Table of Contents

Advertisement

Quick Links

About This Manual
P/N: 4710.01502A01
Product Model: E1V
Release Date: February, 2018
Copyright © 2018 SonoScape Medical Corp. All rights reserved.
Statement
SonoScape Medical Corp. (hereinafter called SonoScape) owns the intellectual property rights to
this manual, and also maintains the contents of this manual as confidential information. This manual
is a reference for the operation, maintenance and cleaning of this product and does not convey any
license under the patent rights of SonoScape, nor the rights of others.
This manual contains the information protected by copyrights or patents. Reproduction, amendment
or translation of this manual in any manner whatsoever without the written permission of SonoScape
is strictly prohibited.
All information contained in this manual is believed to be correct. SonoScape shall not be liable for
errors contained herein or for incidental or consequential damages in connection with the furnishing,
performance or use of this manual. SonoScape does not assume any liability arising out of any
infringements of patents or other rights of third parties.
This manual is based on the maximum configuration and therefore some contents may not apply to
your product.
This manual is subject to change without prior notice and legal obligation.
Manufacturer's Responsibility
SonoScape is responsible for the effects on safety, reliability and performance of this product, only if
all the following requirements are met.
A ll installation operations, expansions, changes, modifications and repairs of this product are
conducted by SonoScape authorized personnel .
The use or application of the product or the use of parts or accessories is approved by
SonoScape.
The electrical installation of the relevant room complies with the applicable national and local
requirements .
The product is used in accordance with the instructions for use.
Documentation
SonoScape provides the documentation consisting of various manuals:
The basic user manual (this manual) describes the basic functions and operating procedures of
the system.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the E1V and is the answer not in the manual?

Questions and answers

Summary of Contents for Sonoscape E1V

  • Page 1 SonoScape is strictly prohibited. All information contained in this manual is believed to be correct. SonoScape shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this manual.
  • Page 2: Contact Information

    Address: 4/F, 5/F, 8/F, 9/F & 10/F, Yizhe Building, Yuquan Road, Nanshan, Shenzhen, 518051, Guangdong, China Tel: +86 755 2672 2890 Fax: +86 755 2672 2850 http://www.sonoscape.com E-mail: sonoscape@sonoscape.net EU Representative: SonoScape Europe S.r.l. Address: Via Luigino Tandura, 74-00128 Rome, Italy Tel: +39 06 508 2160 Fax: +39 06 508 4752 http://www.sonoscapeurope.com...
  • Page 3: Table Of Contents

    Contents 1     S afety   ............................1 1.1 Intended Use ..........................2 1.2 Safety Precautions ........................2 1.2.1 Electrical Safety ......................2 1.2.2 Mechanical Safety ......................4 1.2.3 Accessories Caring ......................4 1.2.4 Biohazard Considerations .....................5 1.3 Acoustic Power Principle ......................5 1.3.1 Biological Safety......................6 1.3.2 ALARA .........................6 1.3.3 Mechanical and Thermal Indices ..................6 1.3.4 Probe Surface Temperature Limits ................8 1.3.5 Imaging Functions that Change Acoustic Output ............8...
  • Page 4 Contents 3.2.3 Installing/Removing the Battery .................26 3.2.4 Status Indicator ......................27 3.3 Powering On/Off the System....................28 3.3.1 To Power On the System .....................28 3.3.2 To Activate Screen Saver ....................28 3.3.3 To Power Off the System ....................28 3.4 Adjusting the Brightness of the Monitor ................28 3.5 Connecting the Probe ......................28 3.6 Connecting the Peripheral Device ..................29 3.6.1 Connecting the Foot Switch (optional) ...............29...
  • Page 5 Contents 6     A cquiring Images ........................51 6.1 Selecting a Probe and an Exam Type ..................52 6.1.1 To Customize a Preset ....................52 6.1.2 To Arrange the Presets Display ...................53 6.2 Acquiring B-Mode Images ....................54 6.2.1 Entering B Mode ......................54 6.2.2 Optimizing B-Mode Images ..................54 6.3 Acquiring M-Mode Images ....................59 6.3.1 M Mode........................59 6.3.2 Optimizing M-Mode Images ..................60...
  • Page 6 Contents 8.2.1 Viewing a Current Image/Cine ...................78 8.2.2 Retrieving an Image ....................79 8.3 Backing up Data ........................79 8.4 Importing Data to the System ....................80 9     W orking with DICOM ......................81 9.1 Verifying Connectivity ......................82 9.2 DICOM Storage ........................82 9.3 DICOM Queue ........................83 10    ...
  • Page 7 Contents 11.3.2 Battery Safety ......................105 11.3.3 Mechanical Safety ....................105 11.3.4 Image Recording .....................105 11.4 Troubleshooting .........................105 11.5 Equipment Disposal ......................106 11.6 Customer Service .......................106 Appendix  A  Specifications ..................... 107 Appendix  B  EMC Guidance and Manufacturer’s Declaration ......... 108 B.1 Electromagnetic Emissions ....................108 B.2 Electromagnetic Immunity ....................108 B.3 Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the Equipment ....................112 Appendix ...
  • Page 8 T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k .
  • Page 9: Afety

