Multi-phase Programmable AC/DC
Power Source
ASR-6000 Parallel Models Series
USER MANUAL
Rev. A
ISO-9001 CERTIFIED MANUFACTURER

Advertisement

loading
Need help?

Need help?

Do you have a question about the ASR-6000 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for GW Instek ASR-6000 Series

  • Page 1 Multi-phase Programmable AC/DC Power Source ASR-6000 Parallel Models Series USER MANUAL Rev. A ISO-9001 CERTIFIED MANUFACTURER...
  • Page 2 This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company. The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
  • Page 3: Table Of Contents

    Table of Contents Table of Contents SAFETY INSTRUCTIONS ............ 4 GETTING STARTED ............8 ASR-6000 Parallel Models Series Overview ....9 Appearance .............. 16 OPERATION ..............25 Set Up ..............26 Input Terminal Connection ........28 Output Terminal Connection ........31 APPENDIX ..............
  • Page 4: Safety Instructions

    ASR-6000 Parallel Models Series User Manual AFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition. Safety Symbols These safety symbols may appear in this manual or on the instrument.
  • Page 5 SAFETY INSTRUCTIONS Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Safety Guidelines Do not place any heavy object on the ASR-6000. General Guideline Avoid severe impact or rough handling that leads CAUTION to damaging the ASR-6000.
  • Page 6 ASR-6000 Parallel Models Series User Manual AC Input voltage range: Power Supply 200 Vac to 240 Vac (3P3W) WARNING 380 Vac to 460 Vac (3P4W) Frequency: 47 ~ 63 Hz To avoid electrical shock connect the protective grounding conductor of the AC power cord to an earth ground.
  • Page 7: Operation Environment

    SAFETY INSTRUCTIONS Location: Indoor, no direct sunlight, dust free, Operation almost non-conductive pollution (Note below) Environment Relative Humidity: 20%~ 80%, no condensation Altitude: < 2000m Temperature: 0°C to 40°C (Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The ASR-6000 falls under degree 2. Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
  • Page 8: Etting Started

    ASR-6000 Parallel Models Series User Manual ETTING STARTED This chapter describes the ASR-6000 parallel model series in a nutshell, including its main features, operating area, accessories and front with rear panel introduction. ASR-6000 Parallel ASR-6000 Parallel ASR-6000 Parallel Model Series in 15u Model Series in 19u Model Series in 23u ASR-6000 Parallel Models Series Overview ......
  • Page 9: Asr-6000 Parallel Models Series Overview

    GETTING STARTED ASR-6000 Parallel Models Series Overview Series lineup The ASR-6000 parallel models series consists of 5 models, which differ in various capacity. Note that throughout the user manual, the term “ASR-6000” refers to any of the models, unless stated otherwise. 1P Output Condition Model Name Power Rating Max.
  • Page 10: Operating Area

    ASR-6000 Parallel Models Series User Manual Operating Area...
  • Page 11 GETTING STARTED...
  • Page 12 ASR-6000 Parallel Models Series User Manual...
  • Page 13 GETTING STARTED...
  • Page 14 ASR-6000 Parallel Models Series User Manual...
  • Page 15: Accessories

    GETTING STARTED Accessories Before using the ASR-6000 parallel models, check the package contents to make sure all the standard accessories are included. Standard Part number Description Accessories 82GW1SAFE0M*1 Safety guide 62SR-6KDSC201 Input terminal cover 62SR-6KDSC301 62SR-6KDSC501 Output terminal cover 62SR-6KDSC601 GTL-246 USB cable (USB 2.0 Type A - Type B cable, approx.
  • Page 16: Appearance

    ASR-6000 Parallel Models Series User Manual Appearance Front Panel...
  • Page 17 GETTING STARTED Item Index Description Power switch button USB interface connector (A Type) LCD screen Function keys (blue zone) Menu key Test key Preset key Scroll wheel Range key/Output mode key Arrow keys Output key Shift key Cancel key Enter key Irms/IPK-Limit button Lock/Unlock button F/F-Limit button...
  • Page 18 ASR-6000 Parallel Models Series User Manual Item Description Power Switch Turn on the mains power USB A Port The USB port is used for data transfers and upgrading software. Also, it is available for screenshot hardcopy. It supports FAT32 format with maximum 32G storage. LCD Screen Displays the setting and measured values or menu system...
  • Page 19 GETTING STARTED Range Key Switches between the 100V, 200V and AUTO ranges Output Mode Selects between the AC+DC-INT, AC-INT, DC-INT, AC+DC-EXT, AC-EXT, AC+DC-ADD, AC-ADD, AC+DC-Sync, AC-Sync and AC- VCA modes. Scroll Wheel Used to navigate menu items or for increment/decrement values one step at a time.
  • Page 20 ASR-6000 Parallel Models Series User Manual IPK-Limit + Used to set the peak output current limit value. Lock/Unlock Key Used to lock or unlock the front panel keys except output key. Simply press to lock, whilst long press to unlock. Used for setting the output frequency (DC mode N/A).
  • Page 21 GETTING STARTED Output Waveform + Selects between the Sine, Square, Triangle and ARB 1~253 waveforms (not available for DC-INT, AC+DC-EXT and AC-EXT). + Switches operation back to local mode Local Mode from remote mode. IPK CLR + Used to clear peak output current value. ALM CLR + Clears alarms.
  • Page 22: Rear Panel

