DMT 560TP User Manual

Tissue puller system
Hide thumbs Also See for 560TP:

Advertisement

Quick Links

TISSUE PULLER SYSTEM - 560TP
USER MANUAL VOL. 3.1
1
Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 560TP and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for DMT 560TP

  • Page 1 TISSUE PULLER SYSTEM - 560TP USER MANUAL VOL. 3.1 Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 2: Trademarks

    DMT560TP-II functions and the Controller Software. No part of this document may be reproduced by any means without the prior written permission of DMT A/S. Copyright © 2019 DMT A/S Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 3: Safety

    Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 4: Ec Declaration Of Conformity

    Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional RF sources), as these may interfere with the proper operation. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 5: Table Of Contents

    3.4 GRAPH TYPE SUBMENU ................. 19 3.5 DRAIN CHAMBER SUBMENU ................20 3.6 PULLER CONTROL SUBMENU ................21 3.7 TEMPERATURE CONFIG SUBMENU ..............22 3.8 FORCE SUBMENU ..................22 Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 6 ..............32 5.4 HOW TO REPLACE THE FORCE TRANSDUCER: ............33 5.5 MOTOR OVERLOAD: ..................33 APPENDIX 1: HOW TO READ THE START POSITION ........35 Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 7: Features & Benefits

    The Puller is specifically designed to perform tensile strength tests of different tissue samples. Force and pull length are plotted in the PC Software to create a force/displacement curve. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 8: Unpacking The Tissue Puller System

    UNPACKING THE TISSUE PULLER SYSTEM 560TP-II Take a few minutes to carefully inspect the new 560TP-II for damage which may have occurred during shipping and handling. If you suspect any kind of damage, please contact DMT immediately, and the matter will be pursued soon as possible. If the packing material appears damaged, please retain it until a possible claim has been settled.
  • Page 9: Chapter 1 - System Overview

    CHAPTER 1 – SYSTEM OVERVIEW Figure 1.1 Tissue Pulling System 560TP - II (front view) Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 10: Setting Up Step-By-Step

    Install the 560TP MyoPULL Control Software on the PC. • Connect the Power adapter to DMT560TP-II. • Switch the Power ON. Run the Control Software on the PC. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 11: Getting Started With The Myopull Software

    When the DMT560TP-II Software is started for the first time, you must type in the license code. Alternatively, continue in Trail Mode (with limited functions). Figure 1.3 License Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 12: Chapter 2 - Tissue Puller 560Tp-Ii Menus On The Interface

    Read the mechanical position from the micrometer positioner and enter it here as shown in Fig 2.1. NOTE: SEE APPENDIX 1 TO LEARN HOW TO READ THE POSITION CORRECTLY. Figure 2.1 Enter mechanical position Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 13: Force Calibration

    However, in order to ensure that the Puller is working at the highest performance, DMT recommends that a new Force calibration is performed before the first use of the Puller. The calibration procedure is described in section 2.2 below.
  • Page 14 Follow the procedure in the popup windows. Figure 2.3 Force Calibration Menu Page 1 Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 15 If the values are out of this range, the force transducer is broken and needs to be replaced. How to replace a broken force transducer is described in section 5.4. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 16: Chapter 3: Description Of Software Functions For Pulls In Myopull

    It is possible to select which icons you want to show on the top Toolbar. Just open the selection window with a click on the down arrow and select. Fig. 3.2 Toolbar Customization Icon Fig. 3.3 Puller menu. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 17: Pull Submenu

    Box for entering the start Position. The green arrow makes the Puller move to the start position. Start Position can be 0 to 50000um Fig. 3.6 Start Position Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 18: Stress - Strain Presets Submenu

    INTO THE SYSTEM TO CALCULATE THE RIGHT STRESS-STRAIN GRAPH, SEE THE EXAMPLE FOR DIFFERENT SHAPES OF SEGMENTS, AND WHERE TO PUT THE RIGHT VALUES. Fig. 3.7 Stress-Strain Presets for Tissue Strips Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 19: Graph Type Submenu

    In Displacement-Force Graphs, the X-axis is shown in μm and the Y-axis in mN. In the Strain - Stress Graphs the X-axis is Strain (Displacement/(Length start)) and the Y-axis is Stress Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 20: Drain Chamber Submenu

    The delay can be 0-10, 20 or 30 sec. Pressing the green ‘play’ icon will activate the drain valve for the selected time and then close again. Fig. 3.11 Valve Delay Selection Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 21: Puller Control Submenu

    3. Reset the physical position (start position) to the actual micrometer screw readings. See appendix Fig. 3.13 Manual Motor Selections 1 on how to read the start position. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 22: Temperature Config Submenu

    The Force submenu for Zero the Force and calibrate the Force. See the Force Calibration section earlier in this manual chapter 2.2, on how to perform the Force calibration. Fig. 3.17 The Force submenu Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 23: Chapter 4: Myopull Software Functions To Analyze The Pull Data

    Figure 4.1 Example of a Tensile Strength Test Curve. The force (mN)readings are shown on the Y-axis, and the displacement in μm is shown on the X-axis. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 24 Max. Force: The max. Force measured under the Pull in mN. Max. Stress: (Only use in Stress-Strain Pulls). The max. Stress measured under the Pull in mN/mm2. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 25 Figure 4.3 Experiment Overlay of Multiple Pull Experiments Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 26 3. Export data in Excel or CSV format of the highlighted area. Copy to Clipboard 4. Export highlighted area to the data sheet. 5. Show the highlighted data in the data sheet. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 27 • Copy to clipboard. • Export to data sheet. • Show data sheet. Fig. 4.7. Data sheet after export of the highlighted area. • Delete data sheet. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 28 Fig. 4.5 Highlighted Area Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 29 Fig. 4.6 Zoom of the Highlighted Area Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 30 Fig. 4.7 Data Sheet Showed after Export of Highlighted Area Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 31: Chapter 5: Maintenance

    If any other problems related to the force transducer are encountered, please contact DMT for advice or further instructions. Fig. 5.1 Illustration of the transducer compartment. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 32: Cleaning Of The Dmt560Tp-Ii

    • 5.3 FORCE TRANSDUCER MAINTENANCE: The DMT560TP-V2 is a very delicate and sophisticated piece of research equipment. DMT recommends that the following sections are read carefully and that the instructions are followed at all times. Force transducer: The force transducer is the most delicate and fragile component of the DMT560TP-V2. Extreme care must be used when handling or touching the force transducer.
  • Page 33: How To Replace The Force Transducer

    By hand, until torque limiter is in place again, it must be in place after max. one turn of the Micrometer Thimble. NOTE: Important after an overload remember to enter the new start position see fig. 3.2 and Appendix 1. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 34 Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 35: Appendix 1: How To Read The Start Position

    0.5 mm. Each division on the thimble scale is 10 μm. If the thimble scale falls between two lines, then a number between 0 and 10 μm must be approximated. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 36 10000 μm B. No additional mark visible: 0 μm C. Thimble reading: 380 μm Total reading: 10380 μm Figure A1.2 Example 1: reading = 10380 μm Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 37 16000 μm B. One additional mark visible: 500 μm C. Thimble reading: 280 μm Total reading: 16780 μm Figure A1.3 Example 2: reading = 16780 μm Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
  • Page 38 Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...