Ranges Functionally, the model 9110E NO analyzer has two physical ranges for determining NO NO and NO concentrations.
  • Page 68: Single Range Mode (Sngl)

    Operating Instructions Model 9110E Instruction Manual 6.5.3. Single Range mode (SNGL) The default range mode for the analyzer is single range, in which all analog concentration outputs (usually A1, A2 and A3) are set to the same reporting range. This reporting range can be set to any value between 100 ppb and 20000 ppb.
  • Page 69 Model 9110E Instruction Manual Operating Instructions Setting analog range limits to different values does not affect the instrument’s calibration. To select the IND range mode, press the following keys: SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL...
  • Page 70: Auto Range Mode (Auto)

    Operating Instructions Model 9110E Instruction Manual 6.5.5. Auto Range Mode (AUTO) In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined ranges (low and high). The same low and high span settings are applied equally to NO, NO and NO readings.
  • Page 71: Dilution Ratio

    Model 9110E Instruction Manual Operating Instructions SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE EXIT EXIT returns SETUP X.X RANGE CONTROL MENU to the main menu.
  • Page 72 Operating Instructions Model 9110E Instruction Manual • Select the range: Use the procedures in Section 6.6. Make sure that the SPAN value entered is the maximum expected concentration of the undiluted calibration gas and that the span gas is either supplied through the same dilution inlet system as the sample gas or has an appropriately lower actual concentration.
  • Page 73: Setup - Internal Variables (Vars)

    Model 9110E Instruction Manual Operating Instructions 6.6. Setup - Internal Variables (VARS) The M9110E has several-user adjustable software variables, which define certain operational parameters. Usually, these variables are automatically set by the instrument’s firmware, but can be manually re-defined using the VARS menu. Table 0-5 lists all variables that are available within the 818 password protected level.
  • Page 74 Operating Instructions Model 9110E Instruction Manual To access and navigate the VARS menu, use the following key sequence: SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE...
  • Page 75: Setup - Diagnostics (Diag)

    Model 9110E Instruction Manual Operating Instructions 6.7. Setup - Diagnostics (DIAG) A series of diagnostic tools is grouped together under the SETUP-MORE-DIAG menu. As these parameters are dependent on firmware revision, the menu trees are listed in detail in Appendix A. The individual parameters, however, are explained in more detail in the following section, indicated in Table 0-6.
  • Page 76: Signal I/O

    Operating Instructions Model 9110E Instruction Manual To access the DIAG functions press the following keys: SAMPLE RANGE = 500.0 PPB NOX=X.X DIAG ANALOG I / O CONFIGURATION < TST TST > CAL SETUP PREV NEXT ENTR EXIT DIAG TEST CHAN OUTPUT...
  • Page 77: Analog Output Step Test

    Model 9110E Instruction Manual Operating Instructions To enter the signal I/O test mode, press: SAMPLE RANGE = 500.0 PPB NOX=X.X DIAG SIGNAL I / O Use the NEXT & PREV keys to move between < TST TST > CAL SETUP...
  • Page 78: Analog I/O Configuration

    Operating Instructions Model 9110E Instruction Manual 6.7.3. Analog I/O Configuration Table 0-7 lists the analog I/O functions that are available in the M9110E. Table 0-7: DIAG - Analog I/O Functions Sub Menu Function AOUTS Shows the status of the analog output calibration (YES/NO) and initiates a CALIBRATED: calibration of all analog output channels.
  • Page 79: Analog Output Signal Type And Range Selection

    Model 9110E Instruction Manual Operating Instructions The following DC current output limits apply to the current loop modules: Table 0-9: Analog Output Current Loop Range Range Minimum Output Maximum Output 0-20 mA 0 mA 20 mA These are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and upper limits.
  • Page 80: Analog Output Calibration Mode

    Operating Instructions Model 9110E Instruction Manual 6.7.3.2. Analog Output Calibration Mode The analog outputs can be calibrated automatically or manually. In its default mode, the instrument is configured for automatic calibration of all channels. Manual calibration should be used for the 0.1V range or in cases where the outputs must be closely matched to the characteristics of the recording device.
  • Page 81 Model 9110E Instruction Manual Operating Instructions DIAG ANALOG I / O CONFIGURATION EXIT to Return to the main PREV NEXT ENTR EXIT Sample Display DIAG AIO AOUTS CALIBRATED: NO < SET> EXIT Press SET> to select the Analog Output channel to be configured.
  • Page 82: Manual Analog Output Calibration

    Operating Instructions Model 9110E Instruction Manual 6.7.3.3. Manual Analog Output Calibration For highest accuracy, the voltages of the analog outputs can be manually calibrated. Note that outputs configured for 0.1V full scale should always be calibrated manually. Calibration is done through the instrument software with a voltmeter connected across the output terminals (Figure 0-3).
  • Page 83: Analog Output Offset Adjustment

    Model 9110E Instruction Manual Operating Instructions FROM ANALOG I/O CONFIGURATION MENU DIAG AIO CONC_OUT_1 RANGE: 5V DIAG ANALOG I / O CONFIGURATION SET> EDIT EXIT PREV NEXT ENTR EXIT DIAG AIO CONC_OUT_1 REC OFS: 0 mV DIAG AIO AOUTS CALIBRATED: NO <...
  • Page 84: Current Loop Output Adjustment

    Operating Instructions Model 9110E Instruction Manual FROM ANALOG I/O CONFIGURATION MENU DIAG ANALOG I / O CONFIGURATION PREV NEXT ENTR EXIT DIAG AIO AOUTS CALIBRATED: NO < SET SET> EXIT Press SET> to select the analog output channel to be configured. Then press...
  • Page 85: Figure 0-3: Setup For Calibrating Current Outputs

    Model 9110E Instruction Manual Operating Instructions See Table 3-2 for pin assignments of the Analog Out connector on the rear panel. V OUT + I IN + V OUT - I IN - Recording Analyzer Device Figure 0-3: Setup for Calibrating Current Outputs CAUTION Do not exceed 60 V between current loop outputs and instrument ground.
  • Page 86: Ain Calibration

    Operating Instructions Model 9110E Instruction Manual Table 0-12: Current Loop Output Calibration with Resistor Voltage for 2-20 mA Voltage for 4-20 mA Full scale (measured across resistor) (measured across resistor) 0.5 V 1.0 V 100% 5.0 V 5.0 V 6.7.3.6. AIN Calibration This is the sub-menu to conduct the analog input calibration.
  • Page 87: Table 0-13: Test Parameters Available For Analog Output A4

    Model 9110E Instruction Manual Operating Instructions To activate the A4 channel and select a test function, follow this key sequence: SAMPLE RANGE = 500.0 PPB NOX=X.X Continue to press NEXT until … < TST TST > CAL SETUP DIAG TEST CHAN OUTPUT PRIMARY SETUP MENU SETUP X.X...
  • Page 88: Optic Test

    Operating Instructions Model 9110E Instruction Manual 6.7.5. Optic Test The optic test function tests the response of the PMT sensor by turning on an LED located in the cooling block of the PMT (Fig. 10-15). The analyzer uses the light emitted from the LED to test its photo-electronic subsystem, including the PMT and the current to voltage converter on the pre-amplifier board.
  • Page 89: Ozone Generator Override

    Model 9110E Instruction Manual Operating Instructions To activate the electrical test press the following keys. SAMPLE RANGE = 500.0 PPB NOX=X.X DIAG SIGNAL I / O < TST TST > CAL SETUP PREV NEXT JUMP ENTR EXIT SETUP X.X PRIMARY SETUP MENU Press NEXT until…...
  • Page 90: Flow Calibration

    Operating Instructions Model 9110E Instruction Manual 6.7.8. Flow Calibration The flow calibration allows the user to adjust the values of the sample and ozone flow rates as they are displayed on the front panel and reported through COM ports to match the actual flow rate measured at the respective pneumatic ports.
  • Page 91: Control Inputs

    Model 9110E Instruction Manual Operating Instructions The status outputs are accessed through a 12 pin connector on the analyzer’s rear panel labeled STATUS. The function of each pin is defined in Table 0-14. STATUS Figure 0-4: Status Output Connector Table 0-14: Status Output Pin Assignments...
  • Page 92: Setup - Communication Ports (Comm)

    Operating Instructions Model 9110E Instruction Manual Table 0-15: Control Input Pin Assignments Input Status Condition when enabled EXTERNAL ZERO Zero calibration mode is activated. The mode field of the display will read ZERO CAL R. EXTERNAL SPAN Span calibration mode is activated. The mode field of the display will read SPAN CAL R.
  • Page 93: Analyzer Id

    Model 9110E Instruction Manual Operating Instructions however, can be configured for half-duplex RS-485 communication or can be used for the TAI Ethernet interface card (optional equipment, Section 5.7.3). There are two options to connect multiple analyzers to a single computer terminal or data logging device over a single serial communications line.
  • Page 94: Com Port Cable Connections

    Operating Instructions Model 9110E Instruction Manual CAUTION Cables that appear to be compatible because of matching connectors may incorpo- rate internal wiring that make the link inoperable. Check cables acquired from sources other than TAI for pin assignments before using.
  • Page 95: Dte And Dce Communication