    ASR-6000 Parallel Models Series User Manual Rear Panel...
  • Page 23 GETTING STARTED Item Index Description Output terminal AC power input terminal Remote sensing input terminal External I/O connector External IN/OUT connection in parallel function RS232 connector Ethernet (LAN) connector USB interface connector (B Type) Optional interface Slot ▪ GPIB card (ASR-003) ▪...
  • Page 24 ASR-6000 Parallel Models Series User Manual External Control Used to control ASR-6000 externally I/O Connector by using the logic signal and monitor Sequence function status. External IN/OUT The IN (Slave) and OUT (Master) Connection in ports are used for connection with Parallel Function external unit in parallel function.
  • Page 25: Operation

    OPERATION PERATION Set Up ................26 Power Up and Procedure .......... 26 Input Terminal Connection ..........28 Input Terminal 3P3W Connection ......29 Input Terminal 3P4W Connection ......30 Output Terminal Connection ........... 31 1P2W Output Connection ......... 31 1P3W Output Connection ......... 33 3P4W Output Connection .........
  • Page 26: Set Up

    ASR-6000 Parallel Models Series User Manual Set Up We take the illustration of 3P4W Input Connection here for example. Please refer to page 28 of the Input Terminal Connection chapter for the detailed information covering the 2 different connection methods. Power Up and Procedure Connect the AC power cords to the...
  • Page 27 OPERATION The AC power cords of 3P4W input are connected with the AC input terminals equipped with protective lid completely. Turn ON circuit breakers in the sequence of MASTER followed by SLAVE. In the case of multiple SLAVE units in parallel connection, turn ON each circuit breaker of SLAVE in proper sequence, e.g., SLAVE-1...
  • Page 28: Input Terminal Connection

    ASR-6000 Parallel Models Series User Manual ▪ If the warning message of “Parallel Communication CAUTION Error” appears in the screen display, turn Off both POWER key and circuit breakers followed by repeating the appropriate power up procedure above. ▪ Contact local dealer in your region if the warning message of “Parallel Communication Error”...
  • Page 29: Input Terminal 3P3W Connection

    OPERATION Input Terminal 3P3W Connection Connect the AC power cords to the AC input terminals. Red → L2 ✓ Green → L1 ✓ Yellow → L3 ✓ ▪ Power input cords are not included in this Note product. ▪ The input & output terminals necessitate connectivity through ring-type connectors.
  • Page 30: Input Terminal 3P4W Connection

    ASR-6000 Parallel Models Series User Manual Input Terminal 3P4W Connection Connect the AC power cords to the AC input terminals. Red → L3 ✓ Green → L2 ✓ Yellow → L1 ✓ Blue → Neutral ✓ ▪ Power input cords are not included in this Note product.
  • Page 31: Output Terminal Connection

    OPERATION Output Terminal Connection The output terminal can output power in three Background modes: 1P2W, 1P3W and 3P4W. Select applicable output mode, via panel configurations, in accordance with varied applications. Be aware of dangerous voltages. Ensure that the WARNING power to the instrument is disabled before handling the power supply output terminals.
  • Page 32 ASR-6000 Parallel Models Series User Manual The protective cover of power output terminals is well installed and fixed on the rear panel. Install the protective lid of power output terminals followed by fastening the single screw to fix the lid firmly into place. The AC power cords of 1P2W output are connected with the AC output terminals equipped with...
  • Page 33: 1P3W Output Connection

    OPERATION 1P3W Output Connection Disconnect the ASR unit from the mains power socket and turn the power switch off before wires connection. Connect the output wires to the AC output terminals as follows: ✓ Yellow → Line (L1) ✓ Green → Line (L2) ✓...
  • Page 34: 3P4W Output Connection