    1     S afety This chapter describes the important information for operating this system. To ensure the safety of both operator and animal, please read the relevant details in this chapter carefully before using this system. You should be thoroughly familiar with the precautions provided in this manual. Otherwise, the manufacturer is not responsible for the effects on safety, reliability and performance of the system.
  • Page 10: Intended Use

    1 Safety 1.1 Intended Use The ultrasound system is intended for the following applications: pet (feline, canine), animal husbandry, laboratory, and horse race applications. The ultrasound system also provides the measurement and calculation packages used for clinical diagnosis purposes. Contraindication: The system is not intended for ophthalmic use or any use causing the acoustic beam to pass through the eye.
  • Page 11 1 Safety Within the environment that is 1.8 meters (6 feet) around a patient, connect peripherals ● to the auxiliary power outlet which is capable of isolation protection; or, power the peripherals by the auxiliary output cable or the isolation transformer complied with EN/IEC 60601-1, or the power input of the same safety level.
  • Page 12: Mechanical Safety

    1 Safety Accessory equipment connected to the analog and digital interfaces must be certified ● according to the respective EN/IEC standards (for example, EN/IEC 60950 for data processing equipment and EN/IEC 60601-1 for medical equipment). Furthermore, all configurations shall comply with the system standards EN/IEC 60601-1. Do not use an endocavitary probe on the body surface of a patient.
  • Page 13: Biohazard Considerations

    1 Safety Only trained physicians or sonographers under ultrasound guidance can handle the ● biopsy needle guides. During the operation, the operator must observe proper needle insertion sequencing with the needle guide to avoid undue discomforts, unnecessary risks or injuries to the patient. Use the legally marketed probe sheath.
  • Page 14: Biological Safety

    1 Safety Do not expose the fetus to prolonged scanning in the Doppler mode. ● Although the output power is automatically controlled for the selected applications, ● high TI values should be kept to a minimum or avoided in obstetric applications. You should be familiar with the performances and operations of the system, observe ●...
  • Page 15 1 Safety ■  MI/TI Explanation In October 1987, the American Institute of Ultrasound in Medicine (AIUM) ratified a report prepared by its Bioeffects Committee (Bioeffects Considerations for the Safety of Diagnostic Ultrasound, J Ultrasound Med.,Sept. 1988: Vol. 7, No. 9 Supplement), sometimes referred to as the StoweReport, which reviewed available data on possible effects of ultrasound exposure.
  • Page 16: Probe Surface Temperature Limits

    1 Safety 1.3.4 Probe Surface Temperature Limits For probes intended for internal applications, e.g. the endocavitary probe or transesophageal probe, the surface temperature of the probe may change by adjusting system parameters. The maximum surface temperature of the endocavitary probes is 43°C. To protect the animal against the harm of excessive temperature, the probe stops working automatically when its temperature reaches the limit.
  • Page 17 1 Safety Symbol Meaning Type BF Applied Part Power On/Off button Alternating current Direct current Degree of IP protection Non-ionizing electromagnetic radiation. Manufacturer Date of manufacture Follow instructions for use Network Port USB Port Fragile Keep dry Maximum stacking limit of packages Maximum of two layers allowed for the system Keep this way upward Basic User Manual...
  • Page 18 1 Safety Symbol Meaning Serial Number Use it in ventilated and dry indoor environment (for AC adaptor only) Rechargeable (for battery only) This symbol indicates that waste electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately.
  • Page 19: Ystem Overview

    2     S ystem Overview This system adopts advanced ultrasonic Doppler technologies. To ensure the performance and availability of this system, you should be thoroughly familiar with the operations of system components, the control panel, the key panel and the basic screen.
  • Page 20: System Configuration

    2 System Overview 2.1 System Configuration The standard system configuration mainly consists of the followings items: ● Main unit ● Probes Others: See the Packing List in the packaging box. ● 2.2 Physical Specifications ● LCD monitor: 15.6’’ color display ● System dimensions: approx.
  • Page 21: Front Panel

    2 System Overview 2.3.1 Front Panel Figure 2-2 Front View 1 Handle 4 Control panel 2 Probe holder 5 Indicators 3 LCD monitor 6 Release latch Basic User Manual...
  • Page 22: Side View

    2 System Overview 2.3.2 Side View Figure 2-3 Side Views 1 HDMI video output port 5 Print control port 2 USB port 6 S-Video output port 3 Audio output port 7 Network port (DICOM) 4 Audio input port 8 Power on/off button Basic User Manual...
  • Page 23: Back View

    2 System Overview 2.3.3 Back View Figure 2-4 Back View 1 Ventilation holes 5 Foot switch port 2 Cable hanger holder (optional) 6 AC adaptor port 3 Probe port 7 Video output port 4 ECG port (optional) 8 Grounding terminal 2.3.4 Bottom View Figure 2-5 Bottom View ...
  • Page 24: Control Panel

