ITEM# P2-S4
P2-S4 PORTABLE
OXYGEN CONCENTRATOR

Advertisement

loading
Need help?

Need help?

Do you have a question about the P2-S4 and is the answer not in the manual?

Questions and answers

David Bickel
April 25, 2025

Can I get a hard manual I can read?

User image 680c099401565

Summary of Contents for Rhythm Healthcare P2-S4

  • Page 1 ITEM# P2-S4 P2-S4 PORTABLE OXYGEN CONCENTRATOR...
  • Page 3 TABLE OF CONTENTS Symbols ....................1-2 Intended Use ................... 3 Contraindications ..................4 General Precautions ................4-7 Parts Diagram ..................8 User Interface Buttons and Display Symbols ........9-12 Alerts ....................13-15 Battery Info .................... 16 Power Supply ..................17 Accessories ..................
  • Page 4 Symbol Description A warning indicates that the personal safety of the patient may be involved. WARNING Disregarding a warning could result in significant injury. A caution indicates that a precaution or service procedure must CAUTION be followed. Disregarding a caution could lead to a minor injury or damage to equipment.
  • Page 5 Symbol Description Serial Number This Side Up Fragile Storage Humidity (Non-condensing) Storage Temperature Magnetic Resonance Unsafe The manufacturer of this POC has determined this device conforms to all applicable FAA requirements for POC carriage and use on board aircraft. Service Required Alert...
  • Page 6 It supplies a high concentration of oxygen and is used with a Nasal Cannula to deliver oxygen from the concentrator to the patient. The P2-S4 is a small and portable device that may be used at the home and can be taken with you while performing your daily activities.
  • Page 7 CAUTION Never leave the P2-S4 in high temperatures/high humidity such as in a car in high heat or a bathroom with high humidity. This can damage the device. WARNING Geriatric patients or any other patients unable to communicate discomfort, hear or see alarms while using this device, may require additional monitoring.
  • Page 8 Always consult your health care practitioner before using the POC. WARNING The settings of the P2-S4 might not correspond with a continuous flow of oxygen. WARNING The settings of other models or brands of oxygen therapy equipment do not correspond with the settings of the P2-S4.
  • Page 9 41˚F to 104˚F (5˚C to 40˚C) or a relative humidity above 90% is expected to adversely affect the flow rate and the percentage of oxygen and consequently the quality of the therapy. WARNING The P2-S4 is not intended for use with a tracheostomized patient.
  • Page 10 • Use only with P2-S4 batteries • Use only approved P2-S4 accessories WARNING Remove the battery from the device if the P2-S4 will not be used for an extended period of time. WARNING Device operation exceeding the voltage, breath rate, temperature, humidity and/or altitude values specified may decrease oxygen concentration levels.
  • Page 11 Parts Diagram Alert/Alarm Light Breath Detection Light Service Required Alert Nozzle Mute Audio On/Off Button Button Set Flow Air Outlet Battery Air Intake Battery AC/DC Latch Power Input...
  • Page 12 User Interface Buttons and Display Symbols Display Panel Symbol Description Instructions On/Off Press once to turn On. Button Press and hold for three seconds to turn Off. Press once to toggle between audible and silent mode. The panel will display the appropriate icon to indicate which mode is enabled: Audible mode------ Audio Alarm...
  • Page 13 User Interface Buttons and Display Symbols Home Screen Battery charge level Battery percentage * Estimated time to fully charge the battery (when plugged in) * Battery remaining time to use (when using battery) Device runtime Flow Setting Icon Description Current flow setting (range is 1 to 4) Battery charge level: •...
  • Page 14 User Interface Buttons and Display Symbols Icon Description Powered by AC or DC only 1. Powered by battery only (not plugged in and charging) 2. Battery level percentage and time remaining (hours: minutes). 3. Device On 1. Battery is charging. 2.