    ASR-6000 Parallel Models Series User Manual 3P4W Output Connection Disconnect the ASR unit from the mains power socket and turn the power switch off before wires connection. Connect the output wires to the AC output terminals as follows: ✓ Yellow → Line (L1) ✓...
  • Page 35: Remote Sensing, Ext I/O And Interface Connection

    OPERATION ▪ The input & output terminals necessitate Note connectivity through ring-type connectors. The diagram is only for reference on wiring method. WARNING Please proceed to wiring in accordance with the color definitions in your local country. Remote Sensing, EXT I/O and Interface Connection Remote sense is used to compensate for the Remote Sensing voltage drop seen across load cables due to...
  • Page 36 ASR-6000 Parallel Models Series User Manual PPENDIX Firmware Update ............37 Function Difference Table ..........40 A Comparison between Stand Alone Type and Parallel Type ... 40 Factory Default Settings ..........41 Error Messages & Messages ..........45 Specifications ..............50 Electrical specifications –...
  • Page 37: Appendix

    APPENDIX Firmware Update The ASR series firmware can be upgraded using Background the USB A-type port on the front panel. See your local distributor or the GWINSTEK website for the latest firmware information. • Both Master and Slave ASR units are required to be Note plugged in USB flash drives with the identical firmware version in order to complete update...
  • Page 38 ASR-6000 Parallel Models Series User Manual 3. The removable cover is removed accordingly. 4. Pull out the plastic frame from ASR Slave unit. 5. The plastic frame was removed and thus the USB A-type port of Slave unit appears. 6. Repeat the previous step 1 to step 5 for each connected ASR Slave unit.
  • Page 39 APPENDIX 9. Use the scroll wheel to go to item 11, Special Function and press Enter. 10. Key in the password when prompted and then press Enter. The password is “5004”. 11. Go to Item 1, Update Firmware and press Enter. Update Firmware Exit [F4] 12.
  • Page 40: Function Difference Table

    ASR-6000 Parallel Models Series User Manual Function Difference Table A Comparison between Stand Alone Type and Parallel Type The difference functions Item Description Stand Alone Type Parallel Type Fast, Medium(default), V Response Medium(default), Slow Slow Output Impedance Setting External Parallel 2~3 units flexible Fixed Operation...
  • Page 41: Factory Default Settings

    APPENDIX Factory Default Settings The following default settings are the factory configuration settings for the ASR-6000 series. For details on how to return to the factory default settings, refer to the User Manual of ASR-6000. Continuous Mode ASR-6450-09 ASR-6600-12 ASR-6450-13.5...
  • Page 42 ASR-6000 Parallel Models Series User Manual MISC Configuration ASR-6450-09 ASR-6600-12 ASR-6450-13.5 ASR-6600-18 ASR-6600-24 T peak, hold(msec) Phase Mode Unbalance Peak CLR Power ON Buzzer Remote Sense V Response Medium Output Relay Enable Measure Unit THD Format External Control V Unit (TRI, ARB) Set Change Phase Monitor Output1 L1 Voltage...
  • Page 43 APPENDIX RS232C ASR-6450-09 ASR-6600-12 ASR-6450-13.5 ASR-6600-18 ASR-6600-24 Baudrate 9600 Databits 8bits Parity None Stopbits 1bit GPIB ASR-6450-09 ASR-6600-12 ASR-6450-13.5 ASR-6600-18 ASR-6600-24 Address CAN BUS ASR-6450-09 ASR-6600-12 ASR-6450-13.5 ASR-6600-18 ASR-6600-24 Baudrate 125K Node ID DeviceNet ASR-6450-09 ASR-6600-12 ASR-6450-13.5 ASR-6600-18 ASR-6600-24 Baudrate 125K MAC ID...
  • Page 44 ASR-6000 Parallel Models Series User Manual Sequence Mode ASR-6450-09 ASR-6600-12 ASR-6450-13.5 ASR-6600-18 ASR-6600-24 Step Time 0.1000 S Jump to Jump Cnt Branch1 Branch2 Term CONTI Sync Code Item L1 L2 L3 L1 L2 L3 L1 L2 L3 L1 L2 L3 L1 L2 L3 0.00,CT 0.00,CT...
  • Page 45: Error Messages & Messages

