Biological Microscope
Model: STM-2036 Series
Instruction Manual
This manual is written for
STM-2036
Series
biological
microscope.
To ensure the safety, obtain
optimum performance and to familiarize you fully with the use of this microscope , it is
recommended strongly that you study this manual thoroughly before using the microscope and
retain this manual in an easily accessible place near the work desk for future reference.

Advertisement

loading
Need help?

Need help?

Do you have a question about the STM-2036 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Stalwart STM-2036 Series

  • Page 1 Biological Microscope Model: STM-2036 Series Instruction Manual This manual is written for STM-2036 Series biological microscope. To ensure the safety, obtain optimum performance and to familiarize you fully with the use of this microscope , it is recommended strongly that you study this manual thoroughly before using the microscope and...
  • Page 2: Table Of Contents

    ……………………………………………………………… 6. Maintenance ……………………………………………………………… 7. Troubleshooting ……………………………………………………………… 8. Outfits ……………………………………………………………… STM-2036 series biological microscopes are designed for college teaching, Medical and clinical identification. The microscope with modern design, steady structure, convenient operation and clear image is suitable for observing various biological specimens, they are mostly applied in colleges and hospitals.
  • Page 3: Parts Name

    1. Parts Name ① Eyepiece ① ⑧ ② Objective ③ Clamp ④ Mechanical Stage ⑤ Condenser ② ⑥ Light Collector C_③ ⑦ Main Body Seidentopf ④ ⑧ Binocular Head (Trinocular Head) D_⑤ Power Input ⑨ Fuse E_⑥ ⑦ Diopter Adjustment Ring Nosepiece Clamp Handle Handle of Iris Aperture Diaphragm...
  • Page 4 2.1 Total magnifications Objective Total Magnifications 100X (Optional) Eyepiece 100X 200X 400X 1000X 160X 320X 640X 1600X 2.2 Objectives STM-2036A, STM-2036 AT Numerical Objective Resolving Working Achromatic Objective Aperture (N.A.) Field Power Distance 0.10 4.5mm 2.8µm 13mm 0.25 1.8mm 1.1µm 6.3mm 20X (S) (Optional) 0.40...
  • Page 5: Installation

    2.3 The other specification 2.3.1 Mechanical tube length: 160mm 2.3.2 Conjugate distance: Infinity (Non- Infinity: 195mm) 2.3.3 Head: Seidentopf Binocular or Trinocular, Inclined 30°, Rotatable 360°, Anti-fungal systems. Interpupillary Adjustable Distance Is 48-75mm, Diopter adjustable range ±5. 2.3.4 Nosepiece: Quadplex nosepiece 2.3.5 Mechanical stage: Size 145mm×140mm, X-Y travel 76mm×52mm 2.3.6 Focusing systems: Coaxial Coarse and Fine Focusing Knobs, Coarse stroke 26mm, Fine division 2µm, Condenser up-down range 22mm...
  • Page 6: Operation

    eyepiece by hand. 3.4 Condenser Put out the condenser of the carton, then turn Condenser Focusing Knob M to lower condenser installing ring. Loose the condenser screw, then install the condenser, and make the condenser graduation face to the front, Tighten the condenser screw, and higher the condenser installing ring to top.
  • Page 7: Image Capture

    5. Image Capture 5.1 Installing Connect the C-mount with CCD camera or connect camera with camera adapter, then connect it with c-mount, finally put it into microscope. 5.2 Using First get a clear image from eyepiece, then put out lever on the side of trinocular head, and collect image with camera.
  • Page 8: Troubleshooting

    dust affect bulb brightness and service life, and insert fully the contact pins into the bulb socket. 6.3.6 Close the lamp housing cover, and screw the knurled thumb screw. ▲After working for above 10 hours continuously, better cut off the microscope about 30 minutes. 6.4 Replacement of fuse 6.4.1 Cut off power of microscope, and pull out the plug.
  • Page 9 Trouble Causation Remedy Size of iris aperture isn’t proper Adjust the size of iris aperture Readjust the position of Position of condenser is too low condenser Make the objective in One side of image is Objective isn’t in correct position correct position dark or image is Place specimen levelly on stage and...
  • Page 10 Items Specification STM-2036A STM-2036B STM-2036C STM-2036D WF 10×-18 mm Standard WF 10×-20mm Standard Standard Standard WF16×-13mm Optional Optional Optional Optional Reticule Eyepiece WF 10×-18 Eyepiece Optional Optional Optional Optional (Reticule 0.1mm) Reticule Eyepiece WF 10×- 20mm Optional Optional Optional Optional (Reticule 0.1mm) 4×...
  • Page 11 Stage Size: 145X140mm Mechanical Travel: 76mmX52mm Standard Standard Standard Standard Stage Coaxial Coarse Fine Focusing Knobs Abbe N.A. 1.25 Condenser Standard Standard Standard Standard (Iris Diaphragm) 3W LED Illumination Systems, Standard Standard Standard Standard Brightness Adjustable Illumination 6V/20W Halogen Lamp, Optional Optional Optional...