    Model 9110E Instruction Manual Operating Instructions Table 0-16: COM1 and COM2 DB-9 Pin Assignments Pin # COM1 (RS-232) COM2 (RS-232) COM2 (RS-485) Not used Not used Not used TRANSMIT DATA* RECEIVE DATA DATA - RECEIVE DATA* TRANSMIT DATA DATA +...
  • Page 96: Table 0-17: Com Port Communication Modes

    Operating Instructions Model 9110E Instruction Manual Table 0-17: COM Port Communication Modes Mode Description Quiet mode Quiet mode suppresses any feedback from the analyzer (iDAS reports, and warning messages) to the remote device and is typically used when the port is communicating with a computer program such as APICOM.
  • Page 97 Model 9110E Instruction Manual Operating Instructions To select a set of communication modes for a one of the COMM Ports, press: SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X CFG DAS RNGE PASS CLK MORE...
  • Page 98: Com Port Baud Rate

    Operating Instructions Model 9110E Instruction Manual 6.9.7. COM Port Baud Rate To select the baud rate of one of the COM Ports, press: SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X...
  • Page 99: Ethernet Port Configuration

    Model 9110E Instruction Manual Operating Instructions To initiate the test press the following key sequence. SAMPLE RANGE = 500.0 PPB NOX =XXX.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X COM1 MODE:0 SETUP X.X SET> EDIT EXIT...
  • Page 100 Operating Instructions Model 9110E Instruction Manual Ethernet and is used to set configuration parameters that enable the Ethernet card to communicate with your local area network. To perform this configuration, you will need to get the following information from your network administrator: •...
  • Page 101 Model 9110E Instruction Manual Operating Instructions Step two in the configuration process is to edit the TCP/IP parameters. Internet Configuration SAMPLE RANGE = 500.0 PPB O3 =XXX.X Keypad Functions < TST TST > CAL SETUP FUNCTION <CH Moves the cursor one PRIMARY SETUP MENU SETUP X.X...
  • Page 102: Hessen Mode Configuration

    Operating Instructions Model 9110E Instruction Manual Step three in the configuration process is to restart the pass-through mode for normal operation. SAMPLE RANGE = 500.0 PPB O3 =XXX.X SETUP X.X COMMUNICATIONS MENU < TST TST > CAL SETUP INET COM1...
  • Page 103: Setup - Data Acquisition System (Das)

    Model 9110E Instruction Manual Operating Instructions 6.10. Setup - Data Acquisition System (DAS) The M9110E analyzer contains a flexible and powerful, internal data acquisition system (iDAS) that enables the analyzer to store concentration and calibration data as well as a host of diagnostic parameters.
  • Page 104: Idas Structure

    Operating Instructions Model 9110E Instruction Manual 6.10.1. iDAS Structure The iDAS is designed around the feature of a “record”. A record is a single data point of one parameter, stored in one (or more) data channels and generated by one of several triggering event.
  • Page 105: Idas Parameters

    Model 9110E Instruction Manual Operating Instructions Property Description Default Setting Range channel values to the RS-232 ports. COMPACT Enables a compact version of the RS-232 OFF or ON REPORT reporting with up to five parameters sharing the same output line.
  • Page 106: Idas Triggering Events

    Operating Instructions Model 9110E Instruction Manual 6.10.1.3. iDAS Triggering Events Triggering events define when and how the iDAS records a measurement of any given data channel. Triggering events are firmware-specific and are listed in Appendix A-5. The most common triggering events are: •...
  • Page 107: Table 0-22: M9110E Default Idas Configuration, Revision C.7

    Model 9110E Instruction Manual Operating Instructions Table 0-22: M9110E Default iDAS Configuration, Revision C.7. List of Parameters List of Channels NAME – MODE/EVENT - PRECISION - NUM SAMPLES Name: CONC NOXCNC1 - AVG - 4 - ON Event: ATIMER Sample Period: 000:00:01...
  • Page 108: Viewing Idas Data And Settings

    Please be aware that all stored data will be erased if the analyzer’s disk-on-chip, CPU board or configuration is replaced/reset. Teledyne Analytical Instruments recommends to download and store CONC, CALDAT and ENGDAT data once per month for permanent documentation and future data analysis. Data can be most conveniently downloaded, graphed and stored using the APICOM remote control program.
  • Page 109: Editing Idas Data Channels

    Model 9110E Instruction Manual Operating Instructions SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL SETUP VIEW KEYPAD FUNCTIONS FUNCTION PRIMARY SETUP MENU <PRM Moves to the next Parameter SETUP X.X PRM> Moves to the previous EXIT will...
  • Page 110 Operating Instructions Model 9110E Instruction Manual SAMPLE RANGE = 500.0 PPB NOX=X.X < TST TST > CAL SETUP PRIMARY SETUP MENU SETUP X.X EXIT will return to the previous SAMPLE CFG DAS RNGE PASS CLK MORE EXIT display. SETUP X.X...
  • Page 111: Trigger Events

    Model 9110E Instruction Manual Operating Instructions To edit the name of a data channel, follow the above key sequence and then press: FROM THE PREVIOUS KEY SEQUENCE … SETUP X.X NAME:CONC <SET SET> EDIT PRINT EXIT ENTR accepts the new string and returns to the SETUP X.X...
  • Page 112: Editing Idas Parameters

    Operating Instructions Model 9110E Instruction Manual 6.10.2.5. Editing iDAS Parameters Data channels can be edited individually from the front panel without affecting other data channels. However, when editing a data channel, such as during adding, deleting or editing parameters, all data for that particular channel will be lost, because the iDAS can store only data of one format (number of parameter columns etc.) for any given channel.
  • Page 113 Model 9110E Instruction Manual Operating Instructions To configure the parameters for a specific data parameter, press: FROM THE EDIT DATA PARAMETER MENU (see previous section) SETUP X.X 0) PARAM=PMTDET, MODE=INST PREV NEXT DEL EDIT EXIT SETUP X.X PARAMETERS: PMTDET SET> EDIT EXIT SETUP X.X...
  • Page 114: Sample Period And Report Period

    Operating Instructions Model 9110E Instruction Manual 6.10.2.6. Sample Period and Report Period The iDAS defines two principal time periods, the SAMPLE PERIOD, which is the time interval at which data are sampled (in volatile memory; not stored permanently), and the REPORT PERIOD, which is the time interval of which data are permanently stored to the Disk-on-Chip data storage and reported to the communication ports.
  • Page 115: Number Of Records

    Model 9110E Instruction Manual Operating Instructions To define the SAMPLE or REPORT PERIOD, press SETUP-DAS-EDIT-ENTR and the following keys: FROM THE DAS PASSWORD MENU Edit Data Channel Menu SETUP X.X 0) CONC: ATIMER, Use the PREV and NEXT Exits to the main...
  • Page 116: Report Function

    Operating Instructions Model 9110E Instruction Manual FROM THE DAS PASSWORD MENU Edit Data Channel Menu SETUP X.X 0) CONC: ATIMER, Exits to the main Data Acquisition PREV NEXT DEL EDIT PRNT EXIT menu SETUP X.X NAME:CONC <SET SET> EDIT PRINT EXIT Press SET>...
  • Page 117: Compact Report

    Model 9110E Instruction Manual Operating Instructions To enable automatic COM port reporting, press SETUP-DAS-ENTR and then: FROM THE DAS PASSWORD MENU Edit Data Channel Menu SETUP X.X 0) CONC: ATIMER, Exits to the main Data Acquisition PREV NEXT DEL EDIT...
  • Page 118: Holdoff Feature

    Operating Instructions Model 9110E Instruction Manual To disable a data channel, press SETUP-DAS-ENTR and the following keys: FROM THE DAS PASSWORD MENU Edit Data Channel Menu SETUP X.X 0) CONC: ATIMER, Exits to the main Data Acquisition PREV NEXT DEL EDIT...
  • Page 119: Remote Idas Configuration

    Model 9110E Instruction Manual Operating Instructions 6.10.3. Remote iDAS Configuration Editing channels, parameters and triggering events as described in Section 6.9.10 is much more conveniently done in one step through the APICOM remote control program using the graphical interface shown in Figure 0-6. Refer to the next Section 6.11 for details on remote access to the M9110E analyzer.
  • Page 120: Remote Operation Of The Analyzer

    6.11.1. Basic Operation 6.11.1.1. Terminal Operating Modes The Model 9110E can be remotely configured, calibrated or queried for stored data through the serial ports. As terminals and computers use different communication schemes, the analyzer supports two communicate modes specifically designed to interface with these two types of devices.
  • Page 121: Help Commands In Terminal Mode

    Model 9110E Instruction Manual Operating Instructions Computer mode is used when the analyzer is connected to a computer with a dedicated interface program such as APICOM. More information regarding APICOM can be found in later in this section or on the TAI website at http://www.teledyne- api.com/software/apicom/.
  • Page 122: Data Types