    APPENDIX Error Messages & Messages The following error messages or messages may appear on the ASR- 6000 screen display during varied operations. Normal Messages Description Protection type Keys Locked All of keys are locked, except Display Message output key, long push “Lock” to Only disable Keys Locked Keys Unlocked...
  • Page 46 ASR-6000 Parallel Models Series User Manual Rear USB Port Connected Rear USB port connected to PC Display Message To PC Only Rear USB Port Rear USB port disconnected Display Message Disconnected From PC from PC Only Reseting... Ready For Recall Factory Default Display Message Only Failed Factory Default Recall Factory Default Failed...
  • Page 47 APPENDIX [Filename] Recall Fail(File Recall file fail message(file Data Display Message Data Error) error(Data out of Range) ) Only Preset M# Deleted Preset M0~M9 Deleted Display Message Only ARB# Deleted ARB1~ARB253 Deleted Display Message Only Save All Data Ready to save all data Display Message (Preset0~9 + SEQ0~9 + SIM0~9 + Only...
  • Page 48 ASR-6000 Parallel Models Series User Manual Saved To ARB# Saved to ARB1 ~ ARB253 Display Message Only Saved To ARB#,V-Limit Saved to ARB1 ~ ARB253,V-Limit Display Message Invalid Invalid Only Saved To ARB#,V-Limit & Saved to ARB1 ~ ARB253,V-Limit Display Message Freq Invalid and Freq Invalid Only...
  • Page 49 APPENDIX Saved to SIM# Saved toSIM0 ~ SIM9 Display Message Only Save Fail! SEQ0 ~ SEQ9 or SIM0 ~ SIM9 Display Message save fail! Only Sequence preparation... Sequence preparation, please Display Message wait some time Only Sequence is ready. Sequence is ready. Display Message Only Simulation preparation...
  • Page 50: Specifications

    ASR-6000 Parallel Models Series User Manual Specifications The specifications apply when the ASR-6000 is powered on for at least 30 minutes. Electrical specifications – ASR-6450-09/ ASR-6600-12 Model ASR-6450-09 ASR-6600-12 Input ratings Power type Three-phase Four-wire Y connection 200 Vac to 240 Vac (Phase Voltage) Voltage range 380 Vac to 460 Vac (Line Voltage) Frequency range...
  • Page 51 APPENDIX 1P3W: 0.00 V 1P3W: 0.00 V to 350.0 V / to 350.0 V / 0.00 V to 0.00 V to 700.0 V 700.0 V 3P4W: 0.00 V 3P4W: 0.00 V to 303.1 V / to 303.1 V / Line voltage setting 0.00 V to 0.00 V to range...
  • Page 52 ASR-6000 Parallel Models Series User Manual Model ASR-6450-09 ASR-6600-12 Output Stability, Total Harmonic Distortion, Output voltage rising time and Ripple noise Line regulation ±0.1% or less (Phase voltage) ±0.5 V / ±1.0 V (phase voltage, 0 to 100%, via output terminal) Load regulation <0.3 % @1Hz to 100Hz, <0.5 % @100.1 Hz to 500 Hz, <1 % Distortion of Output...
  • Page 53 APPENDIX voltage setting. 20) For an output voltage of 100 V / 200 V, a load power factor of 1, with respect to stepwise change from an output current of 0 A to the maximum current (or its reverse). 10% ~ 90% of output voltage. 21) For 5 Hz to 1 MHz components in DC mode using the output terminal on the rear panel.
  • Page 54 ASR-6000 Parallel Models Series User Manual 50/60 Hz only) Range Up to 100th order of the fundamental wave Harmonic current Full Scale 126 A / 63 A, 100% 42 A / 21 A, 100% Effective Resolution 0.01 A / 0.1 A, 0.1% value (rms) Up to 20th: ±...
  • Page 55: General Specifications - Asr-6450-09/ Asr-6600-12

    APPENDIX General Specifications – ASR-6450-09/ ASR-6600-12 Model ASR-6450-09 ASR-6600-12 Type A: Host, Type B: Slave, Speed: 2.0, USB- CDC / USB-TMC MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask Standard External Signal Input Interface External External Control I/O...
  • Page 56: Electrical Specifications - Asr-6450-13.5/Asr-6600-18

