Page 1
FTE-7000A-DWDM Optical Time Domain Reflectometer User's Guide Revision D 01/2021 Information contained in this manual is believed to be accurate and reliable. However, no responsibility is assumed by Terahertz Technologies Inc. for its use nor for any infringements of patents or other rights of third parties that may result from its use.
Page 2
Chapter PAGE Using This Manual.…..……....…................….…………………1 Safety....…...........…….………......…………………...…………2 Quick Start Guide........................…….……..………4 Introduction …………..……......................…….……6 OTDR Inspection and Identification..................………………...……7 Inspection 5.2 Identification and Configuration Description…….……….......................……..……….8 OTDR Physical Description Instrument Enclosure Front Panel Top Plate Home Screen Display Power Requirements Battery Replacement OTDR Operations..........…….....................11 7.1 Key Pad 7.2 Entering OTDR Function 7.3 Icon Definitions 7.4 Main OTDR screen 7.5 OTDR Parameter Settings Screen...
Page 3
Chapter PAGE Table of Contents (Continued) 7.9 Loss Measurements 2 Point Loss dB/km Optical Return Loss Basic Splice Loss Least Square Approximation (LSA) 7.10 Loss Measurement Settings Setting Splice Loss Measurement Areas Splice Loss Adjustment Areas Splice Loss Adjustment Positions LSA Examples 7.11 Event Analysis Enter Event Analysis Threshold Settings Sensitivity Settings Macro bend Analysis Event Analysis Screen Fib-R-Map Event Table Table Definitions Project Management..........................27 Project Management Description Project Management Screen Project Management Menu Project Management File Tab Project Management Edit Tab Project Management View/Help Tabs...
Page 4
Table of Contents Chapter PAGE (Continued) File Management ..............................31 9.1 File Management Description File Management Screen File Management Menu File Management File Tab File Management Edit Tab File Management View/Help Tab 9.2 File Management Operation Saving a Trace Open/View a Trace Dual Trace Delete a File Upload a Trace to CertSoft Rename a Trace File Exit File Management...
Page 5
Table of Contents Chapter PAGE (Continued) 12.4 Video Scope Operations Icon Menu Grading Rings Pass/Fail Label Brightness 12.5 Video Scope Operation Video Probe Tips Viewing/Focusing a Connector Centering a Connector Image Pausing Image Scan Grading Rings Manual Pass/Fail Marking Points of Contamination Auto Pass/Fail Exiting Video Scope Operation 12.6...
Note A special annotation that will assist the user with operational features. Touch Screen Tip These tips will be found throughout the manual to assist with touch screen operation. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 7
Do not use the AC adapter when there are signs of damage to the enclosure or cord ¾ Ensure the correct charger is being used for the local line voltage ¾ Do not use any other charger than the one provided with this instrument. ¾ FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 8
If a live fiber is connected to the OTDR and a scan is attempted, a warning dialog will be displayed as in fig. 2.1. Immediately remove the live fiber from the OTDR and press select to continue testing. Fig 2.1 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
100ns, 300ns, 1us, 3us, 10us or 30us. (Not all pulse widths will be available, only pulse widths appro- priate for the selected range will be presented) ouch the (Range icon) to cycle through the available ranges of 250, 500m, 1km, 2km, 4km, 8km, 16km, 32km, 64km, 100km or 200km. ouch scan or press the Scan button to start a trace. A second touch or press will stop a trace. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Schematic tab in the center/right of the display o save a trace, touch the button and use the left or right buttons to highlight the Quick Save icon and touch it or press . The file will be saved to the active folder. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
We, at Terahertz Technologies Inc., strive to provide premier customer care and technical support by providing timely responsiveness and training. We are proud of our quality and high standards and assure you, our customer, the most user friendly and affordable fiber optic solutions to meet individual needs. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
The instrument’s Model/Part Number, Serial Number and Date of Manufacture are indicated on a label located on the back of the unit. The instrument’s history is filed at the factory by model/part number and serial number. The unit's serial number is also located on the top plate just above the USB Port. Fig 5.1 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Front Panel Protective Rubber Boot Color Touch Screen Key Pad Power Button Fig 6.1 Top panel Unit Serial Number Video Probe Port Power Jack OTDR Port Battery Charge Indicator USB Port Visual Fault Locator Fig 6.2 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
To Prevent Fire or Shock Hazard: Do not install other battery types. Do not use the charger without the batteries installed. Do not expose the battery charger to rain or excessive moisture. Do not use the AC adapter when there are signs of damage to the enclosure or cord. Do not use any charger other than the one provided with this instrument. Any other condition will void the warranty. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Do not use any other charger than the one provided with this instrument. ¾ Failure to follow these caution statements could cause unsafe conditions for the operator and equipment and may void the warranty. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
