Page 2
I. Table of Content I.Getting Started II.Specifications III.Panel Description IV.Installation V.Using the Balance VI.Maintenance and Cleaning U.S. Solid ® • USS-DBS64 Models • All Rights Reserved • Page 1 4560 Johnston Parkway, Cleveland, OH, 44128 • www.ussolid.com...
I. Getting Started Thank you for choosing U.S. Solid’s Precision Balance. We hope you enjoy your purchase. Please maintain the equipment according to all instructions within the manual. This product was developed according to CE marking regulations while emphasizing aesthetics and safety for the user. It has ISO certification, specifically ISO 9001.The high quality of the product, coupled with proper user...
41° F- 95° F with fluctuations of <1.8° F/ Hour Pan Size Φ90mm Relative Humidity 50%-85% Power AC110V-240V 50Hz/60Hz,DC9V U.S. Solid ® • USS-DBS64 Models • All Rights Reserved • Page 3 4560 Johnston Parkway, Cleveland, OH, 44128 • www.ussolid.com...
III. Panel Description Pan Explanations 1. ON/OFF: Turns device on and off 2. UNIT: Select between g, ct, oz, ozt, lb, gn, t, dwt, kg, dr, tlT, N 3. PCS: Piece counting mode 4. CAL: Calibration key 5.TARE/ZERO: Tare the balance Features •...
•Balance •Pan •Power Cable •Calibration Weight In case the balance or any component is damaged during transport, contact us immediately in order to process claims within the proper time frame. U.S. Solid U.S. Solid ® ® • • USS-DBS64 Models • All Rights Reserved • Page5 USS-DBS64 Models •...
2. Put the calibration weight on the pan. Wait until the value of the calibration weight stops flashing and is fixed on the display (e.g. 100.000 g, 200.000g, etc., depending on the model). 3. Remove the weight from the pan. U.S. Solid ® • USS-DBS64 Models • All Rights Reserved • Page 6...
Page 8
2. Press the CAL button repeatedly to cycle through different counts. This function can be used with counts of 10, 20, 50, 100, 200 or 500 pieces. 3. Put the corresponding number of pieces on the pan, press the PCS button to U.S. Solid ® •...
• Never use other objects such as pens, pencils, etc. to press the buttons on the control panel; use only your fingers. • Never place anything on the pan heavier than the maximum capacity of the balance, or the sensor could be damaged. U.S. Solid U.S. Solid ® ®...
Page 10
• Do not submerge the balance, nor spill any liquids on it. • When the balance will be unused for a long period of time, lock the rechargeable battery. • If any liquid does come into contact with the electrical parts of the balance, immediately disconnect it and send it in for service to be checked and adjusted if necessary.
Re-calibrate machine, make sure shows conditions. balance is level, and make sure “Err-2” item being weighed is fully stable. U.S. Solid ® • USS-DBS64 Models • All Rights Reserved • Page 10 4560 Johnston Parkway, Cleveland, OH, 44128 • www.ussolid.com...
Page 12
Please inquire with your local community authorities for the proper disposal location. U.S. Solid ® • USS-DBS64 Models • All Rights Reserved • Page11...