Lat? - Bulb Module Temperature; Mod=N; Mod? - Device Control Mode - THORLABS HPLS301 User Manual

High-power plasma light source, free space
Table of Contents

Advertisement

Free-Space Plasma Light Source
Chapter 8: Command Line Operation
8.3.8. LAT? – Bulb Module Temperature
Queries current lamp temperature
Return:
LAT=n
Where n is the value of lamp temperature in °C.
8.3.9. MOD=n; MOD? – Device Control Mode
The MOD command is used to set or query the device control mode. There are two available modes: local and
external.
MOD=n
Sets the device control mode. The value of n can be one of the following:
n=1: Local
n=2: External
MOD?
Queries current device control mode.
Return:
MOD=n
Where n is current device control mode.
If n is out of current list range, the controller returns CMD_ARG_RANGE_ERR.
Page 23
Rev C, June 12, 2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HPLS301 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Hpls301m

Table of Contents