    2 System Overview 2.3.5 Control Panel 13 14 15 16 17 18 19 23 24 25 26 27 28 Figure 2-6 Control Panel Name Description Probe Select the corresponding exam type. Patient Enter the New Patient screen. Enter the Measurement Report screen. Report Review Display the Patient Exam List screen.
  • Page 25 2 System Overview Name Description Angle ● Adjust the flow direction in the PW mode. ● Adjust the arrow direction when the arrow annotation feature is activated. Save images. Or, the function of this key can be set as required. For details, refer to Section 4.1.3 Save Settings.
  • Page 26 2 System Overview Name Description Enter or exit the M mode. Enter or exit the PW mode. Reserved Enter the B mode. Auto/Gain ● Press it to automatically optimize an image in B/M mode. ● Rotate it to adjust the gain of the current mode. Standby Reserved Indicator...
  • Page 27: Key Panel

    2 System Overview 2.3.6 Key Panel Figure 2-7 Key Panel Name Description <1> Esc key Exit the System Settings screen or New Patient screen. <2> Help key Reserved <3> F3 key The function of this shortcut key can set through the system. The screenshot storage is default settings. For details, refer to Section 4.1.3 Save Settings.
  • Page 28: Basic Screen

    Enter a space. <20> Setup key Enter the System Settings screen. <21> Arrow Add arrows to an image. 2.3.7 Basic Screen New hosp. XXXXX 19700101_203151 2017/07/12 10:39 Sonoscape female MI 0.2 TIS 0.7 cm/s 1/1P 4/20 7.5M Cine Frequency Gray Map Compond μScan...
  • Page 29 2 System Overview Manufacturer logo Cine loop status (in frozen mode) Information bar Imaging parameters Status bar Parameter value/level Clipboard Depth scale and focal position BACK/NEXT key Imaging information Cine loop icon (in frozen mode) 14 Measurement parameter/annotation box Measured result box Grayscale map Imagining area Basic User Manual...
  • Page 30 T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k .
  • Page 31: Reparing The System

    3     P reparing the System System preparation is necessary before using the system. The preparation includes but is not limited to positioning and adjusting the system, connecting the probe and the peripheral devices.
  • Page 32: Positioning The System

    3 Preparing the System 3.1 Positioning the System  Leave at least 20cm at the back and both sides of the system for ventilation. Otherwise, the temperature rise could cause failures. Place this system on a level surface. You can place the system on the trolley (sold separately) provided by manufacturer. For details on installing the system to the trolley, please refer to the relevant user manual.
  • Page 33: Using The Battery

    3 Preparing the System 3.2.2 Using the Battery Do not disassemble or alter the battery. Otherwise, there is a danger of explosion. ● Replace the battery with the type recommended by the manufacturer. ● ● Install/Remove the battery only when the system is powered off. ●...
  • Page 34: Installing/Removing The Battery

    3 Preparing the System Only use the battery if there is an emergency or no reliable AC power supply. It is ● recommended to use the AC supply instead of the battery. ● If you do not attempt to use the battery within 3 months, fully charge the battery and store it in the temperatures of -20°C to 45°C.
  • Page 35: Status Indicator

    3 Preparing the System 3.2.4 Status Indicator The status indicators are located on the control panel to indicate the battery charging and discharging statuses. When the system is powered on, you can move the cursor onto the battery icon by using the trackball to view the current battery capacity. The battery icon is displayed in the status bar.
  • Page 36: Powering On/Off The System

    3 Preparing the System 3.3 Powering On/Off the System ● Do not power off the system during system upgrade or a data transmission. ● An extra 30-60 seconds are needed for powering on the system if it is powered off in an abnormal way previously. ●...
  • Page 37: Connecting The Peripheral Device

    3 Preparing the System Rotate the locking lever to the position to lock the probe firmly. 3.6  Connecting the Peripheral Device To avoid damaging the foot switch, do not place the foot switch on the ultrasound ● system. Use only the foot switch provided by the manufacturer. Otherwise, the ultrasound ●...
  • Page 38: Connecting The Printer

    3 Preparing the System 3.6.2 Connecting the Printer Video printer, laser printer and inkjet printer recommended by the manufacturer can be connected to the system. ■  Video Printer Connect a video printer to the system. VIDEO OUT VIDEO INPUT PRINT REMOTE Ultrasound system Video printer The printer can work without any configurations after a successful connection.
  • Page 39: Ustomizing Your System

    4     C ustomizing Your System The System Settings menu allows you to specify the operating status, system date and time, monitor brightness, etc. All your customized settings can remain even after rebooting the system.
  • Page 40: General System Settings

    4 Customizing Your System After you power on the system, the LCD monitor displays the real time B mode by default. Press the Setup key on the key panel to enter the System Settings menu. To make settings on the System Settings menu, ●...
  • Page 41 4 Customizing Your System Item Description Freeze Response Select the feature to be enabled after you press the Freeze button. End Exam Response Select the feature to be enabled after you press the End Exam button. Trackball Sensitivity Adjust the sensitivity for the trackball movement. A greater value brings about a higher sensitivity.
  • Page 42: Display Settings

    4 Customizing Your System 4.1.2 Display Settings SystemSettings General General Display Save Biopsy Exam Preset Width Ruler System Sleep Measure Screen Saver Wait Report Key Volume Trackball Volume Peripheral Light color Green Brightness&Colors DICOM Keyboard & Monitor Brightness Annotation Auto About Clear Annotation on UnFreeze Keyboard...
  • Page 43: Save Settings

