Free-Space Plasma Light Source
Chapter 8: Command Line Operation
8.3.8. LAT? – Bulb Module Temperature
Queries current lamp temperature
Return:
LAT=n
Where n is the value of lamp temperature in °C.
8.3.9. MOD=n; MOD? – Device Control Mode
The MOD command is used to set or query the device control mode. There are two available modes: local and
external.
MOD=n
Sets the device control mode. The value of n can be one of the following:
n=1: Local
n=2: External
MOD?
Queries current device control mode.
Return:
MOD=n
Where n is current device control mode.
If n is out of current list range, the controller returns CMD_ARG_RANGE_ERR.