    Operating Instructions Model 9110E Instruction Manual executed. Press ? <CR> or refer to Appendix A-6 for a list of available command designators. <CR> is a carriage return. All commands must be terminated by a carriage return (usually achieved by pressing the ENTER key on a computer).
  • Page 123: Remote Access By Modem

    Model 9110E Instruction Manual Operating Instructions instrument and the download of data in electronic format). You can effectively disable the reporting feature by setting the interface to quiet mode (Section 6.9.6, Table 0-18). Status reports include iDAS data (when reporting is enabled), warning messages, calibra- tion and diagnostic status messages.
  • Page 124: Com Port Password Security

    LOGON SUCCESSFUL - Correct password given. • LOGON FAILED - Password not given or incorrect. • LOGOFF SUCCESSFUL - Connection terminated successfully. To log on to the model 9110E analyzer with SECURITY MODE feature enabled, type: LOGON 940331 M9110E Rev B...
  • Page 125: Apicom Remote Control Program

    Ethernet. Section 6.10.3 and Figure 0-6 showed the remote iDAS configuration through APICOM, Figure 0-8 shows the main interface, which resembles and emulates the actual front panel functionality. Note: 9110E and M200E are used interchangeably.
  • Page 126: Additional Communications Documentation

    Operating Instructions Model 9110E Instruction Manual • Retrieve, view, edit, save and upload iDAS configurations. APICOM is very helpful for initial setup, data analysis, maintenance and trouble-shooting. APICOM is included free of cost with the analyzer or can be downloaded for free at http://www.teledyne-api.com/software/apicom/.
  • Page 127: Calibration Procedures

    If this has not been done, please do so before continuing (Section 6.5 for instructions). 7.1.1. Required Equipment, Supplies, and Expendables Calibration of the Model 9110E analyzer requires a certain amount of equipment and supplies. These include, but are not limited to, the following: •...
  • Page 128: Permeation Tubes

    The currently programmed CE is recorded along with the calibration data in the iDAS for documentation and performance analysis (Section 6.10.2.1). The following procedure will cause the Model 9110E to automatically calculate the current conversion efficiency. Step one is to connect a source of calibrated NO span gas as shown below.
  • Page 129: Figure 7-1: Gas Supply Setup For Determination Of No

    Model 9110E Instruction Manual Calibration Procedures Calibrated NO (high concentration) Source of SAMPLE Gas Gas Dilution removed Calibrator during (with O generator procedure option) VENT Zero Air Source Sample MODEL Exhaust Pump Span 9110E Zero Air Figure 7-1: Gas Supply Setup for Determination of NO...
  • Page 130: Manual Calibration

    SAMPLE display. ENTR EXIT 7.2. Manual Calibration The following section describes the basic method for manually calibrating the Model 9110E analyzer. A note on the difference between a calibration and calibration check: pressing the ENTR key during the following procedures recalculates the stored values for OFFSET and SLOPE (instrument response curve) and alters the instrument’s calibration.
  • Page 131: Connect Zero Air And Span Gases To The Analyzer

    Model 9110E Instruction Manual Calibration Procedures 7.2.1. Connect Zero Air and Span Gases to the Analyzer Source of sample gas removed Calibrated NO gas during (high concentration) MODEL 700 Gas calibration Dilution Calibrator (with O generator option) Sample MODEL 701...
  • Page 132: Perform Zero/Span Calibration

    Calibration Procedures Model 9110E Instruction Manual NOTE The expected concentrations for both NO and NO are usually set to the same value unless the conversion efficiency is not equal to 1.000 or not entered properly in the conversion efficiency setting. When setting expected concentration values, consider impurities in your span gas source (NO often contains 1-3% NO and vice versa).
  • Page 133: Calibration Checks

    Model 9110E Instruction Manual Calibration Procedures If the ZERO or SPAN keys are not displayed during zero or span calibration, the measured concentration value is too different from the expected value and the analyzer does not allow to zero or span the instrument. Consult Section 11.3 for calibration problems.
  • Page 134: Calibration With Izs Option

    Calibration Procedures Model 9110E Instruction Manual Connect the sources of zero air and span gas to the respective ports on the rear panel (Figure 3-2) to follow the setup in Figure 7-4. Set the expected NO and NO values according to the procedure described in Section 7.2.
  • Page 135: Figure 7-3: Pneumatic Connections For Manual Calibration Checks With Izs

    Model 9110E Instruction Manual Calibration Procedures permeation tube and through the NO converter (Figure 3-11). It also programs the analyzer to use the NO output from the NO converter to calibrate the span value of both NO and NO NOTE The IZS with NO permeation tube has a limited accuracy of about ±5%.
  • Page 136 Calibration Procedures Model 9110E Instruction Manual SAMPLE RANGE = 500.0 PPB NOX= X.X Set the Display to show the NOX STB test function. < TST TST > CAL CALZ CALS SETUP SAMPLE NOX STB=XXX.X PPB NO = X.X Wait until STABIL <...
  • Page 137: Calibration Checks With Izs Or Zero/Span Valves

    Model 9110E Instruction Manual Calibration Procedures Perform the span calibration procedure: SAMPLE RANGE = 500.0 PPB NOX= X.X Set the Display to show the NOX STB test function. < TST TST > CAL CALZ CALS SETUP SAMPLE NOX STB=XXX.X PPB NOX=X.X X...
  • Page 138: Calibration With Independent Or Auto Ranges

    EXIT SAMPLE display 7.7. Calibration With Independent or AUTO Ranges Additional considerations apply when calibrating a Model 9110E with reporting ranges set to AUTO (automatic) or IND (independent) mode (Sections 6.5.4 and 6.5.5). 7.7.1. Calibration in AUTO Range Mode If the analyzer is operated in AUTO range mode, the high and low ranges must be indepen- dently calibrated.
  • Page 139: Independent Range Mode

    Model 9110E Instruction Manual Calibration Procedures SAMPLE RANGE = 500.0 PPB NOX=X.XX < TST TST > CAL CALZ CALS SETUP SAMPLE RANGE TO CAL: LOW LOW HIGH ENTR SETUP SAMPLE RANGE TO CAL: HIGH LOW HIGH ENTR SETUP ZERO CAL M RANGE = 500.0 PPB...
  • Page 140: Automatic Calibration (Autocal)

    Calibration Procedures Model 9110E Instruction Manual 7.8. Automatic Calibration (AutoCal) The AutoCal feature allows unattended, periodic operation of the zero/span valve options by using the analyzer’s internal time of day clock. The AutoCal feature is only available on the front panel menu (ACAL) if either the zero/span valve or the IZS option is installed.
  • Page 141: Table 7-4: Example Auto-Cal Sequence

    Model 9110E Instruction Manual Calibration Procedures NOTE The CALIBRATE attribute (formerly called “dynamic calibration”) must always be set to OFF for analyzers used in US EPA controlled applications that have IZS option installed. Calibration of instruments used in US EPA related applications should only be performed using external sources of zero air and span gas with an accuracy traceable to EPA or NIST standards and supplied through the analyzer’s...
  • Page 142: Calibration Quality Analysis

    Calibration Procedures Model 9110E Instruction Manual SAMPLE RANGE = 500.0 PPB NOX=X.X SETUP C.4 STARTING TIME:14:15 < TST TST > CAL CALZ CZLS SETUP <SET SET> EDIT EXIT PRIMARY SETUP MENU SETUP X.X SETUP C.4 DELTA DAYS: 1 <SET SET> EDIT...
  • Page 143: Table 7-5: Calibration Data Quality Evaluation

    Model 9110E Instruction Manual Calibration Procedures terms, both individually and relative to each other, indicate the quality of the calibration. To perform this quality evaluation, you will need to record the values of the following test functions (Section 6.2.1 or Appendix A-3), all of which are automatically stored in the iDAS channel CALDAT for data analysis, documentation and archival.
  • Page 144: User Notes

    Calibration Procedures Model 9110E Instruction Manual USER NOTES: M9110E Rev B...
  • Page 145: Epa Protocol Calibration

    Model 9110E Instruction Manual EPA Protocol Calibration 8. EPA PROTOCOL CALIBRATION To ensure high quality, accurate measurements at all times, the M9110E analyzer must be calibrated prior to use. A quality assurance program centered on this aspect and including attention to the built-in warning features of the analyzer, periodic inspection, regular zero/span checks, regular evaluation of test parameters for predictive diagnostics and data analysis and routine maintenance of the instrument are paramount to achieving this goal.
  • Page 146: Table 8-1: Activity Matrix For Epa Calibration Equipment And Supplies

    EPA Protocol Calibration Model 9110E Instruction Manual Table 8-1: Activity Matrix for EPA Calibration Equipment and Supplies Equipment/ Frequency And Method Action If Requirements Acceptance Limits Supplies Of Measurement Are Not Met Recorder Compatible with output signal Check upon receipt Return equipment to of analyzer;...
  • Page 147: Gas Phase Titration (Gpt)

