Kolibri KCD-2000 Calibration Procedure

Hide thumbs Also See for KCD-2000:

Advertisement

Quick Links

KCD-2000
Calibration Procedure
1. Purpose & Scope
This document will demonstrate the regularly required and preventative calibration maintenance procedure.
2. Application
Calibration of the unit is always performed before shipment. Sometimes during transit or after a long period
of use, however, the calibration of the sensors might be altered. These conditions can lead to a higher rate of
false rejects.
3. Definitions & Acronyms
Refer to KCD-2000 User Manual for definitions.
4. Requirements
No tools required.
5. Procedure
1.
Ensure the machine is plugged in.
2.
Press and hold the DOWN ARROW button, then press the POWER button while still holding the DOWN
ARROW button, until the calibration screen appears.
2. Press POWER
while still holding
DOWN ARROW
KB-052019
1. Press and
hold DOWN
ARROW
2019 Kolibri USA

Advertisement

loading
Need help?

Need help?

Do you have a question about the KCD-2000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Kolibri KCD-2000

  • Page 1 These conditions can lead to a higher rate of false rejects. 3. Definitions & Acronyms Refer to KCD-2000 User Manual for definitions. 4. Requirements No tools required. 5. Procedure Ensure the machine is plugged in.
  • Page 2 It is possible, if done carefully, to complete this step in one upward or downward movement. Lift the lid up slowly Lower the lid down slowly Once the calibration is completed the you will be prompted to Exit. KB-052019 2019 Kolibri USA...
  • Page 3 Phone: 1-877-898-8750 Support: https://kolibriusa.com/contact/ If, after calibrating the unit, this does not resolve the issue, please see the support section on the Kolibri USA website at www.kolibriusa.com, or contact Kolibri Support at support@kolibriusa.com for further troubleshooting. KB-052019 2019 Kolibri USA...