Clear and Sanitize Procedures
Table 1: Volatile memory devices (DPO7000, DPO70000/B, and DSA70000/B Series Instruments)
Type
and
minimum
size
Function
SDRAM,
Acquisition
512M X
memory for
16 up to
holding and
512M X
processing
64 bits
waveforms,
and processor
system RAM
PC RAM
PC
(See
motherboard
page 16,
PC Ram
Size.)
SDRAM,
Acquisition
512M X
memory for
16 bits
holding and
processing
digital
waveforms
2
Type
of user
Backed-up
info
by
Method of
stored
battery
modification
User data
No
Indirectly
User data
No
Indirectly
User data
No
Indirectly
Data
input
User
method
Location
accessible
Written
Analog
Yes
by
acquisition
processor
board
system
Written
PC
Yes
by
motherboard
processor
or COM-E
system
module
Written
Digital
Yes
by
acquisition
processor
board
system
To
clear
To sanitize
Remove
Remove
power
power from
from
the instrument
the
for at least
instrument
20 seconds
for at
least
20 seconds
Remove
Remove
power
power from
from
the instrument
the
for at least
instrument
20 seconds
for at
least
20 seconds
Remove
Remove
power
power from
from
the instrument
the
for at least
instrument
20 seconds
for at
least
20 seconds
Declassification and Security
Need help?
Do you have a question about the MSO70000 Series and is the answer not in the manual?