    Model 9110E Instruction Manual EPA Protocol Calibration 8.2. Gas Phase Titration (GPT) 8.2.1. GPT Principle of Operation Gas phase titration (GPT) is recommended for calibration of the M9110E. Those using a NO permeation tube should refer to the CFR The principle of GPT is based on the rapid gas phase reaction between NO and O...
  • Page 148: Example Gpt Calculation

    EPA Protocol Calibration Model 9110E Instruction Manual If more than one analyzer is to be calibrated at the same time, multiply F by the number of analyzers. Calculate the NO concentration [NO] needed to approximate 90% of the URL of the NO...
  • Page 149 Model 9110E Instruction Manual EPA Protocol Calibration assumed that only the volume of the reaction cell, V , and the concentration of the NO standard, [NO] , are known. All flow settings (F , and F ) will be calculated. In many uses, these flow settings are known and need only to be substituted in Equations 8-5 and 8-6 to verify the required conditions.
  • Page 150: Multipoint Calibration Procedure

    EPA Protocol Calibration Model 9110E Instruction Manual Calculate the diluent air flow (F ) required at the mixing chamber, using Equation 8-7: = 2750 cm /min - 260 cm /min - 24.5 cm /min = 2465.5 cm /min 8.3. Multipoint Calibration Procedure...
  • Page 151: Zero Calibration

    Model 9110E Instruction Manual EPA Protocol Calibration Figure 8-1: GPT Calibration System Converter efficiency should be set prior to calibration since its value is used in the computation of the NO and NO concentration outputs. The analyzer should be calibrated on the same range used for monitoring.
  • Page 152: Gpt No Calibration

    EPA Protocol Calibration Model 9110E Instruction Manual Eq 8-8 Enter the respective concentrations using the procedure in Section 7.2. The expected span concentrations need not be re-entered each time a calibration is performed unless they are changed. • Enter the expected NO and NO span gas concentrations: •...
  • Page 153: Calibration Frequency

    Model 9110E Instruction Manual EPA Protocol Calibration CONVERTER EFFICIENCY if the NO reading should drop to less than 96% of its starting value during this step, it indicates the NO converter is in need of troubleshooting or replacement. See Chapter 11 for further details.
  • Page 154: Summary Of Quality Assurance Checks

    EPA Protocol Calibration Model 9110E Instruction Manual Table 8-3: Definition of Level 1 and Level 2 Zero and Span Checks LEVEL 1 ZERO AND SPAN CALIBRATION LEVEL 2 ZERO AND SPAN CHECK A Level 1 zero and span calibration is a A Level 2 zero and span check is an "unofficial"...
  • Page 155: Short Calibration Checks

    Model 9110E Instruction Manual EPA Protocol Calibration Table 8-4: Activity Matrix for Data Quality Frequency and Method Action if Requirements Characteristic Acceptance Limits of Measurement are not Met Shelter Mean temperature between Check thermograph chart Mark strip chart for the temperature 22 °...
  • Page 156: Precision Check

    EPA Protocol Calibration Model 9110E Instruction Manual • Automatic zero/span checks can be performed every night. See Section 7.8 of this manual for setup and operation procedures. • Zero/Span checks through remote contact closure can be initiated through remote contact closures on the rear panel. See Section 7.7.3 of this manual.
  • Page 157: Certification Procedures Of Working Standards

    Model 9110E Instruction Manual EPA Protocol Calibration To establish traceability, data must be obtained routinely by repeated measurements of standard reference samples (primary, secondary, and/or working standards). More specifically, working calibration standards must be traceable to standards of higher accuracy.
  • Page 158 EPA Protocol Calibration Model 9110E Instruction Manual Register, 41 (232), 52688-52692, December 1976 (as amended at 48 FR 2529, Jan 20, 1983). 4. Mavrodineanu, R., and Gills, T. E., Standard Reference Materials: Summary of Gas Cylinder and Permeation Tube Standard Reference Materials Issued by the National Bureau of Standards, Document SP260-108, May 1987.
  • Page 159: Instrument Maintenance

    Model 9110E Instruction Manual Instrument Maintenance 9. INSTRUMENT MAINTENANCE Predictive diagnostic functions including data acquisition, failure warnings and alarms built into the analyzer allow the user to determine when repairs are necessary without perform- ing unnecessary, preventative maintenance procedures. There is, however, a minimal number of simple procedures that, when performed regularly, will ensure that the analyzer continues to operate accurately and reliably over its lifetime.
  • Page 160 Instrument Maintenance Model 9110E Instruction Manual Item Action Frequency Manual Date Performed Check ? Section converter Check Every 3 Yes on 9.3.8 conversion months change, efficiency else no External zero Exchange Every 3 9.3.7 air scrubber chemical months (optional) External dryer...
  • Page 161: Predictive Diagnostics

    Drift of instrument response; clean RCEL NO CONC constant time window, change O air filter chemical. concentration 9.3. Maintenance Procedures The following procedures need to be performed regularly as part of the standard maintenance of the Model 9110E. M9110E Rev B...
  • Page 162: Changing The Sample Particulate Filter

    Instrument Maintenance Model 9110E Instruction Manual 9.3.1. Changing the Sample Particulate Filter The particulate filter should be inspected often for signs of plugging or excess dirt. It should be replaced according to the service interval in Table 9-1 even without obvious signs of dirt.
  • Page 163: Changing The O

    Model 9110E Instruction Manual Instrument Maintenance • To fulfill CE Mark safety requirements, the front panel locking screw must be installed at all times during operation of the analyzer. • Re-start the analyzer. 9.3.2. Changing the O Dryer Particulate Filter ©...
  • Page 164: Changing The Ozone Filter Chemical

    Appendix B of this manual lists the part numbers of the ozone filter rebuild kit. Some older models 9110E contain a Teflon cartridge, which is shown in the inset. Use a special tool FT267 for the red fittings of those cartridges, otherwise follow these instructions: •...
  • Page 165: Rebuilding The External Sample Pump

    Model 9110E Instruction Manual Instrument Maintenance • Pour the contents of the cartridge onto a sheet of white paper. Notice any discoloration of the contents, which is usually white and slightly transparent. Note how much of the chemical is discolored (usually with yellow tint), this may give you an indication of the lifetime of the chemical in your application.
  • Page 166: Changing The Pump And Izs Dust Filters

    Instrument Maintenance Model 9110E Instruction Manual locations). A pump rebuild kit is available from the factory. Appendix B of this manual lists the part numbers of the pump rebuild kit. Instructions and diagrams are included in the kit. A flow and leak check after rebuilding the sample pump is recommended. A span check and re-calibration after this procedure is necessary as the response of the analyzer changes with the RCEL pressure.
  • Page 167: Changing The External Zero Air Scrubber

    Model 9110E Instruction Manual Instrument Maintenance • Unscrew the black aluminum cover of the IZS oven (3 screws) using a medium Phillips-head screw driver. Leave the fittings and tubing connected to the cover. • Remove the old permeation tube if necessary and replace it with the new tube. Make sure that the tube is placed into the larger of two holes and that the open permeation end of the tube (plastic) is facing up.
  • Page 168: Changing The No Converter

    Instrument Maintenance Model 9110E Instruction Manual Figure 9-4: Zero Air Scrubber Assembly 9.3.8. Changing the NO converter The NO converter is located in the center of the instrument, see Figure 3-2 for location, and Figure 9-5 for the assembly. The converter is designed for replacement of the cartridge only, the heater with built-in thermocouple can be reused.
  • Page 169: Cleaning The Reaction Cell

    Model 9110E Instruction Manual Instrument Maintenance • Remove the converter assembly (cartridge and band heater) from the can. Make a note of the orientation of the tubes relative to the heater cartridge. • Unscrew the band heater and loosen it, take out the old converter cartridge.
  • Page 170: Figure 9-6: Reaction Cell Assembly

    Instrument Maintenance Model 9110E Instruction Manual • Disconnect the black 1/4" exhaust tube and the 1/8” sample and ozone air tubes from the reaction cell. Disconnect the heater/thermistor cable. • Remove four screws holding the reaction cell to the PMT housing and lift the cell and manifold out as shown in the inset of Figure 9-6.
  • Page 171: Cleaning Or Changing Critical Flow Orifices

    Model 9110E Instruction Manual Instrument Maintenance 9.3.10. Cleaning or Changing Critical Flow Orifices There are several critical flow orifices installed in the M9110E, Figure 9-6 shows one of the two most important orifice assemblies, located on the reaction cell. Refer to Section 10.2.9 for a detailed description on functionality and locations.
  • Page 172: Checking For Light Leaks

    Instrument Maintenance Model 9110E Instruction Manual Figure 9-7: Critical Flow Orifice Assembly 9.3.11. Checking for Light Leaks When re-assembled or operated improperly, the M9110E can develop small leaks around the PMT, which let stray light from the analyzer surrounding into the PMT housing. To find such light leaks, follow the below procedures.
  • Page 173: Theory Of Operation

    Model 9110E Instruction Manual Theory of Operation 10. THEORY OF OPERATION The M9110E Nitrogen Oxides Analyzer is a microprocessor controlled instrument that determines the concentration of nitric oxide (NO), total nitrogen oxides (NO , the sum of NO and NO ) and nitrogen dioxide (NO ) in a sample gas drawn through the instrument.
  • Page 174: Figure 10-1: M9110E Sensitivity Spectrum