  • Page 15 User Interface Buttons and Display Symbols Audible Mode: Below shows the device has detected an active alarm while in Audible Mode. Example below shows "Absence of Breath". Alerts in Silent Mode: The display below shows that the device has detected an active alarm while in Silent Mode. Example below shows "Absence of Breath":...
  • Page 16 Alerts Adapter plug/unplug: An adapter icon displays when adapter is plugged in and disappears when unplugged. If enabled, an alarm will sound. Battery plug/unplug: A battery icon displays when battery is connected and disappears when disconnected. If enabled, an alarm will sound. Alarm audio selection: An alert will indicate when unit is turned On or Off.
  • Page 17 Alerts (continued) Error Message System Process Screen Display Additional Alarm Info Battery Battery is Exhausted Battery cycle>500 Alarm only Exhausted Replace Battery Or battery health<50% Replace Replace Sieve Bed Alarm only Sieve Bed expired Sieve Bed Contact Provider Auto-switch to battery Low Input Low Input Voltage until the adapter input...
  • Page 18 Alerts (continued) Error Message System Process Screen Display Additional Alarm Info Auto-shut down Sieve Bed Fail Sieve Bed failed or Sieve Bed Fail after 10 seconds Contact Provider is invalid Auto-shut down Compressor Fail Compressor Fail Compressor failed after 10 seconds Contact Provider Auto-shut down Valve Check Fail...
  • Page 19 P2-S4 comes standard with a single battery. Double and triple batteries are optional and can be purchased from your provider. Ensure the battery is installed into the P2-S4. Use the AC power supply to recharge the battery. P2-S4 with Single Battery: Can provide up to 2.5 hours of operation.
  • Page 20 Power Supply AC Power Supply # S3AC The AC Power Supply is used to power the P2-S4 from an electrical outlet. When using the AC charger, the power supply automatically adapts to input voltages from 100V to 240V (50-60HZ) allowing it to be used with most power sources throughout the world.
  • Page 21 CAUTION Select only FDA approved Nasal Cannula (e.g., P2NC-1) Carry Bag # S3CB The P2-S4 Carry Bag allows you to go out for daily activities. See the Carry Bag Instructions for detailed Use. P2-S4 Accessories Item Number...
  • Page 22 Operating Instructions General Operation 1. Find a well ventilated location to place the P2-S4. Ensure that the unit is turned off. Ensure the Intake and exhaust has clear access. The P2-S4 should be placed where alarms can be heard. WARNING Do not use P2-S4 in the presence of flammable anesthetics, detergents, or other chemical vapors.
  • Page 23 Slide the battery into place until the latch returns to the upper position. There will be an audible click when the battery is in position. 3. Connect AC or DC power supply to P2-S4. The green LED on the power adapter will turn on and there will be an audible sound when power is connected.
  • Page 24 Operating Instructions General Operation (continued) 4. Place the P2-S4 into Carry Bag. Position exhaust vents away from the user. 5. Connect the Nasal Cannula to the nozzle fitting. Nozzle fitting is located on the top side of the P2-S4 near the Pre-Filter Cover.
  • Page 25 Operating Instructions General Operation (continued) 7. Set flow rate prescribed by your provider. Press the ± setting button to adjust the P2-S4 to the desired flow rate. The current setting can be viewed on the display from 1 to 4. CAUTION Ensure the power is in a well-ventilated place.
  • Page 26 Troubleshooting The table below lists some common problems and solutions. If you can't resolve a problem, please contact your provider. Problem Possible Cause Recommended Solution Battery is not Remove the battery and reinstall installed correctly it correctly. Use the AC or DC power supply to operate the device (with the battery inserted) to recharge the Battery is depleted...
  • Page 27 Nasal Cannula should be FDA approved and the length should be less than 25ft (7.6m). Filter Cleaning and Replacement Filters are designed for adequate air flow through the device at the front of the P2-S4. Pre-Filter Cover # S4PFCR The Pre-Filter Covers must be cleaned weekly to ensure adequate air flow. Clean with a mild liquid detergent and water.