The up button hides the menu bar or moves cursor up in drop-down menus and the down button opens drop-down menus or move cursor down while in a drop-down menu. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Lambda (Wavelength) Cycles through available wavelengths Averaging Cycles through averaging times Pulse Width Cycles through available pulse widths Range Cycles through available test ranges Index of Refraction Opens the Index of Refraction entry screen FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 18
For finer movement, touch active cursor and drag it to the desired position. Finest movement can be accomplished with the LRUD buttons. Touch Screen Tip All the test parameters may be change by touching the value on the touch screen. A long touch in the test parameter section of the screen will open the parameter settings page. A/B cursor, zoom, loss method and scan may all be operated with the touch screen. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
The digits may now be adjusted with the up and down arrows. Trace Parameter Settings Event Table/Fib-R-Map Sensitivity Settings Pass/Fail Thresholds Settings End of Fiber Pass/Fail Date and Time Threshold Setting Fig 7.4 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 20
Once the desired time is entered, press selected. Exit Parameter Setting Screen Press the menu button or touch Return to exit the parameter page. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Each wavelength stores an independent IOR. Set Wavelength To select the wavelength, press the menu button to display the menu at the top of the display. Use the LRUD buttons to highlight the Lambda (Wavelength) icon, press select to open the page with available wavelengths displayed. Use the touch screen to select a wavelength or move to the next page of available wavelengths. Start Auto Test Press the AutoTest button and the unit will set the average time to short, the appropriate pulse width and range parameters for the fiber under test. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
This will bring up the main OTDR screen with a wavelength indication of 1/# of values selected. Touching scan will cycle through all the wavelengths and save each test. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 23
Touch the desired value to select. the available average times are: RT (Real Time), 15, or 60 seconds and 4, 16 or 40 minutes. Scan Once all the proper settings have been entered, press the Scan button to take the trace. To stop the trace press the scan button again. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
For finer movement, touch active cursor and drag it to the desired position. The finest movement can be accomplished with the left and right buttons. Loss Measurements 2 Point loss 2 Point loss measurements take into account the difference in vertical height between where the A and B cursors cross the fiber trace to determine loss. This information is located on the main OTDR screen in the A-B section of the trace information as shown in Fig. 7.2. This is a quick method of checking the loss between any two points on a fiber optic link. dB/km The dB/Km (dB/Kft or dB/Mi) loss method takes the 2 Point loss in dB and divides by the distance between the cursors in Km (Kilometers), Kf (Kilo feet) or Mi (Miles). For accurate dB/.. loss measurements, the two cursors must be on level back scatter points at least 100m apart (NA will show for distances that are too short). FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Press the menu button at any time to escape the splice lose settings sequence. Note It is necessary to set these areas in clear back scatter to acquire the most accurate measurements, they must not include other events. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 26
Chapter 7 OTDR Operation Splice Loss Adjustment Areas LSA Splice Loss Basic Splice Loss Measurement Areas Measurement Areas Active Splice Zone Cursor Active Splice Zone Cursor LSA Splice Loss Basic Splice Loss Measurements Measurements Fig 7.5 Splice Loss Adjustment Positions Splice Loss Position Splice Loss Position Splice Loss Position Splice Loss Position Fig 7.6 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
The left most green LSA indicator line is not over laying the back scatter of the trace Fig 7.8 Proper LSA Setting: The LSA area and cursor are set properly. The green LSA indica- tor line is over laying the back scatter of the trace properly. Fig 7.9 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Examples of these settings are as follows: Splice: If a 1.00dB splice loss is acceptable, set the threshold to 1.00. If a splice loss of 1.00 is the fail point, set the threshold to 0.99 ORL: If a -50dB reflectance is acceptable, set the threshold to 50. If -50dB is the fail point, set the threshold to 50.5. Link: If a 20dB, link loss is acceptable, set the threshold to 20. If 20db link loss is the fail point, set the threshold to 19.5. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
If event number one was 5 dB at 1550nm and .5 db at 1310nm, the second event will be blank for location, +4.5dB for splice, blank for 2point and DB/KM. The type field will be listed as Macro. To exit dual trace and macro bend analysis, simple start a new scan. Event Analysis Screen Event Table Fib-R-Map Schematic View Trace Parameters Event Map Fig 7.10 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 30
EOF. Event Table Loss Between dB Per/Km Event Location Event Loss in dB Events Between Events Pass/Fail Event Type or Reflectance Value Event Number EOF Data System ORL Link Data Line Link EOF Location Link Loss dB Per/Km Fig 7.12 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 31