    Theory of Operation Model 9110E Instruction Manual Figure 10-1: M9110E Sensitivity Spectrum However, only about 20% of the NO that is formed through reaction 10-1 is in the excited state. In addition, the excited NO can collide with another collision partner M in the reaction cell (mostly other molecules but also cell walls) and transfer its excess energy to its collision partner without emitting any light at all (Equation 10-3).
  • Page 175: Chemiluminescence Detection

    Model 9110E Instruction Manual Theory of Operation → ° Eq 10-4 Once the NO in the sample gas has been converted to NO, it is routed to the reaction cell where it undergoes the chemiluminescence reaction described in Equations 10-1 and 10-2.
  • Page 176: Optical Filter

    Theory of Operation Model 9110E Instruction Manual Figure 10-3: Reaction Cell with PMT Tube 10.1.3.2. Optical Filter Another critical component in the method by which your M9110E detects chemiluminescence is the optical filter that lies between the reaction cell and the PMT (Figure 10-2).
  • Page 177: Measurement Interferences

    Model 9110E Instruction Manual Theory of Operation Figure 10-4: Reaction Cell During the AutoZero Cycle. 10.1.5. Measurement Interferences It should be noted that the chemiluminescence method is subject to interferences from a number of sources. The M9110E has been successfully tested for its ability to reject interference from most of these sources.
  • Page 178: Table 10-1: List Of Interferents

    Theory of Operation Model 9110E Instruction Manual in ambient air measurements, hence provide a constant amount of quenching and the interference of varying CO amounts is negligible at low concentrations. Table 10-1: List of Interferents Interference Type Rejection Method Dilution: Viscosity of CO...
  • Page 179: Light Leaks

    Model 9110E Instruction Manual Theory of Operation 10.1.5.3. Light Leaks The M9110E sensitivity curve includes a small portion of the visible light spectrum (Figure 10-3), hence, it is important to make sure than the reaction cell is completely sealed with respect to light. To ensure this, all pneumatic tubing leading into the reaction cell is either opaque (vacuum exit tubing) in order to prevent light from entering the cell or light penetration is prevented by stainless steel filters and orifices (gas entries).
  • Page 180: Sample Gas Flow

    Theory of Operation Model 9110E Instruction Manual Figure 10-5: External Pump Pack Finally, the M9110E requires a steady, high underpressure, which cannot be achieved reliably over extended periods of time with small vacuum pumps. The external pump used for the M9110E has a very long lifetime and duty cycle and provides a very good vacuum for its entire lifetime.
  • Page 181: Sample Particulate Filter

    Model 9110E Instruction Manual Theory of Operation Table 10-2: M9110E Valve Cycle Phases Phase NO/ NO Autozero Time Activity Figure Valve Valve Index Status Status Open to Open to 0 - 2 s Wait period (NO dwell time). Figure 10-2...
  • Page 182: Ozone Gas Air Flow

    Theory of Operation Model 9110E Instruction Manual 10.2.4. Ozone Gas Air Flow The excess ozone needed for reaction with NO in the reaction cell is generated inside the analyzer because of the instability and toxicity of ozone. Besides the ozone generator itself, this requires a dry air supply and filtering of the gas before it is introduced into the reaction cell.
  • Page 183: Perma Pure Dryer

    Model 9110E Instruction Manual Theory of Operation The M9110E utilizes a dual-dielectric design. This method utilizes a glass tube with hollow walls. The outermost and innermost surfaces are coated with electrically conductive material. The air flows through the glass tube, between the two conductive coatings, in effect creating a capacitor with the air and glass acting as the dielectric.
  • Page 184: Ozone Supply Air Filter

    Theory of Operation Model 9110E Instruction Manual on hydrogen bonds between the water molecule and the Nafion material. Other small polar gases that are capable of hydrogen bonds can be absorbed this way, too, such as ammonia ) and some low molecular amines. The gases of interest, NO and NO , do not get absorbed and pass the dryer unaltered.
  • Page 185: Ozone Scrubber

    Model 9110E Instruction Manual Theory of Operation stream, the output of the O generator flows through a special filter between the generator and the reaction cell. Any NO that may be produced in the generator (from reaction of O or O...
  • Page 186: Figure 10-9: Flow Control Assembly & Critical Flow Orifice

    Theory of Operation Model 9110E Instruction Manual As the pressure on the downstream side of the orifice (the pump side) continues to drop, the speed of the gas flowing through the orifice continues to rise. Once the ratio of upstream pressure to downstream pressure is greater than 2:1, the velocity of the gas through the orifice reaches the speed of sound and remains constant, regardless of any further pressure difference.
  • Page 187: Pneumatic Sensors

    Model 9110E Instruction Manual Theory of Operation sample and O flows. Table 10-3 shows the flow rates for each of the critical flow orifices of the M9110E. See Appendix B for spare part numbers of these orifices. In addition to controlling the gas flows, the two critical flow orifices at the inlets of the...
  • Page 188: Sample Pressure Sensor

    Theory of Operation Model 9110E Instruction Manual Figure 10-10: Vacuum Manifold, Standard Configuration 10.2.10.2. Sample Pressure Sensor An absolute pressure transducer connected to the input of the NO/NO valve is used to measure the pressure of the sample gas before it enters the analyzer’s reaction cell. This is the “upstream”...
  • Page 189: Dilution Manifold

    Model 9110E Instruction Manual Theory of Operation cm³/min), a warning message OZONE FLOW WARNING is displayed on the analyzer’s front panel (Section 6.2.2) and the O generator is turned off. As second warning, OZONE GEN OFF, is displayed. This flow measurement is viewable through instrument’s front panel display as the test function OZONE FL.
  • Page 190: Cpu

    Theory of Operation Model 9110E Instruction Manual of peripheral devices through the motherboard, the main printed circuit assembly on the rear panel (Figure 3-2). Figure 10-12: M9110E Electronic Block Diagram 10.3.1. CPU The CPU is a low power (5 VDC, 0.8A max), high performance, 386-based microcomputer running a version of the DOS operating system.
  • Page 191: Disk On Chip

    Model 9110E Instruction Manual Theory of Operation access (DMA) devices over data busses of either 8-bit or 16-bit bandwidth. The CPU supports both RS-232 and RS-485 serial protocols. Figure 10-13 shows the CPU board. • The CPU communicates with the user and the outside world in a variety of ways: •...
  • Page 192: Flash Chip

    Theory of Operation Model 9110E Instruction Manual 10.3.1.2. Flash Chip The flash chip is another, smaller EEPROM with about 64 kb of space, internally labeled as DOS drive B:\. The M9110E CPU board can accommodate up to two EEPROM flash chips.
  • Page 193: Pmt Cooling System

    Model 9110E Instruction Manual Theory of Operation Figure 10-14: Schematic of Basic PMT Design and Functionality A significant performance characteristic of the PMT is the voltage potential across the electron multiplier. The higher the voltage, the greater is the number of electrons emitted from each dynode of the electron multiplier, making the PMT more sensitive and responsive to small variations in light intensity but also increases random noise (dark noise).
  • Page 194: Pmt Preamplifier

    Theory of Operation Model 9110E Instruction Manual A red LED located on the top edge of this circuit board indicates that the control circuit is receiving power. Four test points are also located at the top of this assembly. For the definitions and acceptable signal levels of these test points see Chapter 11.
  • Page 195: Pneumatic Sensor Board

    Model 9110E Instruction Manual Theory of Operation Figure 10-15: PMT Preamp Block Diagram 10.3.3. Pneumatic Sensor Board The flow and pressure sensors of the M9110E are located on a printed circuit assembly just behind the PMT sensor. Refer to Section 11.5.15 for a figure and on how to test this assembly.
  • Page 196: Relay Board

    Theory of Operation Model 9110E Instruction Manual processing. All sensors are linearized in the firmware and can be span calibrated from the front panel. 10.3.4. Relay Board The relay board is the central switching and power distribution unit of the analyzer. It...
  • Page 197: Motherboard

    Model 9110E Instruction Manual Theory of Operation for 30 seconds, the watchdog circuit will assume lost connection and automatically turn of all valves and heaters and may cause the analyzer to reboot or the I Z bus to hang up.
  • Page 198: Thermistor Interface

    Theory of Operation Model 9110E Instruction Manual • The NO converter temperature is measured with a Type-K thermocouple attached to the NO converter heater and its analog signal is amplified by the circuitry on the relay board. It is sent to the CPU and then digitized and is used to calculate the current temperature of the NO converter.
  • Page 199: External Digital I/O

    Model 9110E Instruction Manual Theory of Operation The output labeled A4 can be set by the user (Section 6.7.2) to carry the signal of any of these TEST parameters: none , PMT DETECTOR, OZONE FLOW, SAMPLE FLOW, SAMPLE PRESSURE, RCELL PRESSURE, RCELL TEMP, MANIFOLD TEMP, IZS TEMP, CONV TEMP, PMT TEMP, BOX TEMP, HVPS VOLTAGE.
  • Page 200: Communications Interface

