Soil weight, soil water tension, and thermal conductivity values will be recorded
automatically in a defined frequency (settings) until the measurement is stopped. See
HYPROP User Manual section
on how to stop measurements.
NOTE: Running a combined HYPROP 2 and VARIOS measurement, both measurements have to be stopped at the
same time.
3.3 POSTPROCESSING A MEASUREMENT
After measurement, the weight of the dry soil of the sample has to be determined because
the LABROS SoilView-Analysis software needs the dry soil weight to calculate the volumetric
water content based on the weight measurements.
1.
Disconnect the sensor cable from the VARIOS.
2.
Disconnect the HYPROP 2 from the magneto cable.
3.
Weigh an oven-safe frying pan and note the weight.
4.
Open the clips of the HYPROP 2 and remove the upper red retaining ring.
MEASUREMENT WITH LABROS SOILVIEW SOFTWARE
Figure 25 Measurement tab
2.6.7 and
VARIOS User Manual
section 2.7.5.2 for instructions
22
Need help?
Do you have a question about the HYPROP 2 and is the answer not in the manual?
Questions and answers