1. FEATURES Standard 94 dB/1000 Hz & 114 dB/1000 Hz sound calibrator, useful to calibrate another " Sound Level Meter " precisely. 2. SPECIFICATIONS Frequency 1000 Hz 2% Sound Pressure 94 dB 0.75 dB. Level 114 dB 0.9 dB. Spec. tested under the environment RF Field Strength less than 3 V/M &...
4. CALIBRATION 1) If the calibrated " SOUND LEVEL METER " used the 0.5" microphone, then on the " Couple Section " (3-1, Fig. 1) should install the " 0.5" Mic. Adapter " (3-2, Fig. 1) first. Consideration : If the calibrated " SOUND LEVEL METER " used the 1 inch microphone, then on the "...
Fig. 2 5. BATTERY CHECK Under the normal operation, slide " Off/94 dB/114 dB Switch " ( 3-3 , Fig. 1 ) to the " 94 dB " or " 114 dB " position, the " Power/Low Battery Indicator " ( 3-4, Fig. 1 ) will light up, then the battery voltage is normal.
2) Remove the old batteries & replace the two new 9V batteries. 3) Make sure the battery is installed to the right position & connected to the proper polarization with the battery snap. 4) Reinstate the battery cover after changing the batteries. Fig. 3 0402-SC-942...