Site Specifications

Plumbing Requirements

Water Supply:
Adequate water supply and drainage is the customer's responsibility, and must be installed to specifications by the
customer before installation of the PROCESSOR begins.
A WATER SUPPLY PANEL for plumbed systems is available from Kreonite (WPS-A), and is recommended.
Kodak will install flexible supply line to PROCESSOR and an electrical connection to the SOLENOID. The WATER
SUPPLY PANEL should be wall-mounted within 4.9 metres (16 feet) of the rear of the PROCESSOR.
Technical support and warranty of WATER SUPPLY PANEL is the responsibility of Kreonite.
Water to a plumbed system must meet the following specifications:
Flow rate: 3.8-20 liters per minute (1-5 gallons per minute)
Temperature: Hot- 59 C (120 F) minimum temperature. Cold-27 C (80 F) maximum temperature. Practically, the
water temperature should be adjusted to 35C (95 F) to minimize the time necessary to bring the PROCESSOR to
operating temperature.
Waste Water:
A floor drain should be located convenient to the PROCESSOR.
The drain system should conform to local specifications for silver-laden bleach-fix, and stabilizer or waste water.
Kodak will install a flexible drain TUBE to the PROCESSOR. The customer will be responsible for connection of
the flexible drain TUBE to a locally approved drain system.
In unplumbed systems, the PROCESSOR is supplied with 10-litre CONTAINERS for easy transport of waste to the
site waste facilities. Waste CONTAINERS will be located under the PROCESSOR.
A sink, 61 cm (24 in.)wide or larger, located near the PROCESSOR, is recommended for use during cleanup.

Venting Requirements

For a roll-to-processor system, 10 complete changes of air in the room per hour are required.
For a roll-to-roll printer, no venting is required.

Other

Telephone: The customer must supply a telephone line (with RJ11C modular phone jack) within 3 metres (10 feet)
of the PRINTER to allow remote diagnostic support via modem. A high-grade analog service line is recommended.
Densitometer: An X-Rite DTP-36 densitometer is strongly recommended for calibration of the PRINTER and for
process control.