Unitech UT-1412 Instruction Manual

Ph indicator/controller

Advertisement

Quick Links

M
Model No. : UT-1412
PH INDICATOR / CONTROLLER
PANEL MOUNT
Flame Proof
@
www.utplindia.in
sales@utplindia.in
+91 7046223333 , 9427301436
www

Advertisement

loading
Need help?

Need help?

Do you have a question about the UT-1412 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Unitech UT-1412

  • Page 1 Model No. : UT-1412 PH INDICATOR / CONTROLLER PANEL MOUNT Flame Proof www.utplindia.in sales@utplindia.in +91 7046223333 , 9427301436...
  • Page 2 230 15% VAC - a switch, fuse (1/2A, 250V) or a circuit breaker. The above items should be installed together with UT-1412 for the products electrical protection. The switch, Fuse or circuit-breaker should be located close to the controller, within easy reach operator.
  • Page 3: Table Of Contents

    PH INDICATOR / CONTROLLER UT-1412 CONTENTS OVERVIEW ....................INTRODUCTION ..................SAFETY NOTES ..................NSTALLATION ........................................MOUNTING ....................EXTERNAL DIMENSION ....................TECHNICAL SPECIFICATIONS ....................WIRING DIAGRAM PROGRAMMING ..................................OPERATION INTERFACE ..............MODE ACCESS FLOW CHART OPERATIONAL MODES ...................
  • Page 4: Overview

    Ph of a substance with temperature the process parameter through a 4 Digit LED Displays . UNITECH UT-1412 Ph meter Type is a precise and compact pH Monitoring System using Solid state devices for direct measurement of pH. It is basically a voltage measuring device With added features. The high input impedance and the stability of its measuring circuit The water-resistant pH sensor quickly detects the pH value and temperature and give signals to meter.
  • Page 5: Installation

    PH INDICATOR / CONTROLLER UT-1412 2. INSTALLATION Mounting SIDE VIEW External Dimension                RL-1 ° RL-2 DIMENSION cutout SIZE 96 X 96 +0.5 +0.5 All dimensions are in mm...
  • Page 6: Technical Specifications

    PH INDICATOR / CONTROLLER UT-1412 Technical Specifications TECHNICAL SPECIFICATIONS Type Of Input PH & Temp. PH Electrode for PH Measurement 0 to 14 pH Measuring Range Input 0 to 150 °C [optional] Temp. 0.1 pH 1 Digit in Reading Accuracy Temp.
  • Page 7: Wiring Diagram

    PH INDICATOR / CONTROLLER UT-1412 Wiring Diagram 1 2 3 4 5 6 7 8 CAUTION WIRING CONSIDERATION RS - 485 RELAY-2 [HIGH AL.] RELAY-1 [LOW AL.] Shielded twisted pair wire is used for all analog I/P, Process sensors RTD, Thermo-couple, mV-mA DC, Volt DC, Linear input and Communication.
  • Page 8: Programming

    PH INDICATOR / CONTROLLER UT-1412 PROGRAMMING Operation Interface FRONT VIEW        pH Present SETUP/CAL. Value Display STATUS        TEMP. Present Relay-1 LED RL-1 Value Display Relay-2 LED °...
  • Page 9: Mode Access Flow Chart

    PH INDICATOR / CONTROLLER UT-1412 Mode Access Flow Chart. Setting Mode Normal Mode       TEMP Mode Selection  Press key to enter into the    Programming mode from Normal mode    ...
  • Page 10: Operational Modes

    PH INDICATOR / CONTROLLER UT-1412 OPERATION Opera ng Procedure. Connect Instrument to 230VAC. On back panel. Connect the Ph & Temp. Connection to their respective connectors. Remove protective cap of the Electrode containing the storage Solution [3.8 Molar KCL]. Wait for some time. The Instruments processor is ready for working.
  • Page 11: Calibration Procedure

    PH INDICATOR / CONTROLLER UT-1412 Calibra on Procedure. Step : 1 When ever you start for a New Calibra on Please must use Default [ ] func on  first. Use the func on as per under showing procedure Press Key...
  • Page 12: Alarm Relay Setting

    PH INDICATOR / CONTROLLER UT-1412 Step : 4 Please clean pH Electrode using D.M water and dry it before Calibration. Dip the Electrode in 9.2 pH Buffers and wait for few seconds to steady the Display reading. Display shows some reading near 9.20 pH.
  • Page 13: Installation Of Ph Electrode

    PH INDICATOR / CONTROLLER UT-1412 Installa on of pH Electode. Hand-tighten only! It is not necessary to apply excessive torque to achieve a liquid tight installation. Severe twisting of the sensor housing could cause internal damage. If necessary, use a wrench for removal only! For valve retractable sensors.
  • Page 14: Error Message

    PH INDICATOR / CONTROLLER UT-1412 Error Message. In case the following error messages appear in the display of controller, please refer to the Error message table below, or call technical support. Sign Description Solution Check Physically the pH Sensor ...
  • Page 15: Communication [Modbus Rtu] Protocol

    DATA+ RS-485 BUS Ethernet DATA- CONVERTER www.utplindia.in UNITECH TECHNOCRATS PVT LTD All content are subject to change without notice FACTORY:- 78/1/Z/3-Makarpura G I D C .., due to continuous improvements. +91 7046223333 , 9427301436 N/R Vadasar Bridge. /p-Gayatri emple Doc.