This is a selectable threshold for the Pass/Fail feature. The "E" event will display N/A as this is not an event that would have meaning. The "L" line of the table is the link and as such does not have an event (Splice) loss. 2POINT: 2 Point Loss is the loss measured from the end of the dead zone of the previous event to the beginning of current event. The value in the "L" line of the table for this column is the link loss. The loss from the beginning of the trace to the event determined to be the EOF. This is the value used to determine if the fiber link passes or fails it's loss threshold. dB/Km: dB per Kilometer if in KM and DB/KF if in Kilo-Feet and DB/MI if in mile unit of measure. This is the calculated loss per Km, Kf or Mi from the end of the dead zone of the previous event to the beginning of the current event. The value in the "L" line of the table for this column is the dB Per/KM, Kf or Mi for the total link. TYPE: Event Type is the type of event or the ORL measurement. If the event has no reflection, “splc” will be displayed and if the event contains a reflection, the event reflectance value will be displayed. For the last event which is labeled "L" in the # column, the value displayed is the system ORL. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
The menu may be pulled down with the stylus the same as the menu on the main trace screen. Once the file menu is open, it is fully operable with the touch screen function. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Upload is used to transfer files to the CertSoft software. Rename allows the highlighted folder name to be modified while maintaining all settings and associated files. Properties allows the user to edit the properties displayed at the bottom of the project window. Fig 8.4 Project Management View and Help Tabs View offers the user the opportunity to move to the file manager from the projects page. With view highlighted, press select or the down button to open the drop-down menu, with the files tab highlighted, press select to move to the file management page. For onboard help pages, highlight the help tab and press select. Fig 8.5 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
The two plugs in the top right corner of the screen should show as connected and green. Highlight the project to be uploaded, press menu to open file manager menu, use the right button to highlight edit tab, press select or the down button to open the drop down menu and with upload highlighted, press select. The folder with all files will be uploaded to the target folder selected on the computer. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Exit Project Management To exit the project management, press menu, use the left or right buttons to highlight the file tab, press select or the down button to open the drop down menu. Use the down button to highlight Exit and press select. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
The file menu may be pulled down with the stylus the same as the menu on the main trace screen. Once the file menu is open, it is fully operable with the touch screen function. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
With view highlighted, press select or the down arrow to drop down the view tab and press select to move to the file management page. For onboard help pages, highlight the help tab and press select. Fig 9.5 9.2 File management Operation Save a Trace File To save a trace, press menu, highlight the file manager icon and press select. Press select or the down button to open the drop down menu and with save highlighted, press select again. The file naming window will open and the onboard QWERTY key board will be available. Use the touch keyboard to name the file and touch save. The file names are limited to 16 characters. This method should be used to set the base file name for a group of traces. These file names are save exactly as entered. To save files with a chronologically numbered suffix, use the quick save. Quick Save uses the last enter base file name. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Exit File Management To exit the file management, press menu, use the left or right buttons to highlight the file tab, press select or the down button to open the drop down menu. Use the down button to highlight Exit and press select. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Never look directly into the end of a connected fiber optic cable or fiber optic interface of optical test equipment, to do so could expose the user to laser radiation and could result in personal injury. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 40
0.40nm for wavelength, 50 GHz for frequency or 0.5 Ch for channel is displayed. ouch the power value and a numeric keypad will be displayed. Enter the desired power value within the ranges set at the top of the display and touch OK. ouch the Laser icon to energize the laser in a CW mode. Touch the Laser icon a second time to turn off the laser. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Active Output Value Slide High End Channel Scale Slide Low End Slide Snap to Highest Laser Output Snap to Lowest Laser Output Laser Output Laser or Laser Output Down Sweep Start/ Pause Fig 5.2 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 42
Moves the active slide to the highest laser output set by the High Snap to highest laser Output End Slide Laser Output Down Moves the low end slide down the scale Laser or Sweep Start/Pause Start and stops the laser/sweep Laser Output Up Moves the high end slide up the scale FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Use these arrows, or the slide to set the low end laser output. Touch the red high end slide to change the output indicator and focus of the arrows to the high end slide to set the high end laser output. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Even if the indicator shows the laser to be off never looking into the end of a fiber connected to the TLS or directly into the connector port. Do not under any circumstance view or inspect the laser output fibers, connectors or the fiber under test through collimating or focusing optics unless the unit is turned off, batteries are removed and the power adapter is disconnected. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Fault Locator Screen The fault locator screen includes the event table, Fib-R-Map (Schematic View) and event map view of highlighted event. Event Table Fib-R-Map (Schematic View) Event Map Trace with adjustable Parameters splice zones Fig 7.2 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 46
