Page 1
Microscope Instruction Manual STM-2081F This manual is for users of STM-2081F Scientific Fluorescent Microscope in laboratory. To ensure your safety, obtain optimum performance, and to familiarize yourself fully with the use of this microscope, we recommended that you study the instruction manual carefully.
Contents Microscope Part ..........................2 Notice For Use ........................... 2 1.Components Name ........................4 2.Structure Overview ........................5 3.1 Open Power Source(fig.1) ....................7 3.2 Adjust Lighting and Switch Light Source (fig 2) ..............7 3.4 Mount Specimen And Adjust Stage( ................... 8 3.5 Adjust Interpupillary Distance( ...................
Microscope Part Notice For Use STM-2081F I. Safety Precaution Symbol Description During use, the power box of microscope become very hot, so should not be touched with bare hands. Before using, please read the user manual carefully. If can not use...
III. Maintenance and Storage All camera lenses are proofread and adjusted. Please tear open outfit Revolving nosepiece and coarse micro focusing mechanism, which structure are precise. Please don’t remove easily. Instruments shall be kept clean, always remove dust. Please care about that don’t contaminate optic components.
2.Structure Overview STM-2081F 1.Filter switching knob 8.”ECO”infrared sensor 2.Coarse and fine focusing handwheel 9.Coaxial pinion for X stage movement 3.Focusing limiting and locking 10.Coaxial pinion for Y stage movement Handwheel 11.Condenser disk 4.Condenser height adjustment wheel 12.Biopsy clip 5.Lighting switching button 13.Screw for fastening extensional board...
Page 7
STM-2081F 8.Focusing limiting and locking 1.pushing rod for switching light path handwheel 2.Condenser fastening screw 3.Condenser height adjustment 9.Screw for fastening lamp chamber wheel 10.Socket of lamp chamber 11.Grounding bar 4.”ECO”button 12.Plug of lamp chamber 5.Aperture size adjustment knob 6.Fine focusing handwheel 13.Lamp chamber...
3.Adjustment And Operation STM-2081F 3.1 Open Power Source(fig.1) Plug power source, toggle the main switch in ‘-’state, which on the back of microscope. 3.2 Adjust Lighting and Switch Light Source (fig 2) 1. Rotating the handwheel according to the direction...
STM-2081F ★ when switching light button, if found the button is not matched with the shown in the figure, please check that 3.4 Mount Specimen And Adjust Stage(fig 4, fig 5) Push the cover glass slowly into the jaw, then clamp Fig 4 glass gently.
Lable of switching light path 3.5 Adjust Interpupillary Distance(fig 6) The range of interpupillary distance is 47mm~78mm. when Pushing rod observing with binocular, holding the prism table and pivoting to adjust the interpupillary distance, until two sides of field roll into one and observe with binocular comfortably. 3.6 Three Light Path Switching(fig 7) Fig 7 Icon...
Handwheel for limiting and On both sides of the fine focusing handwheels are locking focusing removable. According to the direction of arrow, hold one end of handwheel and make another cock. Then stir it from the cock end. About install it, only need to align the center of handwheel and assemble it.
Page 12
Condenser centering: 1. Rotating the focusing handwheel of condenser① and rising the condenser to the top position. 2. Focusing the specimen with the objective 10x. 3. Rotating the field diaphragm ②, shrinking image of field diaphragm to view. 4. Rotating the focusing handwheel ① to focus the image of field diaphragm.
3-11 Using Frosted Sheet(figure 13) As shown in figure 13, the state of using filter is push in the rod. When push out the rod, filter is in an empty state. LBD is a filter of increasing color temperature. Fig 14 ND25 is a filter of 25% light transmittance.
3-14 Replace The Fuse (fig 16, fig17) As shown in the figure 16, first using the general wrench push aside the drawer of fuse. Then referencing figure 17, through the square hole under of the fuse drawer gently drag the fuse upward.
4.Installation STM-2081F 4.1 Installation Diagram The diagram as below shows the order of each component’s installation. The figures in the diagrams show the installation steps. ★Before installing, please ensure that all components are clean. Don’t cut any component or surface of glass.
4-Procedure STM-2081F 4-2-1 Install the Mechanical Slide Bracket (fig 1, fig 2) ★ Install the platform bracket according to the path of ① in figure 1. The dovetail slot needs to align with the slot of platform bracket, sliding it Fig 1 until the screw fix it.
4-2-3 Install Nosepiece (fig 5) Connect the plug of nosepiece with socket, then Ⅲ inset the dovetail groove of microscope according to the path ③ shown in figure 5. At last, use hex Fig 5 wrench tight up the bolt shown as Ⅲ. 4-2-4 Install Camera Lens And Photo tube/0.5X C-mount (fig 6)
★ If not use the ternary camera lens and the photo accessories, please use the corresponding dust cover to cover the ternary attachment interface and eyepiece interface, to prevent into ashes. Fig 7 4-2-5 Install Eyepiece(fig 7) 1.Inset eyepiece into eyepiece sleeve according to the path ⑥...
