uebe visocor OM60 Instructions For Use Manual

uebe visocor OM60 Instructions For Use Manual

Upper arm blood pressure monitor

Advertisement

Upper arm
blood pressure monitor
Instructions for Use

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the visocor OM60 and is the answer not in the manual?

Questions and answers

Summary of Contents for uebe visocor OM60

  • Page 1 Upper arm blood pressure monitor Instructions for Use...
  • Page 2: Table Of Contents

    Table of contents Intended use Thank you for choosing the visocor OM60 upper arm blood pressure monitor (also referred to in the following as the unit). Safety instructions Important instructions for patients These instructions are intended to help the user to use the unit safely...
  • Page 3: A Intended Use

    Intended use Safety instructions 1. Important instructions for patients The intended purpose of this device is to enable adults to automati- cally measure their own systolic blood pressure, diastolic blood pres- sure and pulse at their upper arm. The device can also be used by •...
  • Page 4: Important Technical Details

    Safety instructions Safety instructions • The readings measured using this device do not provide a diagno- – When using the device with a mains adapter, please only use sis. They are no substitute for visiting a doctor. Do not rely solely the visocor U2MC mains adapter, which has been specially on the result obtained;...
  • Page 5: C Operating The Unit

    Operating the unit Operating the unit 1. Unit description 2. Displays Cuff socket Air hose SYS = Systole (upper blood Irregular pulse waves pressure value) Displays Cuff Memory code DIA = Diastole (lower blood Memory recall button Start/Stop button pressure value) Date/time Connecting socket for Battery compartment...
  • Page 6: Important Usage Information

    Operating the unit Operating the unit 3. Important usage information • If the unit is to be operated from the mains power supply, the cable plug of the mains adapter (not included in the scope of de- The measurement results of automated blood pressure monitors can livery) must be inserted into the connection socket on the right- be influenced by the measuring location, posture, previous exertion hand side of the unit.
  • Page 7: Setting The Date/Time

    Operating the unit Operating the unit 6. Setting the date/time • The artery marker on the cuff must be placed over the artery that runs down the inside of • With the device switched off, the arm. press and hold the start/stop button until the four-digit year •...
  • Page 8: Measuring Blood Pressure

    Operating the unit Operating the unit 9. Measuring blood pressure Systolic blood Diastolic blood pressure pressure Classification = upper value = lower value • Switch on the unit by pressing mmHg mmHg the Start/Stop button. The full display appears (Figure 1). strong hypertonia (level 3) ≥...
  • Page 9: Using The Memory

    Operating the unit Operating the unit 12. Using the memory • Press the MEM key in the shut-off state to get to the memory. Measured results are automatically stored in the memory. The mem- ory can store up to 120 results and the average value. The average •...
  • Page 10: D What You Should Know About Blood Pressure

    What you should know about blood pressure Technical information 1. Systolic and diastolic blood pressure values 1. Error messages Failure encoun- Possible cause Corrective action The cardiovascular system has the important function of supplying tered all organs and tissues in the body with sufficient amounts of blood Display Er 1 Sensor error or faulty If the fault occurs several...
  • Page 11: E Technical Information

    Device repairs may only be performed by the manufacturer or an ex- Expected service life: 5 years pressly authorised body. Please contact: Operating mode: Continuous operation UEBE Medical GmbH info@uebe.com Classification: Internal power supply by battery Bgm.-Kuhn-Str. 22 www.uebe.com 97900 Külsheim, Germany 4.
  • Page 12: Technical Inspection / Calibration Check

    “Medical Products Operator Reference number = item number Ordinance” (Medizinprodukte-Betreiberverordnung). Serial number This can be performed either by UEBE Medical GmbH, an authori- ty responsible for metrology or an authorised maintenance service. Protected against solid foreign particles with a diameter of IP21 Please refer to your national regulations.
  • Page 13 Technical information Technical information Guidance and manufacturer’s declaration - electromagnetic immunity Portable and mobile HF communications equipment can affect the unit. Use of the unit in conjunction with non-approved accessories The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
  • Page 14: Maintaining The Unit

    Technical information Maintaining the unit Guidance and manufacturer’s declaration - electromagnetic emissions • Wash your hands after every measurement. If the device is used The device is intended for use in the electromagnetic environment specified below. The by more than one person, wash your hands before and after customer or the user of the device should ensure that it is used in such an environment.
  • Page 15: Warranty

    The cost of returning the device to our factory shall be borne by the sender. Complaints that are sent back without prepayment will not be accepted by UEBE. The statutory claims and rights of the buyer against the seller (for example, defect claims, producer liability) are not restricted by this warranty.
  • Page 17 25060 PZN-16259941 Medical aid code number: 21.28.01.2159 visocor and UEBE are internationally protected trademarks of UEBE Medical GmbH info@uebe.com Bgm.-Kuhn-Str. 22 www.uebe.com 97900 Külsheim Germany Subject to technical modifications. Reprinting, including of extracts, is prohibited. © Copyright 2022 UEBE Medical GmbH www.visocor.de...