Digital Monocular Microscope STM-2010MD Instruction Manual This manual is written for digital monocular microscope STM-2010MD. To ensure the safety, obtain optimum performance and to familiarize the use of this microscope, it is recommended strongly that you study this manual thoroughly before using the microscope and retain this manual in an easily...
Brightness adjusting knob 1. Brief introduction Digital monocular microscope STM-2010MD is widely used in biological, pathological, bacteriological fields for purposes such as observation, research, academic demonstration, clinic, medical inspection. It features more comfortable and reliable operation with modern design and advanced technology.
3. Eyepiece Category Magnification Focal Length f(mm) Field view (mm) Plan 24.95 Ø18 Eyepiece 4. Total Magnification Eyepiece Objective 4X 100X Total 100X 400X 1000X 5. Objective conjugate distance:195mm 6. Stage: double layer stage 125mmX115mm 7. Coarse focusing range: 6.5mm 8.Condenser: Abbe Condenser,NA=1.2 with iris diaphragm 9.Diameter of filter:Ø28mm 10.Illuminator:LED Lamp,brightness adjustable...
Page 4
d) Stage Use double-layer mechanical stage, the slide is fixed with slide clip. e) Illumination system The illumination system consists of illuminator, iris diaphragm and condenser. Abbe condenser can be raised and lowered by simple operation. 5. Assembly A) Environmental Requirement: 1.Ambient temperature: 5°C to 40°C(41°F to 104°F),Maximum R.H:85%。...
2. Place the specimen in the center of the stage, keeping the cover slip upward, and then nip it with the specimen holder. 3. Search the image by lower magnification objective, turn the longitudinal and transversal adjustment knobs which on the mechanical ruler to move the specimen onto the center of the field of view.
8. Troubleshooting Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local franchiser for assistance.
Page 7
2). Mechanical part TROUBLE CAUSE SOLUTION The image can focus Turn inversely when using high The specimen is placed inversely standard magnification The coverslip is too thick coverslip(0.17 mm) objective The objective and the specimen touch Turn inversely when change The specimen is placed inversely standard objectives from low...
Page 8
2. Install the software, follow the clue to finish the installation, after the installation, there will be a shortcut icon on the Desktop and the Start Menu. 3. Please install the video codec ‘wmv9VCMsetup.exe’, it can compress the video files when recording, and greatly reduce the size of the video files.
Page 9
the 32bit operation system. ★ Please unplug the USB cable from the computer immediately after using. It will shorten the lifetime of the camera, sometimes may cause repair service if the camera is plugged in the computer all the time and you do not shut down the computer for a long time. 2) Operation ScopeImage 9.0 is image-processing software professionally designed for digital microscope.
Page 10
3) Special design color database, we can associate the parameters of the color to the database, and can store or delete the color scheme. 4) Various image measurement tools, including line, angle, rectangle and circle, can move, delete or set the color of the objects. And you can also burn all the objects into the image for further using.
Need help?
Do you have a question about the STM-2010MD and is the answer not in the manual?
Questions and answers