  • Page 28 The Cotton Intake Filter cannot be cleaned, it must be replaced. Battery Care and Maintenance The P2-S4 Lithium-ion Battery requires special care to ensure proper performance and long life. Only use P2-S4 batteries (Item #s: S3BY001S, S3BY001D, S3BY001T) with your concentrator. CAUTION Keep liquids away from the battery.
  • Page 29 We suggest that the room temperature should not exceed 75°F (24°C) when the battery is being charged. Battery Time Remaining Clock The P2-S4 continuously displays the battery time remaining. This displayed time is only an estimate and the actual time remaining may vary from this value. CAUTION Store battery in a cool dry place with a charge of 40-50%.
  • Page 30 Maintenance and cleaning of P2-S4 Disposal of Equipment and Accessories Follow your local governing ordinances for disposal and recycling of the P2-S4 accessories. The battery contains lithium-ion cells and should be recycled and must not be incinerated. P2-S4 Replacement Items...
  • Page 31 Available Mode Triggered Pulse Delivery Mode * 1 According to ISO 80601-2-69 the Maximum system sound pressure level of the P2-S4 is 44 dB (A) at Setting 2 and 49 dB (A) at Setting 4. * 2 Data according to test method 14-1 03/2007 MDS-Hi.
  • Page 32 System Specifications Specifications (continued) AC Input: 100 to 240 VAC 50 to 60 Hz DC Input: 11-16 VDC DC Output: 19V 6.3A Power: AC Power Supply Voltage: 14.4 VDC DC Power Supply Single Battery Rated Capacity: 3,400 mAh Rechargeable Battery Double Battery Rated Capacity: 6,800 mAh Triple Battery Rated Capacity: 10,200 mAh Single Battery...
  • Page 33 System Specifications Standards Compliance The device is designed to conform to the following standards: - IEC 60601-1-2, 2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - IEC 60601 - 1: Medical Electrical Equipment - part 1: General Requirements for Basic safety &...
  • Page 34 System Specifications EMC Information The device has been designed to meet EMC standards throughout its Service Life. Guidance and Manufacturer's Declaration - Electromagnetic Immunity: The Concentrator is intended for use in the electromagnetic environment specified below. The user of the Concentrator should make sure it is used in such an environment. Compliance Electromagnetic Immunity Test...
  • Page 35 System Specifications EMC Information (continued) Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications equipment IEC 61000-4-6 150 kHz to 80 MHz 150 kHz to 80 MHz should not be used closer (to any part of the device, Radiated RF 6 Vrms 6 Vrms...
  • Page 36 System Specifications Guidance and Manufacturer's Declaration - Electromagnetic Emissions: The P2-S4 is intended for use in the electromagnetic environment specified below. The user of the P2-S4 should ensure it is used in such an environment. Electromagnetic Environment - Emission Test...
  • Page 37 System Specifications Essential Performance List The P2-S4 is intended for use in the electromagnetic environment specified below. The user of the P2-S4 should ensure it is used in such an environment. Essential Performance Relevant Requirements or Condition Accuracy of concentration...
  • Page 38 This warranty does not cover damage or injury whether to P2-S4 or to personal property or persons caused by accident, misuse, abuse, negligence, failure to install in accordance with Rhythm’s installation instructions, failure to operate under conditions of normal use...
  • Page 39 ÍNDICE Símbolos ..................36-37 Uso Previsto ..................38 Contraindicaciones ................39 Precauciones de Seguridad .............. 39-42 Diagrama de Componentes ..............43 Botones de la Interfaz de Usuario e Íconos de Pantalla ....... 44-47 Alertas ....................48-50 Detalles de la Batería ................51 Fuente de Energía .................
  • Page 40 Símbolo Descripción Una advertencia indica que la seguridad personal del paciente podría estar en ADVERTENCIA riesgo. Ignorar una advertencia podría causar lesiones graves. La señal de precaución indica que se debe seguir un procedimiento o PRECAUCIÓN precaución de servicio. Ignorar una precaución podría provocar una lesión leve o daños al equipo.
