Before using the microscope, please read this user manual carefully to study the instrument design, operation modes and procedures, operational limitations, and safety precautions. Due to the continuous improvements in the microscope design, this manual may not reflect minor design changes that do not affect the microscope performance and operation procedures.
CONTENTS 1 DESCRIPTION Purpose Specifications Microscope kit 2 COMPONENTS Stand Focusing mechanism Microscope head Eyepieces Revolving nosepiece Objectives Illumination system Stage LCD screen 3 UNPACKING AND ASSEMBLING 4 BRIGHTFIELD OBSERVATION PROCEDURE Switching on the illumination Adjusting the LCD screen functions Placing the specimen Observing specimens in tall glassware Focusing on the specimen...
MAGUS Bio V360 Biological Inverted Microscope has been designed and tested in accordance with the international safety standards. If properly used, the microscope is safe for the customer’s health, life, property, and the environment. Proper maintenance of the microscope is a prerequisite for its reliable and safe operation.
Coaxial coarse & fine focusing knobs on both sides. Fine focusing knobs (one flat and one volumetric) can be installed both on the right and on the left Focusing mechanism Coarse focusing travel: 37.7mm/circle Fine focusing scale value: 1μm Coarse focusing lock knob Light source 3W LED Intelligent lighting control...
Page 7
Fig. 1. MAGUS Bio V360 Microscope. View from the right 1. Eyepieces 9. Mechanical stage 2. Interpupillary distance adjustment scale 10. Dish holder 3. Microscope head 11. Dish holder locking screw 4. Slot for relief-contrast slider 12. Condenser 5. Coarse focusing tension adjusting knob 13.
Page 8
Fig. 2. MAGUS Bio V360 Microscope. View from the left 1. Diopter adjustment (on each barrel) 8. Stand 2. Eyepiece tubes 9. ON/OFF switch 3. Beam splitter lever 10. Revolving nosepiece 4. Side camera port 11. Objectives 5. Lighting control knob 12.
COMPONENTS STAND The stand has stable ergonomic design. Parts attached to the microscope stand 8 (Fig. 2): – revolving nosepiece 10 (Fig. 2) with objectives 11 (Fig. 2) – mechanical stage 9 (Fig. 1) – condenser 12 (Fig. 1) – illuminator 12 (Fig. 2) –...
EY EPI EC ES The microscope kit includes eyepieces 1 (Fig. 1). The eyepieces have long eye relief and are designed to work with or without glasses. Eyepiece diameter: 30mm. Eyepiece magnification: 10x. Field of view: 22mm. Both eyepieces have a diopter adjustment to compensate for the observer's ametropia. Eyepieces with a different magnification and a 10x eyepiece with 0.1mm scale are not included and are optional.
IL LU MI NATI O N SYST E M The illuminator 12 (Fig. 2) is fixed in the microscope stand. The light source is a 3W LED. The essential part of the illumination system is the condenser. The condenser 12 (Fig. 1) is fixed on the microscope stand.
UNPACKING AND ASSEMBLING The assembly procedure is given in Fig. 3. Fig. 3. Assembling the microscope 1. Remove the microscope from the package. 2. Check the scope of delivery using Section 8 of the User Manual. 3. Place the stand 1 with the attached microscope head on the flat surface. Remove the packaging and dust cover. 4.
BRIGHTFIELD OBSERVATION PROCEDURE SWITCHING ON T H E I LLU MI N ATI O N Before switching on the ON/OFF switch, make sure that the input voltage of the microscope power supply matches the local mains voltage. If not, do not switch on the microscope. Improper input voltage may result in a short circuit or fire. Make sure that the power cord is plugged into the connector on the back panel of the microscope stand.
Sleep mode time setting Enter the settings and select the Time mode. The screen will show the corresponding message. To set the Sleep mode time, press and hold the knob for 3 seconds. The minute indicator will start blinking. Rotate the knob to increase/decrease the time. You can set up the time from 1 minute upwards.
The stage attachment is driven by a rack and pinion gear. Once it has reached the limit, do not rotate the knob 4 or 5 as this may cause the gear to break. Be careful not to hurt your hand on the protruding rack. Fig.
FOCU S IN G ON T H E S PECI ME N Focusing on the specimen is achieved by coarse and fine focusing knobs. Perform the focusing using the 4x objective. Rotate the revolving nosepiece 1 to place the 4x objective in the optical path. The revolving nosepiece is rotated until locked.
We recommend memorizing your diopter adjustment value for future reference. Adjust the distance between the eyepieces to your interpupillary distance by rotating the eyepiece tubes around the central axis until you see a single circular image when looking through the eyepieces with both eyes.
5 PHASE-CONTRAST TECHNIQUE The condenser allows utilizing the phase-contrast technique. Brightfield observations are performed as described above in the user manual. The following instructions apply only to the phase-contrast operation mode. Set up the phase-contrast device as follows: 1. Fully open the aperture diaphragm by moving the lever 1. 2.
Place the calibration slide on the stage instead of the specimen with the scaled side facing up. Using the scale of the calibration slide, calibrate the eyepiece scale for each objective that will be used for measurements. To do this, bring the image focus of the calibration slide scale into sharp focus in the plane of the eyepiece scale and rotate the eyepiece in the tube, setting the strokes of both scales in parallel.
CALI B RATION SL ID E WI T H A CAM E RA The calibration slide (stage micrometer) is used to calibrate the image analysis software for measurements in actual units. In the calibration mode, you should capture an image of the micrometer scale with every objective magnification and indicate the known distance.
7 TROUBLESHOOTING Potential problems and remedies (Table 4): Problem Cause Remedy ELECTRI CAL C OMPONENTS The ON/OFF switch is off Switch on the ON/OFF switch The LED has burned out Have the LED replaced in the service center by a qualified electronics technician No illumination in the field of view The fuse has blown...
MECHANI CAL C OMPONENTS The image does not remain The coarse focusing tension Adjust the coarse focusing tension sharp during observation adjusting knob is loosened, causing adjusting knob the revolving nosepiece to lower spontaneously The coarse focusing knob is too The coarse tension adjusting knob is Loosen the tension of the coarse focusing tight to rotate...
Page 23
10x/22 eyepiece with a scale Optional 10x/22 eyepiece with a center field pointer Optional 12.5x/16mm eyepiece Optional 15x/16mm eyepiece Optional 20x/12mm eyepiece Optional 30x/8mm eyepiece Optional Centering telescope Eyecup Color filter (green) C-mount camera adapter Hoffman modulation contrast device Optional Relief contrast device Round glass stage plate Optional...
9 CARE AND MAINTENANCE REPL ACING T HE B U LB This microscope employs LED bulbs as a light source. The bulbs should be replaced by the equipment vendor or in a qualified service center. If you replace it yourself, the illumination function may be impaired.
MAGUS provides a 5-year international warranty from date of purchase (valid for the entire life of the instrument). The Levenhuk company warrants the product to be free from defects in materials and workmanship. The Seller warrants that the MAGUS product you have purchased meets specification requirements, provided that the Buyer complies with terms and conditions of transport, storage, and operation of the product.