CHAPTER 4 - CLEANING AND MAINTENANCE CLEANING THE MYODYNAMICS MUSCLE STRIP MYOGRAPH MAINTENANCE OF THE FORCE TRANSDUCER MAINTENANCE OF THE LINEAR SLIDES APPENDIX 1 - READING A MILLIMETRE MICROMETER Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
Each chamber contains supports with mounting hooks to facilitate the mounting of muscle strips with sutures. DMT is also able to deliver a large variety of different mounting supports. Please contact DMT Sales to acquire more information about available mounting supports for the MyoDynamics Muscle Strip Myograph.
FROM USING A PERMANENT MARKER. STORE THE SUPPORTS IN THE PROVIDED PLASTIC CASE. NUMBERING THE SUPPORTS WILL SAVE TIME WHEN THE SUPPORTS ARE REPLACED. IT WILL REDUCE THE AMOUNT OF ADJUSTMENTS REQUIRED AFTER EACH CHANGE. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
Prior to shipping the MyoDynamics Muscle Strip Myograph is subjected to two days of continuous testing, including a final force calibration. However, DMT recommends that a new force calibration is performed before using the MyoDynamics Muscle Strip Myograph for the first time. The force calibration procedure is described in detail in chapter 3.6.1 Force calibration in MyoDynamics Muscle Strip Myograph System -...
2. Position the sutures appropriately on both hooks so that they are ready for use when the muscle is mounted. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
Page 9
In User Gide “How to measure isometric force of striated muscle in vitro in an 820MS or 840MD Myograph”, we have examples with tips, tricks and how to do a proper ‘normalization’ of the mounted muscle before doing the real experiments.
If red or brown discolorations appear on the chamber sides or on the supports, the following cleaning procedure will work in most cases: Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
Page 11
THE CHAMBER AND CLEANED INDIVIDUALLY TO ASSURE PROPER CLEANING OF ALL SUPPORT SURFACES. NEVER SOAK THE SUPPORTS IN ANYTHING STRONGER THAN 8% ACETIC ACID FOR EXTENDED PERIODS OF TIME (I.E. SEVERAL HOURS OR OVERNIGHT). Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
TRANSDUCER PINHOLE IS CHECKED AND SEALED AT LEAST ONCE A WEEK, ESPECIALLY IF THE MYOGRAPH IS USED FREQUENTLY. DMT TAKES NO RESPONSIBILITIES FOR THE USE OF ANY OTHER KINDS OF HIGH VACUUM GREASE OTHER THAN THE ONE SUPPLIED BY DMT.
Page 13
AND the transducer cannot be recalibrated, the force transducer is broken and needs to be replaced. If any other problems related to the force transducer are encountered, please contact DMT for advice or further instructions. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
Page 14
Figure 4.3 - Myograph without the bottom plate 4. Disconnect the connector to the transducer. 5. Remove the two transducer screws that hold the transducer housing, see figure 4.4. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
Page 15
11. Place some vacuum grease at the edge of the bottom plate. Place and tighten the bottom plates. IMPORTANT: CALIBRATE THE NEW FORCE TRANSDUCER BEFORE PERFORMING A NEW EXPERIMENT. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...
Check the linear slides (under the black covers) for grease at least once a week. In case of insufficient lubrication, grease the slides with the “Grease for Linear Slides” included with your system. Danish Myo Technology A/S | Tel.: +45 87411100 | Email: support@dmt.dk | Web: www.dmt.dk...