Soil weight, soil water tension, and thermal conductivity values will be recorded
automatically in a defined frequency (settings) until the measurement is stopped. See
HYPROP User Manual section
on how to stop measurements.
NOTE: Running a combined HYPROP 2 and VARIOS measurement, both measurements have to be stopped at the
same time.

3.3 POSTPROCESSING A MEASUREMENT

After measurement, the weight of the dry soil of the sample has to be determined because
the LABROS SoilView-Analysis software needs the dry soil weight to calculate the volumetric
water content based on the weight measurements.
1.
Disconnect the sensor cable from the VARIOS.
2.
Disconnect the HYPROP 2 from the magneto cable.
3.
Weigh an oven-safe frying pan and note the weight.
4.
Open the clips of the HYPROP 2 and remove the upper red retaining ring.
MEASUREMENT WITH LABROS SOILVIEW SOFTWARE
Figure 25 Measurement tab
2.6.7 and
VARIOS User Manual
section 2.7.5.2 for instructions
22