    4 Customizing Your System 4.1.3 Save Settings SystemSettings General General Display Save Biopsy Exam Preset Save Operation Save Format/Area Save Sound Measure ImageArea+InfoBar Area High Send image to DICOM storage Cine Report Image Send cine to DICOM storage None Peripheral Key Function Freeze Store Save Image...
  • Page 44: Biopsy Settings

    4 Customizing Your System 4.1.4 Biopsy Settings SystemSettings General General Display Save Biopsy Exam Preset Biopsy Grid Biopsy Guide Line Bracket Measure Probe Type Biopsy Angle Report L741 45° C613 12° Peripheral DICOM About Biopsy Parameters Guide Line Type Line Guide Line Display Apply Exit...
  • Page 45: Exam Preset Settings

    4 Customizing Your System 4.2 Exam Preset Settings 4.2.1 Department Settings System Settings General Department Preset Exam Preset Measure Report Animal Husbundry Peripheral DICOM Horse Race About Apply Exit Figure 4-5 Department Settings Set the department according to the practical application. Basic User Manual...
  • Page 46: Probe Settings

    4 Customizing Your System 4.2.2 Probe Settings System Settings General Department Preset Exam Preset Available Items Probe Selection Measure L741 Report Canine ABD M Canine Cardiac Equine Tendon Peripheral DICOM About Selected Items Delete Canine ABD M Canine Cardiac Equine Tendon Apply Load Defaults Exit...
  • Page 47: Measurement Settings

    4 Customizing Your System 4.3 Measurement Settings You can make measurement settings on the General and Formula tabs. 4.3.1 General Settings SystemSettings General General Exam Preset Keep Result Window Measure Report Peripheral DICOM About Apply Load Defaults Exit Figure 4-7 Measurement Settings - General Tab Enable or disable the display of the current result box when unfreezing the image. Set Keep Result Window to ON, the result box is not cleared after you complete the ●...
  • Page 48: Report Settings

    4 Customizing Your System 4.4 Report Settings SystemSettings Logo General Exam Preset Impot/Export bmp: 170×60 Measure Report Peripheral Title/Font/Color Title 1 Title DICOM Title 2 Sub-Title About Title 3 Background Color Content Text Color Display Items For Report Patient Information Patient Information Exam Information Exam Information...
  • Page 49: Peripheral Devices Settings

    4 Customizing Your System Item Description Text Color Set the text color. Display Items for Select items to be displayed on the measurement report. Animal Report and exam information is displayed on the report by default. Other information can also be displayed on the report after you tick Measurements, Image or Comments.
  • Page 50: Network Settings

    4 Customizing Your System Item Description Printer IP Address Set the IP address of the computer connecting a printer. Printer Port Number Set the port number of the printer. Left Key/Right Key Set the function of the foot switch. 4.5.2 Network Settings SystemSettings General Basic...
  • Page 51: Dicom Storage Service Settings

    4 Customizing Your System ■  Wireless Network Item Description Wireless Network Enable or disable the wireless network. Name Displays the name of the wireless network. State Displays the connection status of the wireless network. Level Displays the signal strength of the wireless network. Advance Click Advance to view the information of your current network, and make settings of your local network.
  • Page 52: Viewing System Information

    4 Customizing Your System Item Description Service List ● Select Add from the drop-down list beside Service List, input the relevant parameters information and click Add to add a new server. ● Select any server from the drop-down list and click Delete to delete the server.
  • Page 53: Reparing For An Exam

    5     P reparing for an Exam You can start an exam without entering any animal information. To avoid animal identification errors, you should always verify the identification information.
  • Page 54: Acquiring Animal Information

    5 Preparing for an Exam 5.1 Acquiring Animal Information You can acquire the animal information in the following ways. ● Create new animal information ● Retrieve archived animal information Receive animal information using the DICOM worklist ● 5.1.1 Beginning a New Patient Press the Patient button on the control panel to display the New Patient screen in Figure 5-1.
  • Page 55 5 Preparing for an Exam Animal ID Enter the animal ID. The system will generate the animal ID automatically according to the current system time if this item is left blank. NOTE: The animal ID is an important identification, once it is saved, you cannot modify it.
  • Page 56: Retrieving Archived Information

    5 Preparing for an Exam NOTE: • You must enter the length and weight for the current animal when you select Cardiology as the application mode. BSA will be calculated automatically in accordance with the BSA set on the Measure menu. •...
  • Page 57: Using Dicom Worklist

    5 Preparing for an Exam Search for the archived information Enter the search requirements, and the desired information is displayed in the list. Patient Exam List Animal ID Owner’s Name Animal’s Name Animal Type Reset Date 2015/12/12 2015/12/12 Patient View Current Exam Animal ID Master Name Animal Name...
  • Page 58 5 Preparing for an Exam Select the desired patient information, and then click View to enter the Exam View screen. Select the desired exam, and then click Resume Exam to resume the exam. ■  To Complete an Exam Press the End Exam button on the control panel to end the current exam. You can view the status of the desired exam on the Exam View screen.
  • Page 59: Cquiring Images

    6     A cquiring Images You can acquire images by optimizing the relevant parameters for clinical diagnosis.
  • Page 60: Selecting A Probe And An Exam Type

