MSA Altair 4 Calibration Manual

MSA Altair 4 Calibration Manual

Gas monitor
Hide thumbs Also See for Altair 4:

Advertisement

Quick Links

The MSA Altair 4 Gas Monitor.
THESE SIMPLIFIED INSTRUCTIONS ARE DESIGNED TO
FACILITATE CALIBRATION OF THE GAS MONITOR. THEY DO NOT
REPLACE THE CALIBRATION INSTRUCTIONS AS ISSUED BY THE
MANUFACTURER, WHICH SHOULD ALWAYS BE READ BEFORE
THE INSTRUMENT IS CALIBRATED. THESE INSTRUCTIONS ARE
DESINGED TO PROVIDE AN EASY AND QUICK GUIDE TO
COMPLIMENT YOUR UNDERSTANDING OF THE MANUFACTURER S
1. Calibration.
a. To complete a full sensor calibration the following equipment is
required :-
Multi-mix Gas
Cylinder of
H2S,CO,CH4, &O2
Enter into calibration mode by pressing and holding the up arrow button until
the display asks for a zero operation to be completed..
CALIBRATION MANUAL.
·
Enforcer
·
Cylinder of Multimix Gas containing Methane
(29% LEL), H2S ( 20ppm ) CO ( 60ppm), 15% Vol.
O2 and balance N2
Regulator,
max 0.3
ltrs/min

Advertisement

loading
Need help?

Need help?

Do you have a question about the Altair 4 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for MSA Altair 4

  • Page 1 The MSA Altair 4 Gas Monitor. THESE SIMPLIFIED INSTRUCTIONS ARE DESIGNED TO FACILITATE CALIBRATION OF THE GAS MONITOR. THEY DO NOT REPLACE THE CALIBRATION INSTRUCTIONS AS ISSUED BY THE MANUFACTURER, WHICH SHOULD ALWAYS BE READ BEFORE THE INSTRUMENT IS CALIBRATED. THESE INSTRUCTIONS ARE...
  • Page 2 Check that the label on the gas canister matches the concentration of gas which the instrument requires g. Attach the aspirator / calibration adaptor to the MSA Altair 4. h. The flow of gas is detected automatically and the calibration...
  • Page 3 When it has finished the calibration the MSA Altair 4 display will confirm correct calibration. j. If a fail message is displayed please contact IM&M Ltd for further assistance. k. A note should be made in the vessel log of the date and time of...