    Theory of Operation Model 9110E Instruction Manual right corner of the front panel and back to the relay board, which carries the switching power supplies. AC line power is stepped down and converted to DC power by two switching power supplies, one for +12 VDC (5 A) for various valves and the TEC, and a second supply that provides +5 VDC (3 A) and ±15 VDC (1.5/0.5 A) for logic and analog circuitry as well as the ozone...
  • Page 201: Front Panel

    Model 9110E Instruction Manual Theory of Operation through its external, digital I/O connectors and through the four analog outputs, all located on the rear panel. Figure 10-17: M9110E Electronic Interface Block Diagram 10.3.7.1. Front Panel The front panel of the analyzer is hinged at the bottom and may be opened to gain access to various components mounted on the panel itself or located near the front of the instrument (such as the particulate filter).
  • Page 202: Keypad

    Theory of Operation Model 9110E Instruction Manual ment calibration or certain diagnostic procedures, the instrument’s response messages are also displayed here. Concentration Field: The far right portion of the top line of text displays the concentration of the sample gas currently being measured by the analyzer. The number reported here is the actual concentration of the sample gas reported in whatever units the user selects (Section 6.5.6).
  • Page 203: Software Operation

    Model 9110E Instruction Manual Theory of Operation Figure 10-18: Keyboard and Display Interface Block Diagram 10.4. Software Operation The M9110E NO analyzer’s core module is a high performance, 386-based microcomputer running a version of DOS. On top of the DOS shell, special software developed by TAI interprets user commands from various interfaces, performs procedures and tasks, stores data in the CPU’s memory devices and calculates the concentrations of NO...
  • Page 204: Calibration - Slope And Offset

    Theory of Operation Model 9110E Instruction Manual must be simultaneously met to switch to the short filter. First, the instantaneous concentra- tion must differ from the average in the long filter by at least 50 ppb. Second, the instanta- neous concentration must differ from the average in the long filter by at least 10% of the average in the long filter.
  • Page 205: Temperature/Pressure Compensation (Tpc)

    Model 9110E Instruction Manual Theory of Operation 10.4.3. Temperature/Pressure Compensation (TPC) The software features a compensation of some temperature and pressure changes critical in the measurement of NO and NO concentration. When the TPC feature is enabled (default setting), the analyzer divides the value of the PMT output signal (PMTDET) by a value called TP_FACTOR.
  • Page 206: Converter Efficiency Compensation

    Theory of Operation Model 9110E Instruction Manual 10.4.4. NO Converter Efficiency Compensation Over time, the molybdenum in the NO converter oxidizes and looses its original capacity of converting NO into NO, eventually resulting in a decreased converter efficiency (CE). Even though we recommend to replace the converter if CE drops below 96%, the analyzer’s...
  • Page 207: Troubleshooting & Repair

    Model 9110E Instruction Manual Troubleshooting & Repair 11. TROUBLESHOOTING & REPAIR This section contains a variety of methods for identifying and solving performance problems with the analyzer. NOTE The operations outlined in this chapter must be performed by qualified maintenance personnel only.
  • Page 208: Warning Messages

    Troubleshooting & Repair Model 9110E Instruction Manual 11.1.1. Warning Messages The most common and/or serious instrument failures will result in a warning message displayed on the front panel. Table A-2 in Appendix A-3 contains a list of warning messages, along with their meaning and recommended corrective action.
  • Page 209: Using The Diagnostic Signal I/O Function

    Model 9110E Instruction Manual Troubleshooting & Repair measurements recorded on the factory data sheet may also indicate a failure or a maintenance item. A problem report worksheet has been provided in Appendix C (TAI part number 04503) to assist in recording the value of these test functions. The following table contains some of the more common causes for these values to be out of range.
  • Page 210: Status Led's

    Troubleshooting & Repair Model 9110E Instruction Manual voltage of an input signal or to control the state of an output voltage or control signal. The specific parameter will vary depending on the situation. SETUP ENTER DIAG PASS: 818 SAMPLE RANGE = 500.0 PPB Press ENTR to NOX=XXX.X...
  • Page 211: Cpu Status Indicator

    Model 9110E Instruction Manual Troubleshooting & Repair Figure 11-3: Motherboard Watchdog Status Indicator 11.1.4.2. CPU Status Indicator The CPU board has two red LEDs, the lower of which is the watchdog timer (the device that pulses the motherboard watchdog). This LED is labeled LED2 and blinks about twice per second (twice as fast as the motherboard LED) when operating normally.
  • Page 212: Figure 11-4: Relay Board Pca

    Troubleshooting & Repair Model 9110E Instruction Manual Figure 11-4: Relay Board PCA To enter the signal I/O test mode to manually control I/O functions such as valves and heaters, press the following keys while observing the relay board LEDs when toggling: SAMPLE RANGE = 500.0 PPB...
  • Page 213: Gas Flow Problems

    Model 9110E Instruction Manual Troubleshooting & Repair Table 11-2: Relay Board Status LEDs Color Function Fault Status Indicated Failure(s) LED Row 1 (center of board) Watchdog Circuit; I C bus Continuously Failed or halted CPU; faulty operation. ON or OFF motherboard, keyboard, relay board;...
  • Page 214: Zero Or Low Flow Problems

    Troubleshooting & Repair Model 9110E Instruction Manual When troubleshooting flow problems, it is essential to confirm the actual flow rate without relying on the analyzer’s flow display. The use of an independent, external flow meter to perform a flow check as described in Section 6.7.8 is essential.
  • Page 215: Ozone Flow Is Zero Or Low

    Model 9110E Instruction Manual Troubleshooting & Repair • If the sample and vacuum pressures are around their nominal values (28 and <10 in-Hg-A, respectively) and the flow still displays XXXX, carry out a leak check as described in Section 11.5.2 •...
  • Page 216: High Flow

    Troubleshooting & Repair Model 9110E Instruction Manual 11.2.2. High Flow Flows that are significantly higher than the allowed operating range (typically ±10-11% of the nominal flow) should not occur in the M9110E unless a pressurized sample, zero or span gas is supplied to the inlet ports. Ensure to vent excess pressure and flow just before the analyzer inlet ports.
  • Page 217: No Response

    Model 9110E Instruction Manual Troubleshooting & Repair high, positive values. The M9110E has a built-in Auto-zero function, which should take care of most of these deviations from zero, but may yield a small, residual, negative value. If larger, negative values persist continuously, check if the Auto-zero function was accidentally turned off using the remote variables in Appendix A-2.
  • Page 218: Unstable Zero And Span

    Troubleshooting & Repair Model 9110E Instruction Manual • Carry out an optical test using the OPTIC TEST procedure in the diagnostics menu, see Section 6.7.5. If this test results in a concentration signal, then the PMT sensor and the electronic signal path are operating properly. If the M9110E passes both ETEST and OTEST, the instrument is capable of detecting light and processing the signal to produce a reading.
  • Page 219: Inability To Zero - No Zero Key

    Model 9110E Instruction Manual Troubleshooting & Repair 11.3.5. Inability to Zero - No ZERO Key In general, the M9110E will not display certain keyboard choices whenever the actual value of a parameter is outside of the expected range for that parameter. If the calibration menu...
  • Page 220: Discrepancy Between Analog Output And Display

    Troubleshooting & Repair Model 9110E Instruction Manual between high concentrations span gases and the dilution system, oxygen from ambient air can diffuse into the line and react with NO to form NO . This reaction is dependent on NO concentration and accelerates with increasing NO concentration, hence, affects linearity only at high NO levels.
  • Page 221: Slow Response

    Model 9110E Instruction Manual Troubleshooting & Repair Another possibility of excessive signal noise may be the preamplifier board, the high voltage power supply and/or the PMT detector itself. Contact the factory on trouble-shooting these components. 11.4.2. Slow Response If the analyzer starts responding too slow to any changes in sample, zero or span gas, check for the following: •...
  • Page 222: Subsystem Checkout

    Troubleshooting & Repair Model 9110E Instruction Manual to be replaced. This filter cartridge (Figure 3-2) filters chemicals that can cause chemiluminescence and, if saturated, these chemicals can break through to the reaction cell, causing an erroneously high AutoZero value (background noise).
  • Page 223: Performing A Sample Flow Check

    Model 9110E Instruction Manual Troubleshooting & Repair • Install a leak checker or a tank of gas (compressed, oil-free air or nitrogen) as described above on the sample inlet at the rear panel. • Disconnect the pump tubing on the outside rear panel and cap the pump port. If IZS or zero/span valves are installed, disconnect the tubing from the zero and span gas ports and plug them (Figure 3-3).
  • Page 224: Ac Power Configuration

    Troubleshooting & Repair Model 9110E Instruction Manual 11.5.4. AC Power Configuration The M9110E can be easily configured for two main power regimes, 100-120 V and 220-240 V at either 50 or 60 Hz. The analyzer is correctly configured for the AC power voltage in use if the voltage selector plug is configured properly.
  • Page 225: I 2 C Bus