    6 Acquiring Images 6.1 Selecting a Probe and an Exam Type Once you power on the system, the system enters the real time B mode by default. You can press the Probe button to enter the Application Mode screen (as shown in Figure 6-1) and select the desired application mode. Canine ABD M Canine Cardiac Equine Tendon Figure 6-1 Application Mode Screen...
  • Page 61: Acquiring Images

    6 Acquiring Images User Pre-Set Menu Name Dept. Animal Husbandry Type Bovine ABD Part Canine ABD M Save Exit Figure 6-2 User Pre-set Menu Screen − Enter the name of the application preset in the Name textbox. − Select the department from the Dept. lists. −...
  • Page 62: Acquiring B-Mode Images

    6 Acquiring Images 6.2 Acquiring B-Mode Images B-mode imaging is intended to provide information of anatomical structure of soft tissue. 6.2.1 Entering B Mode The system enters B mode by default after you select a probe model and an exam type. You can press the B button on the control panel to enter B mode from other imaging modes.
  • Page 63 6 Acquiring Images ■  Gain The B mode gain (2D gain) determines the amplifying factors for the received echoes and the brightness of the ultrasound image. The echoes are amplified with the same gain value regardless of depth. You can rotate the Gain knob on the control panel to adjust the value. ■ ...
  • Page 64 6 Acquiring Images ■  Frequency The probe is capable of receiving a broadband signal with a certain start frequency and a certain bandwidth. A higher frequency brings about higher resolution and lower penetration. To adjust the frequency: Press the function knob to select Frequency. Rotate the function knob to adjust the value.
  • Page 65 6 Acquiring Images ■  Line Density Line density refers to the amount of ultrasound beam that makes up the image. Increasing the value of line density improves the resolution and decreases frame rate. Therefore, you have to balance the frame rate and the image quality when adjusting the line density. A higher line density is useful in obtaining a very high resolution, such as the thyroid or testicles.
  • Page 66 6 Acquiring Images NOTE: Expose the animal to the lowest practical transmit power level for the shortest possible period of time to achieve a satisfactory diagnosis. ■  μScan μScan is used to improve the image quality. To use the μScan: Press the function knob to select μScan.
  • Page 67: Acquiring M-Mode Images

    6 Acquiring Images ■  Image Rotation You can rotate the image to get more information. To rotate the image: Press the function knob to select Rotation. Rotate the function knob to adjust the settings. 6.3 Acquiring M-Mode Images M mode is generally used with 2D imaging, and is helpful for cardiac applications. In the M mode, you can position the M line in the 2D image on the anatomy of interest, and then learn about the tissue motion along that line in M-mode trace.
  • Page 68: Optimizing M-Mode Images

    6 Acquiring Images Y axis: depth X axis: time M Image Figure 6-6 M-Mode Imaging Screen − X-axis is the time scale. − Y-axis is the depth scale. NOTE: Once the M mode is activated, you can move the trackball to stop the M trace and adjust the M line.
  • Page 69 6 Acquiring Images You can rotate the Gain knob to adjust the value. ■  Frequency The probe is capable of receiving a broadband signal with a certain start frequency and a certain bandwidth in the real time B mode. The higher frequency brings about higher resolution and lower penetration.
  • Page 70: Acquiring Spectral Doppler Images

    6 Acquiring Images ■  Persistence Persistence is used to average consecutive frames to provide a smoother appearance with less noise. Use lower persistence values for fast-moving organs or tissues and higher persistence values for a smoother appearance. To adjust the persistence: Press the function knob to select Persist.
  • Page 71: Pw Mode

    6 Acquiring Images 6.4.1 PW Mode Pulsed Wave Doppler (PW) is a Doppler mode that measures velocity in a PW sample volume and displays that information in a spectral trace with audio output. PW includes conventional PW and High Pulse Repetition Frequency (HPRF). PW Doppler can be combined with the B mode for rapidly selecting the anatomical site for PW Doppler examination.
  • Page 72 6 Acquiring Images Press the Update button to activate the PW mode. The PW spectrum is displayed at the lower part of the screen after being activated, as shown in Figure 6-8. Y axis: frequency PW spectrum X axis: time Figure 6-8 PW-Mode Imaging Screen −...
  • Page 73: Optimizing Spectral Doppler Images

    6 Acquiring Images Y axis: frequency X axis: time Figure 6-9 Inactivated B+PW Mode 2 Imaging Screen You can also adjust the position and angle of spectral Doppler line, the size of sample volume gate, and the direction of the flow cursor. Press the PW button to exit. 6.4.2 Optimizing Spectral Doppler Images After activating the PW mode, the imaging parameters display at the bottom of the LCD monitor.
  • Page 74 6 Acquiring Images You can rotate the Gain knob to adjust the value. ■  Scale Scale is used to adjust the velocity scale. You can press the Scale switch to adjust the value. NOTE: The value of wall filter may change with the scale value. ■ ...
  • Page 75 6 Acquiring Images Press the function knob to select Frequency. Rotate the function to adjust the value. ■  Sweep Speed Sweep speed is used to set the sweep speed of the Doppler trace. The refresh speeds of the time scale and the frequency scale vary with the sweep speed. To adjust the sweep speed: Press the function knob to select Sweep Speed.
  • Page 76 6 Acquiring Images ■  Steer Steer can be adjusted to change the angle of the spectrum Doppler line when performing a real time mode by using the linear probes. To adjust the steer: Press the function knob to select Steer. Rotate the function knob to adjust the settings.
  • Page 77: Orking With Images

