Moog CURLIN 6000 Worksheet page 2

Ambulatory infusion pump
Hide thumbs Also See for CURLIN 6000:

Advertisement

AMT/DOSE: Enter the amount of each dose, press YES
9.
10. DOSE RATE: Enter the infusion rate, press YES
11. DOSE DURA: Enter the dose duration, press YES
12. DOSE Freq: Enter the dose frequency, press YES
13. KVO RATE: Enter KVO (Keep Vein Open) rate, then press YES (0 to 10 ml/hr)
14. DOSE/BAG: is auto-calculated. Press YES to confirm or NO to change
15. REQ VOL: is auto-calculated. Press YES (no change allowed)
16. TOT TIME: is auto-calculated. Press YES (no change allowed)
17. DONE?: Press YES to accept
When at the Run to Start Screen, press the HELP/OPTIONS key to set the Lock Level
(security level), and to change the Air-In-Line sensitivity and the Down Occlusion
sensitivity from the default settings when the therapy requires these changes:
Lock: Press YES to accept or NO to change Lock setting, press YES to accept new setting.
DN Occlu: Press YES to accept (default setting is LOW) or
press NO to change then press YES to accept.
AIR SENS: Press YES to accept (default setting is 0.5 mL)
or NO to change then press YES to accept.
Press YES at each field to advance to the next field. Press
YES to ACCEPT OPT at the end of the menu.
PROGRAMMED BY:
CHECKED BY:
tel: 801.264.1001
toll free: 800.970.2337
www.curlinpump.com
AMT/DOSE
DOSE RATE
DOSE DURA
DOSE Freq
KVO RATE
DOSE/BAG
REQ VOL
TOT TIME
Lock
DN Occlu:
AIR SENS:
DATE:
DATE:
CURLIN® is a registered trademark of Curlin Medical, Inc. Moog® is a
registered trademark of Moog, Inc. © 2018 Moog, Inc. Moog Medical,
4314 Zevex Park Lane, Salt Lake City, UT 84123 USA. LS-61474 Rev. A

Advertisement

loading
Need help?

Need help?

Do you have a question about the CURLIN 6000 and is the answer not in the manual?

Questions and answers