  • Page 41 Símbolo Descripción Número de serie Este lado arriba Frágil Humedad de almacenamiento (no condensada) Temperatura de almacenamiento No es seguro para la resonancia magnética El fabricante de este concentrador de oxígeno portátil (COP) certifica que este dispositivo cumple con todos los requisitos aplicables de la FAA para su transporte y uso a bordo de aeronaves.
  • Page 42 Proporciona una alta concentración de oxígeno y se utiliza con una cánula nasal para suministrar oxígeno desde el concentrador al paciente. El P2-S4 es un dispositivo compacto y portátil, adecuado para usar en el hogar y llevarlo consigo en sus actividades diarias.
  • Page 43 PRECAUCIÓN Nunca deje el P2-S4 en lugares con altas temperaturas o alta humedad, como dentro de un automóvil en un día caluroso o en un baño con alta humedad. Esto puede dañar el dispositivo.
  • Page 44 COP. ADVERTENCIA Los ajustes del P2-S4 pueden no corresponder con un flujo continuo de oxígeno. ADVERTENCIA Los ajustes de otros modelos o marcas de equipos de terapia de oxígeno no corresponden con los ajustes del P2-S4.
  • Page 45 5˚C a 40˚C (41˚F a 104˚F), o con una humedad relativa superior al 90 %, puede afectar negativamente el flujo y la concentración de oxígeno, comprometiendo la calidad de la terapia. ADVERTENCIA El P2-S4 no está diseñado para su uso en pacientes con traqueotomía.
  • Page 46 • Utilice únicamente las baterías del P2-S4 • Utilice únicamente accesorios aprobados para el P2-S4 ADVERTENCIA Retire la batería del dispositivo si no va a utilizar el P2-S4 durante un periodo prolongado. ADVERTENCIA Operar el dispositivo con valores que superen los especificados para el voltaje, frecuencia respiratoria, temperatura, humedad y/o altitud puede reducir los niveles de concentración...
  • Page 47 Diagrama de Componentes Luz de Alerta/Alarma Luz de Detección de Alerta de Servicio Respiración Requerido Boquilla Botón para Botón de Silenciar Audio Encendido/ Apagado Configurar Flujo Puerta de la Entrada de Aire Entrada de Salida de Aire Batería Batería Corriente CA/CC...
  • Page 48 Botones de la Interfaz de Usuario e Íconos de Pantalla Panel de Pantalla Símbolo Descripción Instrucciones Botón de Encendido/ Pulse una vez para encender. Apagado Mantenga pulsado durante tres segundos para apagar. Pulse una vez para alternar entre modo con sonido y modo silencioso.
  • Page 49 Botones de la Interfaz de Usuario e Íconos de Pantalla Pantalla principal Nivel de carga de la batería Porcentaje de batería * Tiempo estimado para cargar completamente la batería (cuando está enchufado) * Tiempo restante de batería (cuando se usa la batería) Tiempo de funcionamiento del dispositivo Ajuste de flujo Ícono...
  • Page 50 Botones de la Interfaz de Usuario e Íconos de Pantalla Ícono Descripción Alimentado solo por CA o CC 1. Alimentado solo por batería (no enchufado y cargando) 2. Porcentaje de nivel de batería y tiempo restante (horas: minutos). 3. Dispositivo encendido 1.
  • Page 51 Botones de la Interfaz de Usuario e Íconos de Pantalla Modo con Sonido: Lo siguiente muestra que el dispositivo ha detectado una alarma activa mientras está en modo con sonido. El ejemplo a continuación muestra "Falta de respiración": Alertas en Modo Silencioso: La pantalla a continuación indica que el dispositivo ha detectado una alarma activa mientras está...
  • Page 52 Alertas Conexión/desconexión del adaptador: Un ícono del adaptador aparece cuando el adaptador está conectado y desaparece cuando se desconecta. Si está activado, sonará una alarma. Conexión/desconexión de la batería: Un ícono de la batería aparece cuando la batería está conectada y desaparece cuando se desconecta.