    7     W orking with Images You can work with the acquired images by using the features provided by the system, such as the split screen display and annotations.
  • Page 78: Imaging Features

    7 Working with Images 7.1 Imaging Features 7.1.1 Imaging Reverse ● Press the key on the key panel to reverse the real time scan left or right. Press the key on the key panel to reverse the real time scan up or down. ●...
  • Page 79: Auto Optimization

    7 Working with Images ■  Quad-Split Screen Display In the B mode, press the key on the control panel to activate the quad-split screen display. The B-mode image is taken as an example in the following figure. 7.5-10.5 14.0 µS Figure 7-2 Quad-Split Screen Display Press the THI key on the control panel to enter the corresponding mode for the real ●...
  • Page 80: Harmonic Imaging

    7 Working with Images 7.1.4 Harmonic Imaging Harmonic Imaging (THI) uses the harmonic frequencies generated by tissue. The fundamental transmit frequency results in a harmonic frequency echo. The harmonic frequency received from the tissue is a multiple of the fundamental transmit frequency. The fundamental and any accompanying artifacts are missing from the harmonic frequency that the system receives.
  • Page 81: Freezing An Image

    7 Working with Images 7.3 Freezing an Image Press the Freeze button on the control panel to capture an image in a real time scan, Cine is displayed at lower right corner of the screen, and the feature to be applied is also activated.
  • Page 82: Annotations And Body Marks

    7 Working with Images Rotate the function knob corresponding to Set First to select the start frame in the frozen mode. Rotate the function knob corresponding to Set Last to select the end frame. Press the Cine button on the control panel to save the cut cine. 7.5 Annotations and Body Marks The annotation feature allows you to type an annotation or add an annotation from the predefined annotation library.
  • Page 83: Body Mark

    7 Working with Images Position the arrow to the desired place by using the trackball. Rotate the Angle knob to adjust the arrow angle if needed. Press the confirm button to confirm. Repeat the above steps to add more arrows if needed. 7.5.3 Body Mark Perform the following steps.
  • Page 84 7 Working with Images NOTE: You can press the Clear button on the control panel to delete all annotations, arrows, body marks, measurement marks and measured results from the image in any status. Perform this deletion with caution. Basic User Manual...
  • Page 85: Anaging Images/Data

    8     M anaging Images/Data Images in this system include frame images and cine, you can save or review them after the acquisition. Meanwhile, images can also be backed up to a USB drive for future review if it is needed.
  • Page 86: Storing An Image/Cine

    8 Managing Images/Data 8.1 Storing an Image/Cine The following methods can be used to store images/cines. 8.1.1 Storing an Image Store an image as follows. Press the Image button on the control panel in the real time mode or frozen mode to ●...
  • Page 87: Retrieving An Image

    8 Managing Images/Data Review Exam Exam Animal ID:345 Animal Name: XXX Files Number:19 Images Reports 2 × 2 Select All Delete Print DICOM Send Export Export Patient List Exit If the animal has had more than one exam, you can select the desired exam from the Exam drop-down list.
  • Page 88: Importing Data To The System

    8 Managing Images/Data NOTE: • Complete all exams before backing up the data. • Please use a legally marketed USB drive ( FAT, FAT32 and NTFS format) to back up the data. Perform the following steps. Connect a USB drive to the system. Select the Patient button >...
  • Page 89: Orking With Dicom

    9     W orking with DICOM DICOM (Digital Imaging and Communications in Medicine) is a standard created by National Electrical Manufacturers Association (ACR-NEMA) to regulate the distribution and viewing of medical images such as ultrasound images and cine. If the system is configured with the DICOM module, you can: ●...
  • Page 90: Verifying Connectivity

    9 Working with DICOM 9.1 Verifying Connectivity Perform the following steps. Connect the system to the local DICOM network server by using a network cable. Edit the relative settings for the local network and DICOM server. For details, refer to Section 4.5 Peripheral Devices Settings and Section 4.6 DICOM Storage Service Settings.
  • Page 91: Dicom Queue

    9 Working with DICOM 9.3  DICOM Queue The DICOM Queue is used to check the sending status of DICOM storage. Select the Patient button > Patient List > DICOM Queue to enter the DICOM Queue screen. DICOM Queue Reset Animal ID Animal’s Name Type 05/04/2014 05/05/2014 Date...
  • Page 92 T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k .
  • Page 93: Robes And Biopsy Brackets

    10     P robes and Biopsy Brackets  You should become thoroughly familiar with the operations of the probe and biopsy bracket before using them. To ensure the performance of the probe and the biopsy bracket, you should periodically check, clean, disinfect and maintain them.
  • Page 94: Probe

    10 Probes and Biopsy Brackets NOTE: A general instruction for the probe and biopsy bracket is provided in this manual. For details, refer to the relevant manuals. 10.1 Probe This system supports a wide range of probes which make the system remarkably versatile. The schematic diagrams of probes are as follows.
  • Page 95: Probe Usage

