Moog CURLIN 6000 Worksheet
Moog CURLIN 6000 Worksheet

Moog CURLIN 6000 Worksheet

Ambulatory infusion pump
Hide thumbs Also See for CURLIN 6000:

Advertisement

Quick Links

Worksheet for Non-Library Programming

TPN THERAPY

Refer to the CURLIN 6000 User Manual for complete programming instructions.
PATIENT:

PROGRAMMING / SETUP SEQUENCE

NOTE: Use the
(UP) or
data entry. Use the YES and NO keys to accept/confirm or reject/change a field entry.
1.
Turn the pump on and wait for the self-test to complete, note the power source with power graph on the screen.
2.
Highlight PROGRAM from the SELECT menu and press YES
NOTE: If the following screen displays "NEW PROGRAM", the PREVIOUS RX has not been cleared. Refer to the last section of the "CURLIN
6000 TRAINING & QUICK REFERENCE GUIDE FOR BETWEEN PATIENT USE" to clear all Previous RX prior to new patient use.
Highlight TPN from the therapy selection menu and press YES.
3.
NOTE: If any on-screen alerts appear, the Previous RX has not been cleared. Refer to the last section of the "CURLIN 6000
TRAINING & QUICK REFERENCE GUIDE FOR BETWEEN PATIENT USE" to clear all Previous RX prior to new patient use.
With Unidentified highlighted, press YES (this indicates Manual Programming Mode)
4.
DELAY: Press YES to leave the delayed start feature off,
5.
press NO to set a delay then press YES to ON:
TIME: Enter time infusion to begin, press YES
DATE: Enter date Infusion to begin, press YES
ACCEPT?: Press YES
At NEXT? Press YES to continue
6.
BAG VOL: Enter bag volume, press YES. (Bag volume should include
7.
overfill volume if pump priming and/or a KVO is programmed.)
NOTE: To force an auto-calculation, enter any 3 of the 4 entries: Vol TBI, INF RATE, UP/
DN Ramp or TOT TIME, the unknown field will auto-calculate. Bypass the unknown field by
using the DOWN ARROW key (typically the INF RATE). Enter the known fields pressing YES to
confirm each. The unknown field will auto-calculate. Press YES to confirm this field.
8.
Vol TBI: Enter the amount to be infused, press YES
tel: 801.264.1001
toll free: 800.970.2337
www.curlinpump.com
(DOWN) ARROW keys to move highlighter and the numeric keys for
PUMP SERIAL #:

DELAY:

TIME:
DATE:

BAG VOL:

Vol TBI:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CURLIN 6000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Moog CURLIN 6000

  • Page 1 Highlight TPN from the therapy selection menu and press YES. NOTE: If any on-screen alerts appear, the Previous RX has not been cleared. Refer to the last section of the “CURLIN 6000 TRAINING & QUICK REFERENCE GUIDE FOR BETWEEN PATIENT USE” to clear all Previous RX prior to new patient use.
  • Page 2 800.970.2337 www.curlinpump.com CURLIN® is a registered trademark of Curlin Medical, Inc. Moog® is a registered trademark of Moog, Inc. © 2018 Moog, Inc. Moog Medical, 4314 Zevex Park Lane, Salt Lake City, UT 84123 USA. LS-61476 Rev. A...

Table of Contents