    ASR-6000 Parallel Models Series User Manual Electrical specifications – ASR-6450-13.5/ASR-6600-18 Model ASR-6450-13.5 ASR-6600-18 Input ratings Power type Three-phase Four-wire Y connection 200 Vac to 240 Vac (Phase Voltage) Voltage range 380 Vac to 460 Vac (Line Voltage) Frequency range 47 Hz to 63 Hz Power factor 0.95 or higher (typ.) Efficiency...
  • Page 57 APPENDIX Setting AC Mode: 15.00 Hz to 1000.0 Hz, AC+DC Mode: 1.00 Hz to range 1000.0 Hz, Setting resolution: 0.01 Hz / 0.1 Hz Frequency Accuracy ± 0.01% of set Stability ± 0.005% Output on phase 0.0° to 359.9° variable (Free / Fix selectable), 0.1° (1 Hz to 500 setting range Hz), 1°...
  • Page 58 ASR-6000 Parallel Models Series User Manual output current, 45 Hz to 65 Hz and sine wave output only. 3) Can be only set in 3P4W mode. 4) For phase voltage setting in polyphase output. In balance mode all phase are collectively set and in unbalance mode each phases are individually set.
  • Page 59 APPENDIX (0.7 % of rdg + 1 V / (0.7 % of rdg + 1 V / 2 V) 2 V) DC: ± (|0.5 % of rdg| DC: ± (|0.5 % of rdg| AVG value accuracy + 0.5 V / 1 V) + 0.5 V / 1 V) 45 Hz to 65 Hz and 45 Hz to 65 Hz: ±...
  • Page 60 ASR-6000 Parallel Models Series User Manual 3) The accuracy is for output waveform DC or sine wave only. 4) Accuracy values are in the case that the output current is 5% to 100% of the maximum current. 5) The accuracy is for output waveform DC or sine wave only. 6) In the polyphase output, these are the specifications for each phase.
  • Page 61: General Specifications - Asr-6450-13.5/ Asr-6600-18

    APPENDIX General Specifications – ASR-6450-13.5/ ASR-6600-18 Model ASR-6450-13.5 ASR-6600-18 Type A: Host, Type B: Slave, Speed: 2.0, USB- CDC / USB-TMC MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask Standard External Signal Input Interface External External Control I/O...
  • Page 62: Electrical Specifications - Asr -6600-24

    ASR-6000 Parallel Models Series User Manual Electrical specifications – ASR -6600-24 Model ASR-6600-24 Input ratings Power type Three-phase Four-wire Y connection Voltage range 380 Vac to 460 Vac (Line Voltage) Frequency range 47 Hz to 63 Hz Power factor 0.95 or higher (typ.) Efficiency 80 % or higher Maximum power...
  • Page 63 APPENDIX Phase angle 45 Hz to 65 Hz: ±1.0° accuracy 15 Hz to 550 Hz: ±2.0° DC offset ± 20 mV (typ.) Model ASR-6600-24 DC output (only single-phase output) Output capacity 24 kW Mode Floating output, the N terminal can be grounded Setting -250.0 V to +250.0 V / -500.0 V to +500.0 V, Setting Resolution: Range...
  • Page 64 ASR-6000 Parallel Models Series User Manual 13) For an output voltage of 50V or higher, sine wave, same load and voltage condition for all phase. 14) In the case of the AC mode and output voltage setting to 0 V, 23°C ± 5°C 15) For an output voltage of -250 V to -10 V, +10 V to +250 V / -500 V to -20 V, +20 V to +500 V, no load, AC voltage set to 0V (AC+DC mode) and 23°C ±...
  • Page 65 APPENDIX (VA) Accuracy ± (2 % of rdg + 18 VA) ± (2 % of rdg + 6 VA) Resolution 0.1 VAR / 1 VAR / 10VAR Reactive (VAR) Accuracy ± (2 % of rdg + 18 VAR) ± (2 % of rdg + 6 VAR) Range 0.000 to 1.000 Power factor...
  • Page 66: General Specifications - Asr-6600-24

    ASR-6000 Parallel Models Series User Manual General Specifications – ASR-6600-24 Model ASR-6600-24 Type A: Host, Type B: Slave, Speed: 2.0, USB- CDC / USB-TMC MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Standard Subnet Mask External Signal Input Interface External...
  • Page 67: Information Of Name Order

    APPENDIX Information of Name Order The name order of ASR-6000 series has its rules in definition for each character by order. Refer to the following contents for details. The definitions below describe the meanings behind Background each group of alphanumeric characters, in varied colors, of naming code for ASR series models.
  • Page 68: Asr-6000 Parallel Models Dimensions

    ASR-6000 Parallel Models Series User Manual ASR-6000 Parallel Models Dimensions ASR-6000 Parallel Models in 15u Rack Scale = mm...
  • Page 69: Asr-6000 Parallel Models In 19U Rack

    APPENDIX ASR-6000 Parallel Models in 19u Rack Scale = mm...
  • Page 70: Asr-6000 Parallel Models In 23U Rack

    ASR-6000 Parallel Models Series User Manual ASR-6000 Parallel Models in 23u Rack Scale = mm...
  • Page 71: Declaration Of Conformity

    APPENDIX Declaration of Conformity GOOD WILL INSTRUMENT CO., LTD. declare that the below mentioned product satisfies all the technical relations application to the product within the scope of council: Directive: EMC; LVD; WEEE; RoHS The product is in conformity with the following standards or other normative documents: ◎...