  • Page 53 Alertas (Continuación) Proceso del Pantalla de Información Adicional Mensaje de Error Sistema Visualización de la Alarma Batería agotada Ciclo de la batería>500 O Batería agotada Solo alarma Reemplace la batería Salud de la batería<50% Sustituir el lecho del Remplazar el lecho tamiz Solo alarma Lecho del tamiz vencido...
  • Page 54 Alertas (Continuación) Proceso del Pantalla de Información Adicional Mensaje de Error Sistema Visualización de la Alarma El dispositivo se Falla del lecho del Falla del lecho del tamiz El lecho del tamiz falló o apaga luego de 10 tamiz Contacte con el proveedor no es válido segundos El dispositivo se...
  • Page 55 El P2-S4 viene de serie con una batería simple. Las baterías dobles y triples son opcionales y se pueden adquirir con su proveedor. Asegúrese de que la batería esté instalada en el P2-S4. Utilice la fuente de alimentación de CA para recargar la batería.
  • Page 56 Fuente de Energía Fuente de Alimentación de CA # S3AC La Fuente de Alimentación de CA se utiliza para alimentar el P2-S4 en una toma de corriente eléctrica. Cuando se usa el cargador de CA, la fuente de alimentación se adapta automáticamente a voltajes de entrada de 100V a 240V (50-60 Hz), lo que permite su uso...
  • Page 57 Seleccione únicamente cánulas nasales aprobadas por la FDA (por ejemplo, P2NC-1). Bolsa de Transporte # S3CB La Bolsa de Transporte P2-S4 le permite llevar el dispositivo para actividades diarias. Consulte las instrucciones de la bolsa de transporte para un uso detallado.
  • Page 58 útil del lecho de tamiz. El tiempo máximo de operación del P2-S4 es de 24 horas al día, siempre que el dispositivo tenga suficiente energía y tanto el dispositivo como los accesorios estén dentro de su vida útil esperada.
  • Page 59 Deslice la batería en su lugar hasta que el pestillo regrese a la posición superior. Escuchará un clic cuando la batería esté correctamente instalada. 3. Conecte la fuente de alimentación de CA o CC al P2-S4. El LED verde en el adaptador de corriente se encenderá y se escuchará un sonido cuando se conecte la energía.
  • Page 60 Instrucciones de Operación Operación General (Continuación) 4. Coloque el P2-S4 en la bolsa de transporte. Coloque las rejillas de salida de aire alejadas del usuario. 5. Conecte la cánula nasal al accesorio de la boquilla. El accesorio de la boquilla está ubicado en la parte superior del P2-S4 cerca de la cubierta del prefiltro.
  • Page 61 Operación General (Continuación) 7. Ajuste el flujo según lo prescrito por su proveedor. Pulse el botón de ajuste ± para ajustar el P2-S4 al flujo deseado. El ajuste actual se puede ver en la pantalla, con un rango de 1 a 4.
  • Page 62 Solución de Problemas La tabla a continuación enumera algunos problemas comunes y sus soluciones. Si no puede resolver un problema, contacte a su proveedor. Problema Causa Posible Solución Recomendada La batería no está instalada Retire la batería e instálela correctamente correctamente.
  • Page 63 Mantenimiento y Reemplazo del Filtro Los filtros están diseñados para permitir un flujo de aire adecuado a través del dispositivo en la parte frontal del P2-S4. Cubierta del Prefiltro # S4PFCR Las cubiertas del prefiltro deben limpiarse semanalmente para asegurar un flujo de aire adecuado.
  • Page 64 Mantenimiento de la Batería La Batería de ion de litio del P2-S4 requiere cuidados especiales para garantizar un rendimiento adecuado y una larga vida útil. Use únicamente baterías del P2-S4 (N.º de artículos: S3BY001S, S3BY001D, S3BY001T) con su concentrador.