    10 Probes and Biopsy Brackets 10.1.2 Probe Usage Use only the probes provided by the manufacturer. Otherwise, the system may not ● perform and an accident such as a fire may result. Use the probe carefully. If any part of the probe surface is scratched, immediately stop ●...
  • Page 96 10 Probes and Biopsy Brackets Swelling of the acoustic window material. ● Cracks or wear on the probe cable. ● Cracks on the probe connector or any other kinds of visible damage. ● ● Deformed pins or broken pins inside the probe connectors. ●...
  • Page 97: Cleaning The Probe

    10 Probes and Biopsy Brackets 10.1.3 Cleaning the Probe To avoid electrical shock, disconnect the probe from the system before cleaning. ● To avoid potential disease transmission, you should wear sterile medical gloves and ● protective goggles while cleaning. ● Do not use the cleaner that contains sodium sulfate and sodium hypochlorite, or sodium hypochlorite, to clean the probe.
  • Page 98: Disinfecting Or Sterilizing The Probe

    10 Probes and Biopsy Brackets 10.1.4 Disinfecting or Sterilizing the Probe There is a cleared list of sterilants and high-level disinfectants for use in processing ● reusable medical and dental devices on the FDA website. That list can be consulted to find agents that may be useful for probe disinfection. Please refer to the following URL for FDA-Cleared Sterilants and High Level Disinfectants: http://www.fda.gov/ MedicalDevices/DeviceRegulationandGuidance/ReprocessingofSingle-UseDevices/ ucm133514.htm.
  • Page 99 10 Probes and Biopsy Brackets Classification Definition Level of Disinfection Application Critical A device that enters Sterilization Intraoperative, normally sterile tissue biopsy use or or the vascular system. blood contacting probe Semi-critical A device that comes High-level disinfection Endocavitary into contact with intact probe mucous membranes and does not ordinarily...
  • Page 100 10 Probes and Biopsy Brackets Disinfectant Manufacturer Active Level for Active Contact Contact Ingredient Ingredient Type Period T-spray II Pharm. Inc. Quat. Ammonia S p r a y / <10 minutes Wipe T-spray Pharm. Inc. Quat. Ammonia S p r a y / <10 minutes Wipe ■ ...
  • Page 101: Sterilizing The Probe

    10 Probes and Biopsy Brackets 10.1.5 Sterilizing the Probe You should sterilize the intraoperative, biopsy use probe or any probe that comes into contact with blood after each use. Perform the following steps. Clean the probe. For details, refer to Section 10.1.3 Cleaning the Probe. Sterilize the intraoperative probe as shown in Table 10-3.
  • Page 102: Storage And Transportation

    10 Probes and Biopsy Brackets Dry the probe cable with a lint-free soft dry cloth. Table 10-4 Recommended Method for Medium-level Disinfection Disinfectant Manufacturer Active Level of Active Contact Contact Period Ingredients Ingredients Type T-spray II Pharm. Inc. Quat. S p r a y / <10 minutes Ammonia Wipe...
  • Page 103: Biopsy

    10 Probes and Biopsy Brackets 10.2 Biopsy You should properly assemble and verify the biopsy bracket before performing a biopsy. ● Only use the biopsy bracket in conjunction with the probe provided by the manufacturer. Otherwise, the system and probe may be damaged, an accident such as a fire may result in the worst case.
  • Page 104: Assembling The Biopsy Bracket

    10 Probes and Biopsy Brackets 10.2.2 Assembling the Biopsy Bracket ■  To Assemble the Surface Biopsy Bracket Tube Holder Biopsy Guide Tube Fixing Bracket Protrusion Perform the following steps. Wear sterile gloves. Unfold the probe sheath, and apply an appropriate amount of coupling gel into the sheath and onto the face of the probe. Hold the probe and unroll the sheath onto the probe.
  • Page 105 10 Probes and Biopsy Brackets Insert the biopsy needle into the tube. Unfold another probe sheath, and apply an appropriate amount of coupling gel to the inside of the sheath. Hold the probe attached with the biopsy bracket and unroll the sheath onto them. ■ ...
  • Page 106 10 Probes and Biopsy Brackets Slide the inner part into the top housing and push forward until it clicks. Align the biopsy bracket with the orientation mark of probe. Orientation Mark Attach the biopsy bracket to the probe and fasten the biopsy bracket with the locking screw.
  • Page 107: Preparing For A Biopsy

    10 Probes and Biopsy Brackets Unfold another probe sheath, and apply an appropriate amount of coupling gel to the inside of the sheath. Hold the probe attached with the biopsy bracket and unroll the sheath onto them. 10.2.3 Preparing for a Biopsy Before and after using a rectal or transvaginal probe to perform a biopsy, you should clean and sterilize the probe and all the biopsy accessories.
  • Page 108: Performing A Biopsy

    10 Probes and Biopsy Brackets 10.2.5 Performing a Biopsy NOTE: Ensure the biopsy bracket is properly attached to the compatible probe and be verified before you perform a biopsy. Perform the following steps. Choose Setup > General > Biopsy > Biopsy  Guide  Line, set Biopsy Angle and Guide ...
  • Page 109: Sterilizing The Biopsy Bracket

