M526-E024N 2024-06 Collimator Type R-300/RC-300 Operation Manual Read the manual thoroughly before you use the product. Keep this manual for future reference.
Page 2
About the Symbols Appearing in this Operation Manual Throughout the text in this manual, warnings and other information essential when using this unit, such as cautionary or prohibited items, appear classified as per the following: Mark Description Indicates an imminently hazardous situation which, if not avoided, will result in serious injury or death.
Page 3
Preface Thank you for purchasing a Collimator Type R-300/RC-300 (hereafter referred to as "equipment"). This operation manual contains the information for using the collimator correctly. Read this manual before using the collimator, and use it correctly. Failure to observe the cautions in this manual may result in damage to the collimator or harm to the patient (person being X-rayed).
Page 4
(2) Confirm that the ground is connected properly. (3) Check all wiring for proper and correct connections. (4) Check the condition of the external electric circuit, which will be directly connected to a patient. (5) Check the condition of the battery power source. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 5
When the equipment is found to be out of order, do NOT try to repair it. Display an appropriate sign to indicate that the equipment is out of order, and contact a Shimadzu service representative for repair. Do NOT modify any part of the equipment.
Never modify the equipment. In general, modifications are strictly prohibited by the Regulatory requirements of the law of the country where the equipment is installed. Please contact your Shimadzu service representative if it is necessary to modify the equipment. Perform periodic inspection.
Page 7
If the operator has no experience in operating the equipment, be sure that he or she receives instruction on how to operate it from Shimadzu service personnel or someone who has adequate experience in using the equipment.
Page 8
Should such liquid drip on equipment surfaces, wipe it off immediately. Any such liquid entering into system electronics may cause failure or malfunction. Should liquid drip on the equipment or enter the covers, immediately turn off the power and contact your Shimadzu service representative. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 9
Do NOT open the covers of the equipment. Otherwise, electric shock may result. When opening the covers for maintenance, contact your Shimadzu service representative. Always be very careful when moving the equipment to avoid contact with the patient or operator and to ensure that the patient or operator does not become caught between the equipment and any neighboring devices.
"1.4 Environmental Conditions" P.3 ^ Reference Note also that there must be no sudden changes in temperature or humidity. This causes condensation, which can lead to equipment failure. viii R-300/RC-300 OPERATION MANUAL M526-E024...
Particular ways of using the equipment may increase the scatter dose absorbed into the patient, which may cause a radiation hazard. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 12
Doing so may result in unnecessary radiation exposure to the patient. Be sure to carry out a warm-up (running-in operation of the X-ray tube unit) before taking an X-ray radiograph. Follow the warm-up procedure described in the X-ray high-voltage generator operation manual. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 13
The shorter the distance becomes, the greater the amount of scatter dose absorbed into the patient, which may cause a radiation hazard. Pay extra attention when irradiating X-rays for a long time or repeatedly. It may cause a radiation hazard. M526-E024 R-300/RC-300 OPERATION MANUAL...
Wipe down all contact surfaces using a cloth moistened with an appropriate disinfectant or cleaner. Make sure the cloth is NOT too wet. If it is, liquid may enter into system electronics, causing failure or malfunction. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 15
Repeated disinfection over a long time may lead to discoloring and cracking on the equipment surface, and deterioration of rubber and plastic. If any abnormality is found on the equipment after disinfection, stop using the equipment immediately. Contact your Shimadzu service representative for repair. xiii...
Page 16
On completing the work, check the following points before switching the power ON again. • There must be no water or disinfectant adhering to the equipment. • The tools used in cleaning and disinfecting work must be tidied away. "3.1 Turning the Power ON/OFF" P.22 ^ Reference R-300/RC-300 OPERATION MANUAL M526-E024...
ON. Such devices can exceed the EMC standard limitations, and under some conditions this can impair the proper functioning of the equipment. In the worst case, this can cause serious injuries or clinical errors. M526-E024 R-300/RC-300 OPERATION MANUAL...
Do NOT use this equipment adjacent to, or stacked with, other equipment. If adjacent or stacked use is necessary, check to be sure that this equipment works properly in the environment "7.1 Environmental Conditions of EMC (Electromagnetic Compatibility)" P.58 ^ Reference R-300/RC-300 OPERATION MANUAL M526-E024...