    Model 9110E Instruction Manual Troubleshooting & Repair Table 11-4: DC Power Supply Acceptable Levels CHECK relay board Test Points Power Voltage From Test Point To Test Point Min V Max V Supply Name Name DGND +4.80 +5.25 AGND +13.5 +16.0...
  • Page 226: Relay Board

    Troubleshooting & Repair Model 9110E Instruction Manual 11.5.8. Relay Board The relay board circuit can most easily be checked by observing the condition of its status LEDs as described in Section 11.1.4.3, and the associated output when toggled on and off through the SIGNAL I/O function in the DIAG menu, see Section 11.1.3.
  • Page 227: Status Outputs

    Model 9110E Instruction Manual Troubleshooting & Repair Table 11-6: Analog Output Test Function - Nominal Values Full scale Output Voltage 100mV Step Nominal Output Voltage 0 mV 20 mV 40 mV 60 mV 80 mV 100 mV 11.5.9.3. Status Outputs The procedure below can be used to test the Status outputs.
  • Page 228: Cpu

    Troubleshooting & Repair Model 9110E Instruction Manual • Connect a second jumper from the ‘-‘ pin on the STATUS connector to the A pin on the CONTROL IN connector. The instrument should switch from SAMPLE mode to ZERO CAL R mode.
  • Page 229: Modem Or Terminal Operation

    Model 9110E Instruction Manual Troubleshooting & Repair the green LED should also illuminate. If not, set the DCE/DTE switch to the other position. See also Section 6.9.5. • Note that some laptops do not enable their RS-232 port when in power-saving mode.
  • Page 230: High Voltage Power Supply

    Troubleshooting & Repair Model 9110E Instruction Manual 11.5.14. High Voltage Power Supply The HVPS is located in the interior of the sensor module and is plugged into the PMT tube (Section 10.3.2). It requires 2 voltage inputs. The first is +15 V, which powers the supply.
  • Page 231: Reaction Cell Pressure

    Model 9110E Instruction Manual Troubleshooting & Repair 11.5.15.1. Reaction Cell Pressure Measure the voltage across test points TP1 and TP5. With the sample pump disconnected or turned off, the voltage should be 4500 ± 250 mV. With the pump running, it should be 800- 1700 mV depending on the performance of the vacuum pump.
  • Page 232: Ozone Flow

    Troubleshooting & Repair Model 9110E Instruction Manual 11.5.15.3. Ozone Flow Measure the voltage across TP1 and TP3. With proper ozone flow (80 cm³/min at the ozone dryer inlet), this should be approximately 2.0 ± 0.2 V (this voltage will vary with altitude).
  • Page 233: O 3 Generator

    Model 9110E Instruction Manual Troubleshooting & Repair cumulative product of the NO concentration times the exposure time to that concentration). However, this lifetime heavily depends on many factors such as absolute concentration (temporary or permanent poisoning of the converter is possible), sample flow rate and pressure inside the converter, converter tempera- ture, duty cycle etc.
  • Page 234: Box Temperature

    Troubleshooting & Repair Model 9110E Instruction Manual The semi-permeable PTFE membrane of the permeation tube is severely affected by humidity. Variations in humidity between day and night are usually enough to yield very variable output results. If the instrument is installed in an air-conditioned shelter, the air is usually dry enough to produce good results.
  • Page 235: Disk-On-Chip Replacement

    Model 9110E Instruction Manual Troubleshooting & Repair 11.6.1. Disk-on-Chip Replacement Replacing the Disk-on-Chip (DOC) will cause all of the instrument configuration parameters to be lost unless the replacement chip carries the exact same firmware version. iDAS data will always be lost and, if possible, should be downloaded prior to changing the DOC. If the analyzer is equipped with at least one EEPROM flash chip (standard configuration), the configuration settings are stored on the EEPROM.
  • Page 236: O 3 Generator Replacement

    Troubleshooting & Repair Model 9110E Instruction Manual • When looking at the electronic circuits from the back of the analyzer, locate the EEPROM chip in the left-most socket of the CPU board. The chip is almost square with one corner cut off, the socket is shaped accordingly and the chip is recessed into the socket.
  • Page 237: Pmt Sensor Hardware Calibration

    Model 9110E Instruction Manual Troubleshooting & Repair • Attach the replacement dryer to the mounting bracket in the same orientation as the old dryer. Fix the dryer to the bracket using new tie wraps. Cut off excess length of the wraps.
  • Page 238: Figure 11-6: Pre-Amplifier Board Layout

    Troubleshooting & Repair Model 9110E Instruction Manual Figure 11-6: Pre-Amplifier Board Layout • Adjust the HVPS fine adjustment such that the NORM PMT value is 800-830 mV. The fine adjustment typically increments the NORM PMT value by about 30 mV. It may be necessary to go back and forth between coarse and fine adjustments if the proper value is at the threshold of the min/max coarse setting.
  • Page 239: Replacing The Pmt, Hvps Or Tec

    In case of PMT, HVPS or TEC failure, the sensor assembly needs to be opened in order to change one of these components. Refer to Figure 11-7 for the structure of the 9110E sensor assembly and follow the steps below for replacement of one of its components. We recommend to ensure that the PMT, HVPS or TEC modules are, indeed, faulty to prevent unnecessary opening of the sensor.
  • Page 240: Technical Assistance

    11.7. Technical Assistance If this manual and its trouble-shooting / repair sections do not solve your problems, technical assistance may be obtained from Teledyne Analytical Instruments, Customer Service, 16830 Chestnut St. City of Industry, CA 91748. Phone: 626-934-1673. Fax: 626-...
  • Page 241 Model 9110E Instruction Manual Troubleshooting & Repair 961-2538. Email: ask_tai@teledyne.com. Before you contact customer service, fill out the problem report form in Appendix C M9110E Rev B...
  • Page 242: User Notes

    Troubleshooting & Repair Model 9110E Instruction Manual USER NOTES: M9110E Rev B...
  • Page 243: Appendix A: Software Documentation

    Model 9110E Instruction Manual APPENDIX A APPENDIX A: Software Documentation APPENDIX A: SOFTWARE DOCUMENTATION............229 APPENDIX A-1: M9110E Software Menu Trees and Index, Revision C.8 ......230 APPENDIX A-2: Setup Variables For Serial I/O, Revision C.8..........236 APPENDIX A-3: Warnings and Test Measurements, Revision C.8 ........245 APPENDIX A-4: M9110E Signal I/O Definitions, Revision C.8..........
  • Page 244: Figure A-1: Basic Sample Display Menu

    APPENDIX A Model 9110E Instruction Manual APPENDIX A-1: M9110E Software Menu Trees and Index, Revision C.8 Figure A-1: Basic Sample Display Menu 044100120 Rev B...
  • Page 245: Figure A-2: Sample Display Menu - Units With Z/S Valve Or Izs Option Installed

    Model 9110E Instruction Manual APPENDIX A Figure A-2: Sample Display Menu - Units with Z/S Valve or IZS Option installed 044100110 Rev B...
  • Page 246: Figure A-3: Primary Setup Menu (Except Idas)

    APPENDIX A Model 9110E Instruction Manual Figure A-3: Primary Setup Menu (Except iDAS) 044100120 Rev B...
  • Page 247: Figure A-4: Secondary Setup Menu (Comm & Vars)

    Model 9110E Instruction Manual APPENDIX A Figure A-4: Secondary Setup Menu (COMM & VARS) 044100110 Rev B...
  • Page 248: Figure A-5: Secondary Setup Menu (Diag)

    APPENDIX A Model 9110E Instruction Manual Figure A-5: Secondary Setup Menu (DIAG) 044100120 Rev B...
  • Page 249: Figure A-6: Internal Data Acquisition (Idas) Menu

    Model 9110E Instruction Manual APPENDIX A Figure A-6: Internal Data Acquisition (iDAS) Menu 044100110 Rev B...
  • Page 250: Appendix