    10 Probes and Biopsy Brackets 10.2.7 Sterilizing the Biopsy Bracket Disinfect the biopsy bracket after each use. ● Use the liquid sterilant that meets local laws and regulations. ● ● If you use the recommended sterilant, ensure the requirements for the sterilant level and biopsy bracket soak period are met.
  • Page 110 T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k .
  • Page 111: Ystem Maintenance

    11     S ystem Maintenance To maintain the safety and functionality of the system, periodically perform the maintenance for the system and accessories.
  • Page 112: Cleaning The System

    11 System Maintenance To avoid electric shock and damage to the system, power off and unplug the system ● from the AC power outlet before cleaning. ● To maintain the safety and functionality of the system, maintenance must be performed at least once every year. Electrical safety tests must also be performed at regular intervals as specified by local safety regulations.
  • Page 113: Maintenance Checks

    11 System Maintenance 11.3  Maintenance Checks To maintain the safety and functionality of the system, the following maintenance checks must be performed by service personnel at least every 1 year. 11.3.1 Electrical Safety Integrity of power cable ● ● Protective conductor resistance ●...
  • Page 114: Equipment Disposal

    Any feedback or inquires concerning our product or services should be directed to the following address. Address: 4/F, 5/F, 8/F, 9/F & 10/F, Yizhe Building, Yuquan Road, Nanshan, Shenzhen, 518051, Guangdong, China Zip Code: 518051 Tel: +86-755-26722890 Fax: +86-755-26722850 E-mail: service@sonoscape.net Basic User Manual...
  • Page 115: Appendix A Specifications

    Appendix  A  Specifications Classifications Ty p e o f p r o t e c t i o n Class I ME equipment (with Adaptor); a g a i n s t e l e c t r i c a l Internally Powered ME equipment (when not shock connected to supply mains)
  • Page 116: Appendix B Emc Guidance And Manufacturer's Declaration

    Appendix  B  EMC Guidance and  Manufacturer’s Declaration B.1 Electromagnetic Emissions The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the EQUIPMENT should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment and Guidance RF emissions...
  • Page 117 Appendix B EMC Guidance and Manufacturer’s Declaration Immunity Test IEC 60601 Test Compliance Electromagnetic Environment and Level Level Guidance Electrical fast ±2 kV for power ±2 kV for power Mains power quality should be that transient/burst supply lines; supply lines; of a typical commercial or hospital environment.
  • Page 118 Appendix B EMC Guidance and Manufacturer’s Declaration The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment. Immunity IEC 60601 Compliance Electromagnetic Environment and Guidance Test...
  • Page 119 Appendix B EMC Guidance and Manufacturer’s Declaration Immunity IEC 60601 Compliance Electromagnetic Environment and Guidance Test Test Level Level Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
  • Page 120: Recommended Separation Distances Between Portable And Mobile Rf Communications Equipment And The Equipment

    Appendix B EMC Guidance and Manufacturer’s Declaration B.3 Recommended Separation Distances between Portable and  Mobile RF Communications Equipment and the Equipment The equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the equipment as recommended below, according to the maximum output power of the communications equipment.
  • Page 121: Appendix C In Situ, Derated, And Water Value Intensities

    Appendix  C  In Situ, Derated,  and Water Value Intensities All intensity parameters are measured in water. Since water absorbs very little acoustic energy, these water measurements represent a worst case value. Biological tissue does absorb acoustic energy. The true value of the intensity at any point depends on the amount and type of tissue and the frequency of the ultra-sound that passes through the tissue.
  • Page 122: Appendix D Recommended Coupling Gel, Cleaner And Disinfectant

    Appendix  D  Recommended Coupling Gel, Cleaner and Disinfectant NOTE: • Coupling gels, cleaners and disinfectants listed in this manual are recommended because of their chemical compatibility with product materials, not their biological effectiveness. For the biological effectiveness of a disinfectant, refer to the guidelines and recommendations of the disinfectant manufacturer, Association for Practitioners in Infection Control, U.S.
  • Page 123: Cleaner

    Appendix D Recommended Coupling Gel, Cleaner and Disinfectant Probe Model Applicable Inapplicable 2P1, 5P1, 10L1, C311, C321, C343, Aquasonic 100, Aquasonic Clear, Sterile Aquasonic, Clear Image/ C344, C351, C352, C361, C362, C511, Single, Sterile Ultraphonic Gel, C521, C541, C542, C543, C611, C612, Transonic gel, Clear Scanning Gel, C721, L541, L561V, L662V, L741, Multigel Gel...
  • Page 124: Disinfectant

    Appendix D Recommended Coupling Gel, Cleaner and Disinfectant D.3 Disinfectant Probe Model Applicable Inapplicable MPTEE Mini 7-4, MPTEE 7-4, MPTEE, MPTEE mini, C i d e x O P A M e t r i c i d e 2 8 , 12C-ER Solution, Cidex, M e t r i c i d e 3 0 ,...
  • Page 125: Appendix E Acoustic Output Data

    Appendix  E  Acoustic Output Data Please consult the accompanying CD.

Table of Contents