★ when operating, first under of the objective 10x search the sample and focus, then switching to other ratio of objective lens. ★ When switching objectives, turn the objective nosepiece. It does not ensure the objective lens into the heart of the light path until hear the “click” Fig 9 sound 4-2-7 Install the Electric...
★ In order to ensure the appropriate position of camera interface, first use eyepiece to observe and adjust the sample to a clean imaging position. Then install the camera and observe. At the same time, accordingly adjust the position of camera interface until to the clearest position, then screw the number Ⅺ...
5.Technical Specification STM-2081F (1) Main Technical Specification Optical System NIS60 Infinite optical system STM-2081F Seidentopf Trinocular Head, 30° inclined, interpupillary distance: 47mm-78mm; splitting ratio Eyepiece: Trinocular Standard =100:0 or 20:80 or 0:100 Viewing Head Ergo Tilting Trinocular Head, adjustable 0-35° inclined,...
Page 22
C-Mount 0.5× Optional Polarization Kit Simple Polarizing kit with polarizer and analyzer Optional DIC Attachment Optional Phase contrast Kit Turret Phase Contrast Attachment Optional Dark-field Condenser (Dry) Optional Dark field device Dark-field Condenser (Oil) Optional Fluorescence device Standard Note: Standard Standard Outfit, Optional Optional (3) Objective Parameters Cover glass Conjugate...
6.Trouble shooting guide 6.1 Optical Part TROUBLE CAUSE SOLUTION The edge of the field The nosepiece is not in the located position Locate the nosepiece of view is dark or (objective and light path are not coaxial) properly where it clicks the brightness is not The image of filament is not centered Center the filament...
Page 24
There was bleb in the oil Eliminate the bleb Have used the unsuitable oil Use standard immerse The size of the aperture diaphragm is too Minify it large There are stains on the incident lens of the Clean up binocular tube The size of the aperture diaphragm is too Open it up small...
6.2 Mechanical System TROUBLE CAUSE SOLUTION The image can not focus The specimen is placed Turn inversely when using high inversely Use the standard coverslip magnification objective The coverslip is too thick (0.17mm) The objective touches the The specimen is placed Turn inversely specimen when...
6.3 Electrical System TROUBLE CAUSE SOLUTION Check the connection of the No power power cord The lamp can’t light when the switch is turned on The bulb is not inserted Insert it correctly The bulb burns out Replace it Use the specified lamp to The lamp burns out Use a substandard lamp replace, if the problem is not...
FL-900 Epi- fluorescent Attachment This manual is written for FL-900 Epi-fluorescent attachment. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this attachment, it is strongly recommended that you read this manual carefully before using it.
Notice for use The FL-900 epi-fluorescent attachment has been designed for STM-2081 scientific microscope. ⅠSafety Note 1. The epi-fluorescent attachment is a precise instrument. Open the box carefully, and avoid dropping the accessories to ground and causing damage to them. 2.
Page 29
Symbol Explanation Indicates that the surface becomes hot, and should not be touched with bare hands。 Indicates that high voltage (upper 1KV) inside, improper handling could result in an electric shock to the user. Before use, carefully read the user manual. Improper handling could result in personal injury to the user and/or damage to the equipment.
Page 30
1.Components Name ●FL-900 Epi-fluorescent Attachment includes: (FL-900)=the body of fluorescent Attachment +(NFP-1) (shown in figure) ① Main body of the Epi-fluorescent Attachment ② Power supply unit NFP-1 ③ Power cord (please use the power cord provided) ④ Ultrahigh mercury burner ⑤...
Page 31
2. Assembly ● STM-2081F Scientific Fluorescent Microscope: STM-2081F = STM-2081. Assembly of STM-208 1F Laboratory Fluorescent Microscope: 1. Loosen the set screw① and take the trinocular Viewing Head ③ from the body of scientific microscope. 2. Insert the epi-fluorescent attachment into the scientific microscope correctly and tighten the setscrew①...
Page 32
2.1Preparation Open the box carefully, remove all packing material and take the attachment out. 2.2 Mounting the Mercury Burner (fig.1, fig.2) 1. Loosen the burner socket clamping screw ① (fig.1), and remove the burner socket (fig.2). 2.After removing the foam backstop②, securely insert the + pole (the wide head) of the specified mercury burner③...
Page 33
★To prevent any hazard, always turn the main switch on the power supply unit to “O” (OFF), unplug the power cord plug from the mains outlet, and wait for at least 10 minutes before replacing the burner. ⑤ 2.3 Mounting Filter Blocks ⑥...
Page 34
2.4 Mounting Protection Barrier (fig.5) Install the protection barrier on the attachment by tightening the screw. Screw Protection barrier Fig 5 Assembly Fluorescent Attachment (fig.6) ⑤ Mount the lamp housing① into the other end of the attachment② and fix it with two screws③. ⑥...