  • Page 65 Sugerimos que la temperatura ambiente no exceda los 24°C (75°F) mientras la batería se está cargando. Reloj de Tiempo Restante de la Batería El P2-S4 muestra continuamente el tiempo restante de la batería. Este tiempo mostrado es solo una estimación, y el tiempo real restante puede variar de este valor. PRECAUCIÓN Almacene la batería en un lugar fresco y seco con una carga del 40-50 %.
  • Page 66 Mantenimiento y Limpieza del P2-S4 Instrucciones Para Desechar el Equipo y Accesorios Siga las ordenanzas locales para la eliminación y reciclaje de los accesorios del P2-S4. La batería contiene celdas de litio-ion y debe reciclarse; no debe incinerarse. Artículos de Reemplazo del P2-S4 Número de Parte...
  • Page 67 Modo disponible Modo de suministro de pulso activado * 1 Según ISO 80601-2-69, el nivel máximo de presión sonora del sistema del P2-S4 es de 44 dB(A) en el ajuste 2 y 49 dB(A) en el ajuste 4. * 2 Datos según el método de prueba 14-1 03/2007 MDS-Hi.
  • Page 68 Especificaciones del Sistema Especificaciones (Continuación) Potencia: Entrada de CA: 100 a 240 V CA 50 a 60 Hz Fuente de Entrada de CC: 11-16 V CC Salida de CC: 19V 6,3A alimentación de CA Voltaje: 14,4 V CC Fuente de Capacidad nominal de la batería simple: 3.400 mAh alimentación de CC Capacidad nominal de la batería doble: 6.800 mAh...
  • Page 69 Especificaciones del Sistema Cumplimiento de Normas El dispositivo está diseñado para cumplir con las siguientes normas: - IEC 60601-1-2, 2014, Equipo médico eléctrico - Parte 1-2: Requisitos generales para la seguridad básica y el rendimiento esencial - Norma colateral: Perturbaciones electromagnéticas - Requisitos y pruebas - IEC 60601-1: Equipo médico eléctrico - Parte 1: Requisitos generales para seguridad básica y rendimiento esencial...
  • Page 70 Especificaciones del Sistema Detalles de EMC El dispositivo ha sido diseñado para cumplir con los estándares EMC durante toda su vida útil. Guía y declaración del fabricante - inmunidad electromagnética: El concentrador está destinado para su uso en el entorno electromagnético especificado a continuación.
  • Page 71 Especificaciones del Sistema Detalles de EMC (Continuación) RF Conducida 3 Vrms 3 Vrms El equipo de comunicaciones RF portátil IEC 61000-4-6 150 kHz a 80 MHz 150 kHz a 80 MHz y móvil no debe usarse más cerca del dispositivo RF radiada 6 Vrms 6 Vrms...
  • Page 72 Especificaciones del Sistema Guía y Declaración del Fabricante - Emisiones Electromagnéticas: El P2-S4 está diseñado para su uso en el entorno electromagnético especificado a continuación. El usuario del P2-S4 debe asegurarse de que se utilice en dicho entorno. Prueba de Emisión Cumplimiento Entorno Electromagnético - Guía...
  • Page 73 Especificaciones del Sistema Criterios Esenciales de Rendimiento El P2-S4 está diseñado para su uso en el entorno electromagnético especificado a continuación. El usuario del P2-S4 debe asegurarse de que se utilice en dicho entorno. Rendimiento Esencial Requisitos o Condición Relevante Dentro de 93% +3/-6% en todos los niveles Precisión de la concentración...
  • Page 74 P2-S4. Esta garantía no cubre daños que puedan ocurrir durante el envío. Esta garantía no se aplica a ningún producto o parte individual de un producto que haya sido reparado o alterado por alguien que no sea Rhythm o un centro de servicio autorizado por Rhythm.
  • Page 76 Rhythm Healthcare Email: contactus@rhythmhc.com Website: www.rhythmhc.com V3: 09/14/24...