Shenzhen Linshang Technology Co., Ltd. LS212 Ultrasonic Thickness Gauge User Manual V2.3 Please read this manual carefully before using and reserve it for reference.
Shenzhen Linshang Technology Co., Ltd. I. Product introduction Ultrasonic thickness gauge adopts the principle of the reflected plus ultrasonic measurement, which is specialized in thickness measurement of materials that can transmit ultrasonic waves such as metals (e.g. steel, aluminum, copper, etc.), plastics, ceramics, glass, etc. The instrument employs a professional design of time-chip with a resolution of up to 0.001 mm.
Shenzhen Linshang Technology Co., Ltd. 2. Host parameters Resolution 0.8-10mm:0.001mm 10-100mm:0.01mm 100-350mm:0.1mm Sound Velocity Range 1000~9999m/s Display 240×160 dot matrix LCD Unit mm/inch Power Supply 2pcs of 1.5V AA alkaline battery Host Size 142*72*28 mm Weight About 230g Operation Temperature Range -10~50℃,0~85%RH(No condensation)...
Shenzhen Linshang Technology Co., Ltd. effective measured values, statistics can be restarted before measurements are taken. 7. The instrument can adjust sound velocity in 3 ways: setting sound velocity based on material/ thickness/ manually. 8. The instrument intelligently identifies the model of the probe, and the instrument adapts the display interface according to the probe type.
Shenzhen Linshang Technology Co., Ltd. ◆ Instrument malfunction and it need to be repaired returning to the factory. 3. General Measurement Apply coupling fluid on the surface of the material, press the probe of the instrument tightly and keep it still, then the user can get the thickness of the material.
Shenzhen Linshang Technology Co., Ltd. workpiece coating. This function is achieved by measuring two consecutive bottom echoes of the substrate, which can penetrate the coating thickness of 0.2~2.5mm and measure the thickness of the substrate (45# steel) of 4~60mm. In the system setting interface there is a switch selection for the penetration coating function, set the...
Page 7
Shenzhen Linshang Technology Co., Ltd. Language: Short press the button to enter the language selection. These two buttons are available to select language. A short press on the button completes the setting. Unit: Short press on the button to enter the unit selection. These two buttons are available to select units.
Page 8
Shenzhen Linshang Technology Co., Ltd. (2) Sound velocity Set Short press the buttons to select [Set by material/Set by thickness/Manual input/Return]. A short press on the button confirms the selection. Set by material: The user can set the sound velocity according to the known material. Short press the...
Page 9
Shenzhen Linshang Technology Co., Ltd. prompt interface "Please press the probe tightly against the material with known thickness". Press the probe of the instrument tightly against the surface of the material with couplant and keep it still. The instrument will automatically jump to the interface of "Enter the actual thickness of the material" where...
Page 10
Shenzhen Linshang Technology Co., Ltd. (4) Exit A short press on the button exists the main menu and enters the measurement interface. (5) Shutdown A short press on the button turns off the instrument. 6. Check measurement records In the measurement mode, short press the buttons to enter the browsing interface to view historical data.
Shenzhen Linshang Technology Co., Ltd. V. Bluetooth connection APP function The instrument has built-in Bluetooth communication module, which can be connected to cell phone APP. VI. APP Function 1. APP Installation The measuring instrument APP supports 7.0 and above Android operating system, search “UT and HL” on google play, follow the instructions to download and install the APP.
Page 12
Shenzhen Linshang Technology Co., Ltd. Bluetooth connection interface If the APP has a bound Bluetooth device, the logo interface will automatically search and connect the bound Bluetooth device after 3 seconds, and the successful connection will automatically enter the "measurement" interface.
Shenzhen Linshang Technology Co., Ltd. Measurement interface 3. Measurement interface (1) Bluetooth connection status The APP is connected successfully, the upper right corner of the interface shows "Connected", and the APP is disconnected, showing "Not connected". When not connected, tap it, and the APP will automatically reconnect the bound device.
Page 14
Shenzhen Linshang Technology Co., Ltd. (5) Instrument information Display instrument setup material, unit, connected probe type. (6) Create groups Click Create Groups icon to automatically save the previous group and generate a new group. (7) All groups Click All Groups icon to select a set of data for open and delete operations.
Shenzhen Linshang Technology Co., Ltd. which requires the use of coarse crystalline probes for measurement.5. The probe is easy to be scratched on the rough surface. Try to reduce the sliding of the probe on the rough surface. If the probe is seriously worn, it should be replaced in time.