Bar instead of horizontal. Event Table Loss Between dB Per/Km, Kf or Mi Previous and Event Loss in Between Events Event Location Current Events Event Type or Pass/Fail Reflectance Value Event Number EOF Event Information System ORL Link Link dB Per/ Information Link Loss Km, Kf or Mi Fig 7.4 FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 47
This is a selectable threshold for the Pass/Fail feature. The "E" event will display N/A as this is not an event that would have meaning. The "L" line of the table is the link and as such does not have an event (Splice) loss. 2POINT: 2 Point Loss is the loss measured from the end of the dead zone of the previous event to the beginning of current event. The value in the "L" line of the table for this column is the link loss. The loss from the beginning of the trace to the event determined to be the EOF. This is the value used to determine if the fiber link passes or fails it's loss threshold. dB/Km: dB per Kilometer if in KM and DB/KF if in Kilo-Feet and DB/MI if in mile unit of measure. This is the calculated loss per Km, Kf or Mi from the end of the dead zone of the previous event to the beginning of the current event. The value in the "L" line of the table for this column is the dB Per/KM, Kf or Mi for the total link. TYPE: Event Type is the type of event or the ORL measurement. If the event has no reflection, “splc” will be displayed and if the event contains a reflection, the event reflectance value will be displayed. For the last event which is labeled "L" in the # column, the value displayed is the system ORL. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
To View a test in the full OTDR mode, touch an event in the table and the scan will be displayed in trace mode with the active cursor at the selected even. Exit Fault Locator To exit the fault locator mode touch the Home icon. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
12.3 Video Scope File/help Icon Menu The Icons at the top of the display are Home, file save, project management, file management and Help. Home Press the menu button, use the left and right buttons to highlight the home icon and press select to return to the home Screen. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
To set the pass/fail indicator, press the menu button, use the left or right buttons to highlight the pass/fail icon and press select to cycle through the pass/fail states. Touch Screen Tip The touch screen may also be used to make these selections. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
When the touch screen is used to roughly center the connector image, the menu mode is automatically turned off. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 52
The video scope will evaluate the image and grade it with Pass or Fail. Contamination will be displayed on the image. Exiting Video Scope Operation Enter the menu mode, use the left or right buttons to highlight the Home icon and press select. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Page 53
No limit ≤ 3µm Cladding Zone 25µm to 120µm 5 from 2µm to 5µm None > 3µm None > 5µm Adhesive Zone No limit No limit 120µm to 130µm Contact Zone 130µm to 250µm No limit None ≥ 10 µm FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
It also aids in conserving battery life. The usable range for fault location depends on many factors, the type of fiber, the type of cable, the overall loss. To activate the VFL, tap the VFL icon to cycle through the available states of off, Continuous and Modulated. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Memory Type Internal TTI reserves the right to change specifications without notice. Tu n a b l e L a s e r S o u r c e Tu n a b l e L a s e r S o u r c e Frequency 191.4 - 196.25 THz...
Page 56
Emitter Type with interchangeable FC Adapter Wavelength 650nm ±5nm Connector Type 2.5mm Universal Output Power 1mW Max. TTI reserves the right to change specifications without notice. General 4" color touch screen Display 100-240V universal Power Supply / Charger Li-ion 10 hr typ. Battery -20 to 60 C Storage Temperature -10°C to + 40°C...
This warranty does not apply to expendable parts such as batteries or optical panel connectors, nor to any instrument or component which has been subjected to misuse, alteration, or fiber connector damage. It is the customer’s responsibility to understand all the instructions and specifications prior to operating this instrument. This warranty does not extend to any loss or damage consequent to the failure of the warranted product. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Version Control Through a Program of continuous improvement the upgrade of the features and performance of this instrument are an on going process. The instrument firmware version is accessible in the lower right corner of the home screen. The version changes and approximate release dates are as follows. OTDR V1.0.0.2 - 05/2017- Original release V1.0.1.4 - 05/2018- Added clean sweep operation V1.0.11.0 - 12/2021- Adds 50GHz/100GHz-A/100GHz-B switch to wavelength selection screen. FTE-7000A-DWDM OTDR User's Guide Rev D 1/2021...
Need help?
Do you have a question about the FTE-7000A-DWDM and is the answer not in the manual?
Questions and answers