Product Specifications - Braun IRT6525KO Manual

Ear thermometer
Table of Contents

Advertisement

Available languages

Available languages

Product specifications

Displayed temperature range:
Operating ambient temperature range:
Storage temperature range:
Operating relative humidity
Storage relative humidity
Display resolution:
Accuracy for displayed temperature range
35 – 42 °C (95 – 107 .6 °F):
Outside this range:
Clinical repeatability:
Battery Type:
Battery life:
Service life:
Accessory:
The measured temperature is converted to an oral equivalent value .
This thermometer is specified to operate at 1 atmospheric pressure or at altitudes with an atmospheric pressure
up to 1 atmospheric pressure (700 - 1060 hPa) .
40 °C
10 °C
Equipment with
See Instruction for use
Operating temperature
type BF applied parts
LOT
SN
Caution
Batch code
Serial number
Manufacturer
Date of manufacture
Keep it dry
Subject to change without notice .
This medical device carries the CE mark and is manufactured in conformity with RoHS Directive 2011/65/EU and other
applicable directive and/or regulations as specified in the EU Declaration of Conformity . Full responsibility for the
conformance of the product to the standard is assumed by Kaz Europe Sarl, Place Chauderon 18, CH-1003 Lausanne,
Switzerland .
ASTM laboratory accuracy requirements in the display range of 37 to 39 °C (98 to 102 °F) for IR thermometers is ±0 .2 °C
(±0 .4 °F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and
E1112-86 is ±0 .1 °C (±0 .2 °F) .
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC . NOTE: Do not use this device in the
presence of electromagnetic or other interference outside the normal range specified in EN 60601-1-2 . For detailed
description of EMC requirements please visit website (Support / Document center) or contact service center .
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT .
This product contains batteries and recyclable electronic waste . To protect the environment, do not dispose of
it in the household waste, but take it to appropriate local collection points .
8
34 – 42 .2 °C
(93 .2 – 108 °F)
10 – 40 °C
(50 – 104 °F)
-25 – 55 °C
(-13 – 131 °F)
10-95% RH (noncondensing)
15-95% RH (noncondensing)
0 .1 °C or °F
Maximum Laboratory Error
±0 .2 °C
(±0 .4 °F)
±0 .3 °C
(±0 .5 °F)
±0 .14 °C
(±0 .26 °F)
2 X 1 .5V type AA (LR 06) batteries
2 years/350 measurements/ 100 minutes in nightmode
5 years
Lens filter, LF40/LF20, single use
10
Storage temperature
Operating humidity
REF
MD
Catalogue number
Medical device
Lens filter: Do not re-use
Calibration
The thermometer is initially calibrated at the time of manufacture . If this thermometer is used according to the use
instructions, periodic re-adjustment is not required . If at any time you question the accuracy of the temperature
measurement, please contact your authorized service center .
Serial Number and Manufacturing date are given by the UDI barcode .
Serial Number(S/N): please refer to the UDI barcode that starts with '(21)DDDYYKNNNNN'
Mfg . date: please refer to the UDI barcode that starts with '(11)YYMMDD'
An example: (10)210618, this product has been manufactured on 18 Jun 2021 .