Philips PM32671 Service Manual page 70

100 mhz dual-channel oscilloscope
Table of Contents

Advertisement

INPUT VOLTAGE
Time-marker signal
repetition time:
0.1
0.2
0.5
5
P
0.1
ms
0.2 ms
0.5 ms
1 ms
Magnifier
Square-wave signal
repetition time Ips
Continuous control
As 4.9.2.
I
SETTINGS
-
Depress DTB of
S2
-
MTB
DTB
S13
S15
0.1
0.05
0.2
0.1
0.2 ps
0.5 ps
0.5
5
P
5
P
0.1
ms
0.1 ms
0.2
ms
1 ms
0.5 ms
2
1
ms
- Depress DTB of
S2
- TIME/DIV MTB
- TIME/DIV DTB
- Pull X MAGN
S7
X pos R 5 n
-
-
- TIME/DIV DTB
- Depress DTB of
S2
-
Continuous control
I
I
R EQU I R EM ENTS
Measured over 8 div. in horizontal
screen centre:
+/-
Coefficient error
3%
+/-
Coefficient error
3%
+/-
3%
Coefficient error
+/-
Coefficient error
3%
+/-
3%
Coefficient error
+/-
Coefficient error
3%
+/-
3%
Coefficient error
+/-
Coefficient error
3%
+/-
3%
Coefficient error
+/-
Coefficient error
3%
+/-
3%
Coefficient error
+/-
Coefficient error
3%
Coefficient error
3%
t/-
+/-
Coefficient error
3%
- MAGN led B2 lights up
+/-
- Coefficient error
5%
- Trace adjustable over
- UNCAL led B4 lights up
- Continuous range
1
:
MEASURING RESU LTS