Commissioning
42
9.5

Setting the operating language

9.5.1
Local display
Setting the operating language
To set the operating language, the display first needs to be unlocked:
1. Press the  key for at least 2 s.
 A dialog box appears.
2. Unlock the display operation.
3. Select the Language parameter in the main menu.
4. Press the  key.
5. Select the desired language with the  or  key.
6. Press the  key.
Display operation locks automatically:
• after 1 min on the main page if no key has been pressed
• after 10 min within the operating menu if no key has been pressed
Display operation - locking or unlocking
The  key must be pressed for at least 2 seconds in order to lock or unlock the optical
keys. Display operation can be locked or unlocked in the dialog that appears.
Display operation locks automatically (except in the SIL wizard):
• After 1 minute on the main page if no key has been pressed
• After 10 minutes within the operating menu if no key has been pressed
9.5.2
Operating tool
See the description of the operating tool.
9.6

Configuring the device

9.6.1
Commissioning with keys on the electronic insert
The following functions are possible via the keys on the electronic insert:
• Position adjustment (zero point correction)
The orientation of the device may cause a pressure shift
This pressure shift can be corrected by a position adjustment
• Setting the lower range value and upper range value
The pressure applied must be within the nominal pressure limits of the sensor (see the
specifications on the nameplate)
• Resetting the device
Carrying out position adjustment
1. Device installed in desired position and no pressure applied.
2. Press the "Zero" and "Span" keys simultaneously for at least 3 s.
3. When the LED lights up briefly, the pressure present has been accepted for position
adjustment.
Setting the lower range value (pressure or scaled variable)
1. The desired pressure for the lower range value is present at the device.