Fuse Replacement ③ ② (fig.8 ,fig. 9) 1 Set the main switch ④ to “O”(OFF)and unplug the power cord before replacing fuses. ① 2 Using a flat-blade screwdriver, remove each Fig 8 of the fuses holders ⑦ by tuning it counter- clockwise and pulling out.
3. Adjustment and Operation 3.1 Name of Components Diaphragm Centering Screw Mirror centering knob Lamp centering knob Filter block turret Condenser Filter block label adjustable knob Aperture diaphragm lever Light shutter lever Field Accessorial diaphragm slot lever ND Filter slot...
Page 37
◎ There are 6 fluorescent mirror blocks (filter Fluorescent mirror block block) mounted in the filter block turret at (filter block ) the most. ● a mirror block includes a diachronic mirror, a barrier filter, an excitation filter. (There are kinds of excitation filters).
Page 38
3.2 Operation 3.2.1 Preparation 1.Verify that the voltage and the frequency of the AC mains outlet match the setting of the voltage switch and the frequency switch on the rear of the power supply units. 2.Make sure the cord is connected firmly. 3.When transmitted light observation is required, pull out the filter system and make the hole in the light path.
Page 39
3.2.2 Selecting Fluorescent Filter Combination Select fluorescent filters combination according to the fluorescent dye you use. Excitation Diachronic Excitation Barrier Filter Application Mirror Filter DM400 BP330-385 BA420 · Auto-fluorescence observation · DAPI: DNA · Hoechest 332528, 33342: Chromosome DM455 BP400-410 BA455 ·...
Page 40
3.2.3 Switching on Electrical Source Set the main switch of the power supply unit to “I” (ON). It will stabilize in 5 to 10 minutes after ignition. ◎ Some mercury burners may not ignite the first time the power is turned on due to variance in production, and the safety mechanism in the starter in such a case.
3.2.4 Centering the Field Iris Diaphragm (fig.1) ③ 1. Switch the light shutter ① to “●”position. 2. Revolve filter block turret to engage the B-excitation mirror in the light path. 3. Switch the light shutter ① to “O” position. ① 4.
Page 42
observation target region. According to the objective in use, adjust the diaphragm image using the field diaphragm lever so that the field of view is circumscribed by the field diaphragm to exclude stray light. 3.2.5 Centering the Aperture Iris Diaphragm (fig3) 1....
Page 43
9 . Adjust the aperture iris diaphragm centering screws④ with attached wrench to superpose the image of aperture iris diaphragm on the cross of centering plate. Adjusting the aperture iris diaphragm (fig.3) The aperture iris diaphragm adjusts image resolution and contrast. For fluorescent observation, push in the aperture iris diaphragm lever③.
3.2.6 Centering the Mercury Burner (fig.4, fig.6) ◎Before proceeding to center the burner, wait for the arc image to stabilize to protect against glare during arc image centering, it should be viewed across the excitation light protective shield. Fig 4 1.
Page 45
are image and the mirror reflected arc image in the symmetrical position.(B) ⑦ 9. Adjust the mirror focusing knob⑥ (Fig.6) to ⑥ sharpen the mirror reflected arc image. (C) 10. Turn the burner adjustin,,g knob ⑤ to overlap the are image with the mirror reflected are image. Fig 5 ◎...
Page 46
3.2.7 Mounting ND filter (fig.7) 1.The ND filter can reduce the excitation light intensity to delay color fading of the specimen. Use the ND filter as far as this does not hinder observation. 2. There are two kinds of ND filters for option:ND6 and ND25 for position①...
Page 47
4. Troubleshooting Guide Under certain conditions, performance of the attachment may be adversely affected by factors other than defects. if problems occur, please review the following list and take remedial action as needed. PROBLEMS CAUSE SOLUTION I. Optical Part The light shutter closes the light Switch the light shutter to “○”position path The ND filter is engaged in the light...
Page 48
The field of view doesn’t open fully Open it fully ND filter is stopped in halfway in the Pull in the filter slider until it clicks into light path place The mercury burner is not centered. Center it The collector focus position is not Adjust it to an optimum position correct d)Shadow exists in the field...
Page 49
5. Characteristics of Mirror Block’s Wave Length Blue excitation Green excitation Ultraviolet excitation Violet excitation...
Page 50
Fluorescent Filter Excitation Dichroic mirror Barrier Filter block B Excitation BP460~490 DM500 BA520 Standard Epi-Fluorescent Illumination G Excitation BP510~550 DM570 BA590 Standard U Excitation BP330~385 DM400 BA420 Optional V Excitation BP400~410 DM455 BA455 Optional Lamp 100W GCQ Ultra Hi-voltage Spherical Mercury Lamp Standard Protection Barrier Barrier to Resist the Ultraviolet Light...