Page 14
– Smaller CureWraps will need less water. 1. Remove this 1. Remove this 2. Place the CritiCool unit near the bed and lock the front wheels cap to fill cap to fill 3. Power ON ON switch - after a short Self Test, Mode Select screen appears...
Page 22
• CritiCool cannot thermoregulate without a valid Core CORE CORE readout – Therefore, water will not flow into CureWrap until the core probe is connected to CritiCool and the core probe is attached to the patient NOTE: An accurate core temperature reading is essential for proper thermoregulation.
Page 23
SETTING UP CRITICOOL – CONNECT HOSES 1. Connect metal fittings on blue hose to metal fittings on the front of CritiCool – Listen for a click – Tug slightly to ensure securely inserted Water in coupling Water in coupling connector...
Page 29
SELECTING SET POINT TEMPERATURE Set Point = Patient’s Target Temperature • The CritiCool cools/warms the patient to reach and maintain the preset Set Point Adjust Adjust • TTM Mode Default Set Point: TTM Mode Default Set Point: TTM Mode Default Set Point:...
2. Select Standby o Water flow to the CureWrap will stop o CureWrap will deflate o CritiCool will continue to monitor patient’s temperature (as long as core temperature probe remains in place) To resume thermoregulation: 1. Touch MENU icon/button 2.
Page 42
– This mode has no rewarming steps – This mode features fewer alarms related to low core temperature • The CritiCool warms the patient to reach and maintain the selected Set Point • Default Set Point is the default set point for the preceding mode •...
Page 44
• The hourly rewarming rate is preset in the Settings screen (“Rewarming Step”) according to the hospital protocol. • The patient will be warmed to the Target Temperature (“Target Temp.”), the normothermic end temperature. The CritiCool will reach the Target Temp. Target Temp. gradually by raising Target Temp. Target Temp.
Page 50
– Positioning of the core temperature probe (depth and secureness) – Placement of the skin temperature probe (location and secureness) • Other factors that can be adjusted via CritiCool by the clinician: – CureWrap Positioning (see CureWrap Instructions for Use) –...
Page 54
ESPT334MG PTB Usage Label v5-2023.pdf. Refer to the dilution chart/usage table in the NaDCC manufacturer instructions to determine how much NaDCC to use. • A full water tank (six liters or 1.6 gallon) allows for optimal cleaning of the CritiCool water tank. Add sterile water (or 0.22 micron filtered water) as needed.
Page 61
Controlled Rewarming Mode via the Mode Select screen. After the user acknowledges the message on the screen, water will once again flow to the wrap and CritiCool will not alarm for that condition for 30 minutes.
Page 65
*After the user acknowledges the message on the screen, water will once again flow to the wrap and CritiCool will not alarm for that condition for 30 minutes. If after 30 minutes the alarm conditions apply, it will once again stop water flow and alert the user.
– CritiCool turns off thermoelectric coolers (TECs) if water temperature >43°C (no water flow) – CritiCool shuts off if water temperature reaches 44°C with Halt 3 (no water flow) • Note: High Water Temperature does not control the water temperature; it alarms when the value is reached.
• Use sterile water or 0.22 micron filtered water only. • CritiCool cools the water upon startup. If normothermia is the goal, to avoid cooling the water, limit the time the device is turned on but not fully set up.