6.2
Disinfecting
6.3
Sterilizing
7
FUNCTION CHECK
32 • seca 750; seca 761/760; seca 762
1. Disinfect the device at regular intervals with a disinfectant suitable for sensi-
tive surfaces and acrylic glass (e.g. 70 % ethanol).
2. Follow the instructions for use for the disinfectant.
3. Disinfect the device as follows:
► Moisten a soft cloth with disinfectant and wipe down the device with it.
► Comply with the intervals, see table:
Interval
Component
every
Before
measurement
Weighing plat-
with direct skin
form
contact
every mea-
After
Weighing plat-
surement with di-
form
rect skin contact
Protecting ring
As required
and viewing
window
The device must not be sterilized.
► Perform a function check before each use.
A complete function check includes:
• A visual inspection for mechanical damage
• A test of the alignment of the device
• A visual and function check of the display elements
• A function check of all the controls shown in the section entitled "Overview"
• A function check of the optional accessories
If you find faults or deviations during the function check, first try to remedy the
fault with the aid of the "Troubleshooting" section in this document.
CAUTION!
Personal injury
If you find faults or deviations during the function check which you are
unable to remedy with the aid of the "Troubleshooting" section in this
document, you must not use the device.
► Have the device repaired by seca Service or an authorized service
partner.
► Follow the section entitled "Servicing" in this document.