Important Information:
NOTE: The
GastroCH
ECK® Gastrolyzer®
should be wiped down before and after each test. For a full list
4
of recommended cleaning wipes and solutions, please visit: https://www.bedfont.com/cleaning-bedfont-
monitors
How to perform a calibration:
2
3
5
1
7
8
To perform a calibration test, firstly make sure that the
GastroCH
ECK® Gastrolyzer® is plugged into the mains and to turn
4
on the monitor. Please note calibration is only possible if the
monitor has been switched on for 1 hour
3
Press the calibration icon to start the calibration
5
Connect the direct sample line to GastroCH
there should be an audible click
2
Press the settings icon to open the settings menu
4
Allow the GastroCH
If the monitor is too cold or hot please wait for it to come to
ambient temperature before continuing
6
To perform a calibration using gas, firstly attach the gas canister
ECK® monitor,
4
to the gas regulator by screwing the canister into the regulator
ECK® monitor to zero.