    APPENDIX A Model 9110E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision C.8 Table A-1: M9110E Setup Variables, Revision C.8 Please note that variables listed below under “Setup Variables Available Through serial I/O” should not be changed unless specifically instructed by Teledyne-API customer service or engineering! Changing those variables may cause significant problems with analyzer performance.
  • Page 251 Model 9110E Instruction Manual APPENDIX A Setup Variable Numeric Default Value Description Units Value Range SPRESS_TPC_GAIN — 0-10 Sample pressure compensation attenuation factor. CE_FACTOR1 — 0.8-1.2 Moly converter efficiency factor for range 1. CE_FACTOR2 — 0.8-1.2 Moly converter efficiency factor for range 2.
  • Page 252 APPENDIX A Model 9110E Instruction Manual Setup Variable Numeric Default Value Description Units Value Range SG_NO_DWELL Seconds 0.1–30 Dwell time after switching valve to NO position in single-gas measure modes. NO_SAMPLE Samples 1-30 Number of samples to take in NO mode.
  • Page 253 Model 9110E Instruction Manual APPENDIX A Setup Variable Numeric Default Value Description Units Value Range NO2_SPAN2 Conc. 4-20000 Target NO concentration during converter efficiency calibration of range 2. NOX_SLOPE2 PPB/mV 0.25-4 slope for range 2. PPM/mV NOX_OFFSET2 -10000- offset for range 2.
  • Page 254 APPENDIX A Model 9110E Instruction Manual Setup Variable Numeric Default Value Description Units Value Range RCELL_SAMP_RATIO — 0.53 0.1-2 Maximum reaction cell pressure / sample pressure ratio for valid sample flow calculation. STD_BOX_TEMP ºK 1-500 Standard box temperature and Valid limits:...
  • Page 255 Model 9110E Instruction Manual APPENDIX A Setup Variable Numeric Default Value Description Units Value Range BAUD_RATE2 — 115200 300, RS-232 COM2 baud rate. Enclose 1200, value in double quotes (") when 2400, setting from the RS-232 interface. 4800, 9600, 19200,...
  • Page 256 APPENDIX A Model 9110E Instruction Manual Setup Variable Numeric Default Value Description Units Value Range TEST_CHAN_ID — NONE NONE, Diagnostic analog output ID. PMT DE- Enclose value in double quotes (") TECTOR, when setting from the RS-232 OZONE interface. FLOW,...
  • Page 257 Model 9110E Instruction Manual APPENDIX A Setup Variable Numeric Default Value Description Units Value Range RCELL_PRESS_SET In-Hg 0-100 Reaction cell pressure warning Warnings: limits. Set point is not used. 0-15 RCELL_CYCLE Seconds 0.5-30 Reaction cell temperature control cycle period. RCELL_PROP 1/ºC...
  • Page 258 APPENDIX A Model 9110E Instruction Manual Setup Variable Numeric Default Value Description Units Value Range CLOCK_FORMAT — “TIME=%H: Time-of-day clock format flags. %M:%S” character Enclose value in double quotes (") in the when setting from the RS-232 allowed interface. character “%a”...
  • Page 259: Appendix

    Model 9110E Instruction Manual APPENDIX A APPENDIX A-3: Warnings and Test Measurements, Revision C.8 Table A-2: M9110E Warning Messages, Revision C.8 Name Message Text Description WSYSRES SYSTEM RESET Instrument was power-cycled or the CPU was reset. WDATAINIT DATA INITIALIZED Data storage was erased.
  • Page 260 APPENDIX A Model 9110E Instruction Manual Table A-3: M9110E Test Measurements, Revision C.8 TEST Message Text Valid Data Range Description Function RANGE RANGE=500.0 PPB 50-20000 ppb Analog output range in single or auto- range modes. RANGE1 RANGE1=500.0 PPB 50-20000 ppb D/A #1 range in independent range mode.
  • Page 261: Appendix

    Model 9110E Instruction Manual APPENDIX A APPENDIX A-4: M9110E Signal I/O Definitions, Revision C.8 Table A-4: M9110E Signal I/O Definitions, Revision C.8 Signal Name Bit or Channel Description Number Internal inputs, U7, J108, pins 9-16 = bits 0-7, default I/O address 322 hex...
  • Page 262 APPENDIX A Model 9110E Instruction Manual Signal Name Bit or Channel Description Number ST_SPAN_CAL 0 = in span calibration 1 = not in span ST_DIAG_MODE 0 = in diagnostic mode 1 = not in diagnostic mode Spare B status outputs, U27, J1018, pins 1-8 = bits 0-7, default I/O address 324 hex...
  • Page 263 Model 9110E Instruction Manual APPENDIX A Signal Name Bit or Channel Description Number Temperature MUX Spare Spare SAMPLE_PRESSURE Sample pressure RCELL_PRESSURE Reaction cell pressure REF_4096_MV 4.096V reference from MAX6241 OZONE_FLOW Ozone flow rate TEST_INPUT_11 Diagnostic test input CONV_TEMP Converter temperature...
  • Page 264 APPENDIX A Model 9110E Instruction Manual EXITHS Exit high span calibration mode EXITMP Exit multi-point calibration mode SLPCHG Slope and offset recalculated Events that can be accessed through the remote interface EXITDG Exit diagnostic mode AZEROW Auto-zero warning OFLOWW Ozone flow warning...
  • Page 265 Model 9110E Instruction Manual APPENDIX A Name Description Units NOCNC2 NO concentration for range #2 N2CNC1 concentration for range #1 N2CNC2 concentration for range #2 STABIL Concentration stability AZERO Auto zero offset (range de-normalized) O3FLOW Ozone flow rate cm³/min RCPRES...
  • Page 266: Appendix A-6: Terminal Command Designators, Revision C.8

    APPENDIX A Model 9110E Instruction Manual APPENDIX A-6: Terminal Command Designators, Revision C.8 Table A-7: Terminal Command Designators, Revision C.8 Command Additional Command Syntax Description Display help screen and this list of ? [ID] commands LOGON [ID] password Establish connection to instrument...
  • Page 267 Model 9110E Instruction Manual APPENDIX A The command syntax follows the command type, separated by a space character. Strings in [brackets] are optional designators. The following key assignments also apply. Terminal Key Assignments Abort line CR (ENTER) Execute command Ctrl-C...
  • Page 268: User Notes

    APPENDIX A Model 9110E Instruction Manual USER NOTES: 044100120 Rev B...
  • Page 269: Appendix B: Spare Parts And Expendables

    Model 9110E Instruction Manual APPENDIX B APPENDIX B: Spare Parts and Expendables NOTE Use of replacement parts other than those supplied by Teledyne-API may result in non-compliance with European standard EN 61010-1. The following listings contain spare parts for individual ordering as well as recommended stocking levels for operation of multiple analyzers to minimize downtime for any one analyzer unit.
  • Page 270 APPENDIX B Model 9110E Instruction Manual Part Number Description Notes AMBIENT 040900000 ORIFICE HOLDER, M9110E REACTION Orifice holder only; does not include fitting, O- CELL rings or any orifice assembly parts 041710000 CPU, CONFIGURATION, "E" SERIES CPU board including EEPROM. Needs...
  • Page 271 Model 9110E Instruction Manual APPENDIX B Part Number Description Notes PS0000029 EOS SWITCHING PS, +5V, +/-15V, 5 V and ±15V power supply for all electronics PS0000031 EOS SWITCHING PS, 12V, 60W 12 V Power supply for PMT cooler and valves...
  • Page 272 APPENDIX B Model 9110E Instruction Manual Quantity included in one kit. Part Number Description Notes 046060000 AKIT, ORIFICE EXPENDABLES, Expendables kit for one orifice M9110Ex assembly This kit contains the following items FL0000001 FILTER, SS Sintered stainless steel filter HW0000020...
  • Page 273 Model 9110E Instruction Manual APPENDIX B USER NOTES: 044100120 Rev B...
  • Page 274: Intentionally Blank

    APPENDIX B Model 9110E Instruction Manual INTENTIONALLY BLANK 044100120 Rev B...
  • Page 275 Model 9110E Instruction Manual APPENDIX B RECOMMENDED SPARE PARTS STOCKING LEVELS 9110E UNITS PART NO. DESCRIPTION 2- 5 6-10 11-20 21-30 010150100 Subassy, Moly Con Cartridge 011310000 * Assy, Dryer, NOX 011930000 PMT, NOX 014080100 Assy, HVPS, SOX/NOX 014610000 Replacement Cooler Assembly...
  • Page 276 APPENDIX B Model 9110E Instruction Manual USER NOTES: 044100120 Rev B...
  • Page 277: Appendix C: Warranty/Repair Questionnaire Model 9110E

    APPENDIX C: Warranty/Repair Questionnaire Model 9110E CUSTOMER: ____________________________________ PHONE: ___________________________________________ CONTACT NAME: _______________________________ FAX NO. ___________________________________________ SITE ADDRESS: ______________________________________________________________________________________ MODEL 9110E SERIAL NO.: ________________________FIRMWARE REVISION: _______________________________ 1. ARE THERE ANY FAILURE MESSAGES? ____________________________________________________________ _____________________________________________________________________________________________________ _____________________________________________________________________________________________________ PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST...
  • Page 278 Model 9110E Instruction Warranty/Repair TELEDYNE Manual Questionnaire INSTRUMENTS Advanced Pollution Instrumentation Model 9110E A Teledyne Technologies Company 2. WHAT IS THE RCELL & SAMPLE PRESSURES WITH THE SAMPLE INLET ON REAR OF MACHINE CAPPED? RCELL PRESS - IN-HG-A SAMPLE PRESSURE - IN-HG-A 3.
  • Page 279: Appendix D: Diagrams And Schematics

    Model 9110E Instruction Manual APPENDIX D APPENDIX D: Diagrams and Schematics Table D-1: List of Included Diagrams and Schematics Document # Document Title Number of pages 04574 Document, M9110E Detailed Pneumatic Diagrams for all configura- tions (similar to Figure 3-11 and Figure 3-12)
  • Page 280 APPENDIX D Model 9110E Instruction Manual USER NOTES: 044100140 Rev C...

Table of Contents