Page 19
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the R-300/RC-300, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Page 20
GSM 180, CDMA 1900, GSM 1900, DECT, LTE Band 1,3,4,25, UMTS should not be used no closer than 45 cm to any part of the R-300/RC-300, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
DSA, or cineradiography) can adversely affect the CMOS circuit in implantable pacemakers and defibrillators. Such affects may cause oversensing in these devices that can temporarily inhibit pacing pulse output and result in an inappropriate heart rate. M526-E024 R-300/RC-300 OPERATION MANUAL...
The system is warranted to be free from defects in material and workmanship for one year from the date of delivery. If found to be defective, the system must be offered to Shimadzu for inspection and examination. Upon examination, Shimadzu, at its sole option, will repair or replace at no charge, the system or any part found to be defective.
In some jurisdictions, some of the foregoing warranty disclaimers or damage limitations may not apply. Shimadzu will be indemnified for any claim, liability, or damage arising out of the misuse or non-compliance with this manual by the purchaser or lessee of the system.
WEEE Mark Contact your Shimadzu service representative when the equipment has reached the end of its life. They will advise you regarding the equipment take-back. With your co-operation we are aiming to reduce contamination from waste electronic and electrical equipment and preserve natural resource through re-use and recycling.
Inspect the safety labels periodically (once a year). If any label is peeled or unreadable by stain or scratch, replace it with a new one. For new labels, contact your Shimadzu service representative. • R-300...
Explanation Source Image Distance Light Emitting Diode Digital Radiography Computed Radiography Flat Panel Detector Options The "Options" described in this manual may be provided as standard, or may be unavailable, depending on the sales format. xxiv R-300/RC-300 OPERATION MANUAL M526-E024...
Intended Use The collimator R-300 is fitted to the X-ray tube unit and is used to adjust the X-ray irradiation field. Its size can be known by the emission field of visible light and the indication on the display panel.
Even under the prescribed conditions, avoid rapid changes of temperature or humidity. Condensation may occur and cause failure. Also, rust or corrosion may occur inside the equipment. M526-E024 R-300/RC-300 OPERATION MANUAL...
SONIALVISION G4 system) Be sure to use the power supply specified in the operation manual. Using a power supply other than the one specified may cause equipment malfunction or serious accidents such as fire, smoke emission, or explosions. R-300/RC-300 OPERATION MANUAL M526-E024...
1.4.4 Grounding Grounding resistance: less than 100 Ω Be sure to connect the equipment only to a (commercial) power outlet with a ground terminal. If the outlet does not have a ground terminal, electric shock may occur. M526-E024 R-300/RC-300 OPERATION MANUAL...
For Use in Flammable Atmosphere Do NOT use the equipment or system in the presence of flammable anesthetics gas. It may cause an explosion. Classification of Installation Type Permanently installed equipment R-300/RC-300 OPERATION MANUAL M526-E024...
Language English Experience All operators should be informed about operation before using the device. Corrected visual acuity is over 20/28 (6/9). Permissible impairments Impaired hearing or color blindness is permitted to operate the equipment. M526-E024 R-300/RC-300 OPERATION MANUAL...
On name plate Manufacturer On the right side of the equipment Refer to operation manual The Type B Applied Part Symbol is indicated to meet the requirements of IEC 60601-1:1988+A1:1991+A2:1995 as this may come into contact with patients. R-300/RC-300 OPERATION MANUAL M526-E024...
Lamp button P.25, P.39 On pressing it again, the collimator lamp and line marker are turned off. Irradiation field adjustment knob (for H-leaves) Used to adjust the irradiation field size. P.26 Irradiation field adjustment knob (for V-leaves) R-300/RC-300 OPERATION MANUAL M526-E024...
Page 41
Tape measure Used to measure the exposure distance. P.24 Displays a cross marking in the center of the irradiation field Reticule when the lamp has been turned on. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 42
Lamp button P.25, P.39 On pressing it again, the collimator lamp and line marker are turned off. Irradiation field adjustment knob (for H-leaves) Used to adjust the irradiation field size. P.26 Irradiation field adjustment knob (for V-leaves) R-300/RC-300 OPERATION MANUAL M526-E024...
Page 43
Tape measure Used to measure the exposure distance. P.24 Displays a cross marking in the center of the irradiation field Reticule when the lamp has been turned on. M526-E024 R-300/RC-300 OPERATION MANUAL...
2 Part Names and Functions Operation Panel (R-300) 2.2.1 When Combined with General Radiography Equipment Name Function Filter selection button Enables you to select from among three types of X-ray filter. P.35 On pressing this button its LED lamp lights and the interlock...
Page 45
2.2 Operation Panel (R-300) Name Function Used to perform radiography without changing the size of the irradiation field, for example examining the cervical vertebrae from four directions. When combined with a system that allows the irradiation field size to be changed for each exposure and the preset radiography...
Page 46
Contact your Shimadzu service representative when changing the location of any button on the operation panel. The preset values to be loaded by the SID change button can be changed. Contact your Shimadzu service representative if you wish to change them. R-300/RC-300 OPERATION MANUAL M526-E024...
2.2 Operation Panel (R-300) 2.2.2 When Combined with Fluoroscopy Equipment (FLEXAVISION system) Name Function Enables you to select from among three types of X-ray filter. Filter selection button P.35 This button cannot be used when performing fluoroscopy. Used to make the irradiation field adjustment knobs effective Knobs/remote selection with the DR technique.
When DR technique is selected, the LED comes on if this button is pressed only if the button is arranged on the console. Contact your Shimadzu service representative when changing the location of any button on the operation panel. R-300/RC-300 OPERATION MANUAL M526-E024...
When using RC-300, buttons are displayed on the touch panel of the X-ray tube support CH-200. For details, refer to the operation manual of the X-ray tube support CH-200. CH-200 Operation Manual ^ Reference Display Panel (R-300) Name Function Irradiation field The irradiation field is displayed here.
Page 50
2 Part Names and Functions This page is intentionally left blank. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 51
Chapter Operation Contents Turning the Power ON/OFF ..... 22 Setting the Irradiation Field ..... . 23 Auto-Filter .
This positioning operation ends about 10 seconds after the power has been turned ON. When the positioning operations of each of the motors have ended the irradiation field size is displayed and the collimator can be used. R-300/RC-300 OPERATION MANUAL M526-E024...
With General Radiography or Simple Cassette Radiography The operator must take due care to avoid unnecessary exposure, and set the irradiation field before starting radiography. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 54
• The tape measure is marked out in both inches and centimeters. • The scale on the tape measure shows the distance from the focal spot of the X-ray tube. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 55
The high temperature on the rear of the collimator may burn injury. If the message "Lamp-ON time limited" appears on the display panel, the lamp does not light. After the message disappears, turn the lamp on. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 56
It is possible to exchange the functions of the knobs at the right and left sides, and to set them to operate with the same turning directions as on other collimators manufactured by Shimadzu (the R-20J/R-30H and so on). The maximum size setting of the irradiation field can be changed.
Page 57
3.2 Setting the Irradiation Field When using R-300, the time that the lamp stays on can be set from 5 to 30 seconds at installation. When using RC-300, the time that the lamp stays on can be set from 5 to 60 seconds at installation.
Page 58
X-ray field as possible, before performing X-ray exposure. When the collimator is turned, corners of the X-ray field may be missed depending on the degree of turning, the amount of narrowing, and the maximum X-ray field of the X-ray tube unit. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 59
If the stopper is left not fully tightened, the angle of the collimator may change. When the detent (option) is fitted, as you turn it back into place, you will sense a click that makes it easy to locate the position. M526-E024 R-300/RC-300 OPERATION MANUAL...
The radiography procedure is the same as for general radiography or simple cassette radiography when the collimator is combined with general radiography equipment. "With General Radiography or Simple Cassette Radiography" P.23 ^ Reference R-300/RC-300 OPERATION MANUAL M526-E024...
Page 61
The irradiation field size can be adjusted within a range that doesn't exceed the size of the irradiation field of the FPD or image intensifier. Irradiation field adjustment knob for the H-leaves Lamp button Irradiation field adjustment knob for the V-leaves V-leaves H-leaves M526-E024 R-300/RC-300 OPERATION MANUAL...
P.30 adjustment knobs on the equipment. With General Radiography Always check the X-ray irradiation region using the collimator lamp. Irradiating a patient with X-rays outside the required region risks exposing the patient to unnecessary radiation. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 63
The irradiation field size can be adjusted within a range that doesn't exceed the size of the irradiation field of the FPD or image intensifier. Irradiation field adjustment knob for the H-leaves Lamp button Irradiation field adjustment knob for the V-leaves V-leaves H-leaves M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 64
When radiography conditions other than general radiography have been selected, it is not possible to enlarge the selected irradiation field size. Turning the Collimator This facility cannot be used when the collimator is used in combination with fluoroscopy equipment. R-300/RC-300 OPERATION MANUAL M526-E024...
• The auto-filter is installed when the collimator is shipped from the factory. • Notify your Shimadzu service representative before changing the type of X-ray filter. 3.3.1 Types of X-ray Filter The material composing the X-ray filter is indicated in the display panel.
With SONIALVISION G4 system using techniques that allow fluoroscopy (such as DR technique), it is possible to select the radiography filter. The information of the radiography filter appears blinking on the display panel and changes to the information of the fluoroscopy filter after 3 seconds. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 67
Chapter Options Contents Line Marker ........38 Dose Calculation Unit .
Looking at it directly could cause loss of sight. Only in cases where the line marker is required, operate the line marker shielding slide switch on the bottom face of the collimator to open the shutter. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 69
Operate the line marker shielding slide switch on the bottom face of the collimator to open the shutter. Line marker shielding slide switch Press the lamp button. The lamp and line marker will light. Lamp button M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 70
The lamp and line marker will go out automatically after the set amount of time even if you don't press the lamp button. When using R-300, the time that the lamp stays on can be set from 5 to 30 seconds at installation.
The area dosimeter must be periodically calibrated in order to maintain its precision in indicating dose area product. For details, contact your Shimadzu service representative. M526-E024 R-300/RC-300 OPERATION MANUAL...
(transversal direction of the tabletop). For details, refer to the operation manual of the system that is being used in combination with the collimator. This option is for fluoroscopy equipment. (It cannot be used with general radiography equipment.) R-300/RC-300 OPERATION MANUAL M526-E024...
If an error occurs, an error message appears in the relevant display location on the display panel for R-300. When using RC-300, an error message and the button are displayed on the touch panel of the X-ray tube support CH-200.
Page 75
Turn the power OFF and back ON. The motor driving polygonal collimator (C-leaves)/single-acting H mask may have failed, and the leaves cannot be moved to the set position. ERR2 Turn the power OFF and back ON. M526-E024 R-300/RC-300 OPERATION MANUAL...
Cause and Corrective Action Parts inside the collimator may have failed. A burning smell Turn the power OFF immediately and contact your Shimadzu service representative. The lamp circuit may have failed or the lamp may have blown. The lamp does not light.
When a user performs the maintenance check for himself or herself, please keep in mind his or her own safety. Shimadzu shall not be liable for any damage resulting from the checks other than those by Shimadzu or Shimadzu representative.
Daily Inspections (Inspections Carried Out by the User) Carry out the following inspection before starting work, and if there is any abnormality contact your Shimadzu service representative or the qualified personnel approved by them. 6.2.1 Checklist for Daily Inspection Date of inspection:...
6.2.3 Checking the Warning and Caution Labels Periodically (once a year) inspect the labels attached on the equipment. If any label is peeled or unreadable by stain or scratch, contact your Shimadzu service representative for replacement of a new one.
Wipe down all contact surfaces using a cloth moistened with an appropriate disinfectant or cleaner. Make sure the cloth is NOT too wet. If it is, liquid may enter into system electronics, causing failure or malfunction. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 82
Repeated disinfection over a long time may lead to discoloring and cracking on the equipment surface, and deterioration of rubber and plastic. If any abnormality is found on the equipment after disinfection, stop using the equipment immediately. Contact your Shimadzu service representative for repair. R-300/RC-300 OPERATION MANUAL...
Page 83
Rubbing alcohol, organic solvents or non-neutral detergents may lead to deformation or crack of resin parts such as reticule of the collimator. Wipe them off immediately if they adhere to the reticule of the collimator. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 84
When turning the power ON after cleaning, make sure the examination room is properly ventilated. Turning the power ON while any flammable gas remains in the examination room could lead to fire, smoke, explosion or electrocution. R-300/RC-300 OPERATION MANUAL M526-E024...
The inspections require good knowledge of the internal mechanisms and can also be dangerous. Contact your Shimadzu service representative to request a periodic inspection. It is recommended to conduct periodic inspections every 6 months. A fee is charged for periodic inspections after expiry of the warranty periods.
Page 86
6 Maintenance This page is intentionally left blank. R-300/RC-300 OPERATION MANUAL M526-E024...
For replacement parts of internal components, make sure to apply the cables supplied by Shimadzu. The use of non-cable devices, accessories, or cables other than those sold by Shimadzu as replacement parts for the internal components may result in increased emissions or decreased immunity of the equipment. R-300/RC-300 OPERATION MANUAL...
Page 89
Essential performances of this equipment are as followings; • Adjustment of X-ray field If an essential performance of this equipment is degraded or lost due to electromagnetic disturbances, the following may occur. • X-ray irradiation field exceeds the accuracy specified in IEC 60601-1-3. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 90
Note *1: The ISM bands between 150 kHz and 80 MHz are 6.765 to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 91
*2: As an alternative to FM modulation, the carrier may be pulse modulated using a 50 % duty cycle square wave signal at 18 Hz. While it does not represent actual modulation, it would be worst case. M526-E024 R-300/RC-300 OPERATION MANUAL...
231 (W) 317 (L) 259 (H) mm External Dimensions Mass Approx. 10 kg 24 VDC, 150 VA (without polygonal collimator (C-leaves)) Power Supply 240 VA (with polygonal collimator (C-leaves) or when combined with SONIALVISION G4 system) 45 Turning Mechanism R-300/RC-300 OPERATION MANUAL M526-E024...
Page 93
Distance between the focus and fitting face 60 mm External Dimensions 231 (W) × 343 (L) × 280 (H) mm Mass Approx. 10 kg Power Supply 24 VDC, 150 VA (without polygonal collimator (C-leaves)) 45 Turning Mechanism M526-E024 R-300/RC-300 OPERATION MANUAL...
P.42 field from rectangle to octagon. Use the single-acting H mask to block a half of Single-acting H Mask the irradiation field/light field horizontally P.42 (transversal direction of the tabletop). R-300/RC-300 OPERATION MANUAL M526-E024...
7.3 Labels Labels 7.3.1 R-300 125678 Name Plate 1 Name plate Power supply capacity: 150 VA Type Power supply capacity: 240 VA Type Warning Labels 2 Warning label 3 Caution label (for Halogen Lamp Type) M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 96
4 Caution label (option: Only if laser model number is LDS163LSS.) Others 5 Label, cTÜVus mark (For USA) 5 Label, CE (For CE) 6 Refer to operation manual 7 Type B Applied Part Label 8 UDI Label R-300/RC-300 OPERATION MANUAL M526-E024...
Page 97
IEC 60601-1:1988+A1:1991+A2:1995 as this may come into contact with patients. Package for Transportation The following shows the label on the package for transportation, describing about storage environment and transport environment. Symbol Location Meaning Package Keep away from rain in during transport. M526-E024 R-300/RC-300 OPERATION MANUAL...
Page 98
7 Specifications 7.3.2 RC-300 1245 Name Plate (Identification label) 1 Name plate Warning Labels 2 Warning label 3 Caution label (option: Only if laser model number is LDS163LSS.) R-300/RC-300 OPERATION MANUAL M526-E024...
Page 99
4 Refer to operation manual 5 UDI Label Package for Transportation The following shows the label on the package for transportation, describing about storage environment and transport environment. Symbol Location Meaning Package Keep away from rain in during transport. M526-E024 R-300/RC-300 OPERATION MANUAL...
Statement of Compliance [For Europe] 7.4.1 Regulatory Information For Europe: The product complies with the requirement of the Medical Device Directive 93/42/EEC and RoHS Directive 2011/65/EU. Product Name: COLLIMATOR TYPE R-300 Manufacturer: SHIMADZU CORPORATION Medical Systems Division 1, NISHINOKYO-KUWABARACHO, NAKAGYO-KU, KYOTO, 604-8511, JAPAN...
7.5 Statement of Compliance with Standards Statement of Compliance with Standards • X-RAY EQUIPMENT for RADIOGRAPHY and RADIOSCOPY ... COLLIMATOR TYPE R-300 IEC 60601-2-54:2009+A1:2015+A2:2018 • X-RAY EQUIPMENT for RADIOGRAPHY and RADIOSCOPY ... COLLIMATOR TYPE R-300 EN 60601-2-54:2009+A1:2015+A2:2019 Manufacturer Information Manufacturer:...
Installation Manual. The operator must set the focal spot to skin distance as large as possible in order to keep the absorbed does to the patient as low as reasonably achievable. R-300/RC-300 OPERATION MANUAL M526-E024...
Page 104
Appendix Check List for Troubleshooting Fill following blanks and contact your Shimadzu service representative. Hospital: Phone: FAX: Serial Number: Date of Installation: Item Check Did the malfunction injure anyone? When did the malfunction occur? Did the system malfunction suddenly? Were there any indications of a problem beforehand? ...
Page 105
Signature Signature • At the time of installation or when changing the designated operator, your Shimadzu service representative will explain the safety instructions described in this manual to the maintenance and operation staff, and record the information. •...
Page 106
Appendix This page is intentionally left blank. R-300/RC-300 OPERATION MANUAL M526-E024...