USER MANUAL
Manufacturer's guide and declaration - Electromagnetic emissions
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
From 80 MHz to 800 MHz is applied the higher frequency range.
1
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
2
reflection from structures, objects and people.
a) The intensity of the field for fixed transmitters such as base stations for radio, mobile and cordless phones
and land radio mobile, amateur radio, radio transmitters in the AM and FM and TV transmitters cannot
be predicted theoretically with accuracy. To establish an electromagnetic environment due to fixed RF
transmitters, it should consider the electromagnetic survey of the site. If the field strength measured at the
place where you use the instrument exceeds the applicable level of compliance of the above, the device
should be observed to verify normal operations. If you notice abnormal performance, it may take additional
measures such as a different orientation of the device or re-locate it.
b) The field strength over a frequency range of 150 kHz to 80 MHz should be less than 3 V/m.
RW 2.0 BASIC (WU150) - REV.06 - 13/01/2025
IEC 60601 Compliance Level
3Vrms 150kHz to 80MHz (for
appliances that are not life
supporting)
3 V/m
80MHz to 2,7 GHz
(for appliances that are not
life equipment)
EN
Electromagnetic environment-guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the product
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1.2 √ P
d = 1.2 √ P from 80 MHz to 800 MHz
d = 2.3 √ P from 800 MHz to 2.5 GHz
P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic
site survey
, should be less than the compliance level
1
in each frequency range
the proximity of equipment marked with the following
symbol: