Summary of Contents for BRONKHORST miniCORI-FLOW MI MKII Series
Page 1
Instruction Manual mini CORI-FLOW™ MI-series MKII Industrial Coriolis Mass Flow Meters/Controllers for Liquids and Gases Doc. no.: 9.17.206 rev. A Date: 19-03-2024 ATTENTION Please read this document carefully before installing and operating the product. Not following the guidelines could result in personal injury and/or damage to the equipment. Keep this document for future reference.
Page 2
At the same time a report should be submitted to your Bronkhorst representative. · If the product is damaged, it should not be put into service. In that case, contact your Bronkhorst representative for service. ·...
Page 3
Selected parts should be obtained only through Bronkhorst, to maintain accuracy and functionality of the product. If you are unsure about the suitability of a replacement component, contact your Bronkhorst representative for information.
Page 4
Bronkhorst® Instruction Manual mini CORI-FLOW™ MI-series MKII 9.17.206A...
(e.g. pressure, temperature) as specified at ordering time. If you are planning to use the product (including any third party components supplied by Bronkhorst, such as pumps or valves) with other media and/or other conditions, always check the wetted materials (including seals) for compatibility. See the technical specifications of the product and consult third party documentation (if applicable) to check the incorporated materials.
The analog output and the digital measured value are scaled accordingly. Switching between ranges can be done via the RS-232 interface or the fieldbus interface (if applicable), or with a Bronkhorst® readout and control unit E-8000.
The MI-series MKII comes with all necessary documentation for basic operation and maintenance. At some points this document refers to other documents, most of which can be downloaded from the Bronkhorst website. Insofar as agreed upon within the framework of the sales agreement, calibration certificates, test certificates and material certificates are included in the scope of delivery.
Page 10
Bronkhorst® The documentation listed in the following table can be downloaded from www.bronkhorst.com/downloads : Type Document Document no. General documentation EU Declaration of Conformity 9.06.059 Instruction manuals Manual CANopen 9.17.131 Manual DeviceNet™ 9.17.026 Manual EtherCAT® interface 9.17.063 Manual Ethernet/IP 9.17.132 Manual FLOW-BUS interface 9.17.024...
Before installing the MI-series MKII, check that the functional and technical properties of the product match your requirements. If you have a question about the product or if you find the product does not meet the specifications as ordered, do not hesitate to contact your Bronkhorst representative. See section Service for contact information.
Bronkhorst® Model key The model key contains information about the technical properties of the instrument as ordered. The actual properties of your instrument can be retrieved from the diagram below. Customized I/O options At ordering time, the firmware allows the configuration of two additional input/output functions, customized I/O options.
Page 13
Bronkhorst® Code Description 3.8…20.8mA sourcing output, controller Analog signal for Badger Meter valve with TEIP11 signal converter (control signal only) Digital output, min/max alarm During a min/max alarm, M8/A2 is pulled down to 0Vdc. Digital output, counter alarm During a counter alarm, M8/A2 is pulled down to 0Vdc.
Page 14
Bronkhorst® Code Description Digital input, reset counter The counter resets when M8/A2 is connected to 0Vdc. Digital input, reset alarm The alarm resets when M8/A2 is connected to 0Vdc. Instruction Manual mini CORI-FLOW™ MI-series MKII 9.17.206A...
Bronkhorst® Installation Location in fluid system To minimize the risk of gas entrapment due to cavitation, the position of the following system components is important: · Install a control valve downstream from the instrument. · Install a pump upstream from the instrument.
When in doubt about the suitability of your cabling, contact your Bronkhorst representative. Never power the instrument simultaneously from two different power sources (e.g. fieldbus and Plug-in Power Supply).
Bronkhorst® Opening top cover Closing top cover All wiring terminals are suitable for cables with a core diameter from 0.25 to 1.5 mm 3.4.2 Installing glands · To ensure proper shielding, connect the cable shielding with the glands before re-installing them to the instrument housing.
Bronkhorst® Operation After correct installation and taking all necessary safety precautions, the MI-series MKII can be used to measure and/or control flow. Powering up After cleaning at high temperature, allow the instrument to return to ambient temperature before turning it on.
Continuous purging of the housing interior with a dry, inert gas like Nitrogen can help prevent condensation. Contact your Bronkhorst representative for setting up an optimal purging configuration. 4.7.2 Cleaning temperature ·...
Bronkhorst® Communication interface The following table lists the communication interfaces the MI-series MKII can be equipped with (ex factory): Connector/terminal Type Communication standard Fieldbus/protocol Main Analog 0…5Vdc 0…10Vdc 0…20mA 4…20mA Digital RS-232 ProPar RS-485 FLOW-BUS Modbus (RTU/ASCII) HART HART Fieldbus...
Microsoft Office or custom made software, built with third party development software like LabVIEW or a SCADA platform. The FlowWare tools and associated documentation can be downloaded from the product pages on the Bronkhorst website: www.bronkhorst.com/products For more information about communication through the RS-232 interface, consult the RS-232 manual (document no.
Bronkhorst® Hardware interface · The instrument cover may only be opened in a clean, dry and non-hazardous environment. · Before opening the cover, make sure that the operating environment is perfectly safe and that the electrical components cannot be affected by moisture or aggressive atmospheric components.
Bronkhorst® The tables below list the different LED indications: Mode Pattern Time Indication continuous Power-off or program not running continuous Normal operation mode short flash 0.1 sec on, No bus communication, valves are in safe state 2 sec off blink 0.2 sec on,...
Bronkhorst® 4.9.2.1 Normal operating functions · In order to access these functions, press and hold the switch while the instrument is in normal operation mode (green LED lit continuously). · As long as the switch is held, the LEDs show a repeating sequence of patterns, where each pattern indicates a function.
Bronkhorst® 4.9.2.3 Control mode - readout/change Reading control mode · By briefly pressing the switch 2 times within 1 second in normal operation mode, the instrument shows its current control mode with a series of consecutive LED indication patterns. ·...
Bronkhorst® 4.9.2.4 Network settings - readout/change Reading network settings · By briefly pressing the switch 3 times with intervals of up to 1 second in normal operation mode, the instrument shows its current node address and baud rate with a series of consecutive LED indication patterns:...
Bronkhorst® Changing network settings · By briefly pressing the switch 5 times with intervals of up to 1 second in normal operation mode, the instrument enters a state in which the node address and baud rate can be changed (non-Ethernet based protocols only; for Ethernet based protocols, network parameters are configured by the fieldbus master and cannot be set on the instrument).
4.10 Adjusting zero point The zero point of a Bronkhorst® flow meter/controller (the measurement signal that indicates the absence of a flow) is factory adjusted at approximately 20 °C and atmospheric pressure (ambient conditions), with the instrument positioned upright. Under normal circumstances (i.e. at stable process conditions), the zero point will remain stable. However, over time several factors can induce a slight deviation of the measured value from the zero point, causing the instrument to detect a flow when in reality there is none.
4.10.1 Digital procedure Bronkhorst FlowSuite and FlowPlot provide an easy way to adjust the zero point of an instrument using RS-232 communication; the Autozero function automatically performs the procedure described here. To adjust the zero point using digital communication, set parameter values in the following sequence (see section...
Modbus address blocks are two bytes long. Larger data types use up to 8 subsequent address blocks, resulting in a maximum variable length of 16 bytes. Values longer than the maximum length are truncated. For more detailed information about setting up a Modbus network with Bronkhorst® instruments, consult the Modbus manual (see Documentation).
Bronkhorst® Measurement and control Measure Type Access Range FlowDDE FLOW-BUS Modbus Unsigned int 0…41942 0x0020/33 (65535*) This parameter returns a dimensionless representation of the measured flow rate or pressure. The value 32000 corresponds to 100 %, the maximum value corresponds to 131.07 %.
Liquid volume Liquid/gas mass Gas volume Alarms Alarm settings are most easily accessible using Bronkhorst FlowSuite, FlowPlot or FlowView or a Bronkhorst® readout and control unit. The built-in alarm functionality can be used to handle different alarm types: · system errors and warnings ·...
Page 33
Bronkhorst® Alarm Mode Type Access Range FlowDDE FLOW-BUS Modbus Unsigned char 0…3 97/3 0x0C23/3108 Available modes: Value Description Alarm off Alarm on absolute limits Alarm on limits related to setpoint (response alarm) Alarm at power-up(e.g. after power-down) (On DeviceNet™ instruments, only modes 0 and 1 are available)
Automatically (when alarm conditions no longer apply) Counter · Counter settings are most easily accessible using Bronkhorst FlowSuite, FlowPlot or FlowView or a Bronkhorst® readout and control unit. · When the instrument is powered down, it remembers the state of the counter. If the counter is active when the instrument is powered down, it is activated when powered up and then continues to count from the value at the time of power down.
Page 35
Bronkhorst® Counter Unit supports the following values: Mass Normal volume Standard volume Custom volume (1.01325 bar(a), 0 °C) (1.01325 bar(a), 20 °C) (Capacity Unit Pressure, Capacity Unit Type Temperature) ug, mg, g, kg uln, mln, ln, uls, mls, ls, ul, ml, l,...
Bronkhorst® The following methods are supported: Value Description By hardware switch (if present) Externally (obsolete) By parameter Reset Automatically (e.g. when counter value is reset) Network configuration Changes made to the network settings will not be restored by a factory reset.
Bronkhorst® Fieldbus1 Baud Rate Type Access Range FlowDDE FLOW-BUS Modbus Unsigned long 0…1.0E10 125/9 0xFD48…0xFD49/64841…64842 Fieldbus1 Parity Type Access Range FlowDDE FLOW-BUS Modbus Unsigned char 0…2 125/12 0x0FAC/4013 The following values are supported: Value Description No parity Odd parity Even parity...
Bronkhorst® This parameter represents the unit in which Capacity 100% is expressed. Available units: Mass flow Normal volume flow Standard volume flow Custom volume flow (1.01325 bar(a), 0 °C) (1.01325 bar(a), 20 °C) (Capacity Unit Type Pressure, Capacity Unit Type...
Controller The picture below is a simplified visualization of the PID controller algorithm (proportional, integral, derivative) used by digital Bronkhorst® instruments. The controller speed controls the overall performance of the controller algorithm. Basically, to adjust the controller response, only the controller speed needs to be changed.
Bronkhorst® PID-Kp Type Access Range FlowDDE FLOW-BUS Modbus Float 0…1E+10 114/21 0xF2A8…0xF2A9/62121…62122 PID controller proportional action, multiplication factor. PID-Ti Type Access Range FlowDDE FLOW-BUS Modbus Float 0…1E+10 114/22 0xF2B0…0xF2B1/62129…62130 PID controller integral action in seconds. PID-Td Type Access Range FlowDDE...
Unsigned char[20] 113/3 0xF118…0xF11F/ 61721…61728 Instrument serial number for identification. BHT Model Number Type Access Range FlowDDE FLOW-BUS Modbus Unsigned char[35] 113/2 0xF110…0xF117/ 61713…61720 This parameter shows the Bronkhorst® instrument model type information. 9.17.206A Instruction Manual mini CORI-FLOW™ MI-series MKII...
Identification Number Type Access Range FlowDDE FLOW-BUS Modbus Unsigned char 0…255 113/12 0x0E2C/3629 Bronkhorst® (digital) device type identification number. Device Type Type Access Range FlowDDE FLOW-BUS Modbus Unsigned char[6] 113/1 0xF108…0xF10A/ 61705…61707 Device type information string; this parameter contains an abbreviation referring to the identification number.
Page 43
Control Mode 18 prevents the instrument from assuming its default/safe state in the event of a digital communication failure. · The column labeled List option shows the control modes as used in Bronkhorst® software. Calibration Mode Type Access Range FlowDDE...
Bronkhorst® 5.9.1 Default control mode IO Status Type Access Range FlowDDE FLOW-BUS Modbus Unsigned char 0…255 114/11 0x0E4B/3660 The instrument is set to accept a setpoint from either an analog or a digital source. Although this setting can be changed...
4. Locate the defective fuse. 5. Replace the fuse with the same type (Littelfuse NANO2® Slo-Blo 2A; no 0454002). If you are unsure about the suitability of a replacement fuse or if replacement fuses keep blowing, contact your Bronkhorst representative for support.
(see Multifunctional switch) · with the restore function of a Bronkhorst® readout and control unit E-8000 Changes made to the network settings (bus address, baud rate, parity) will not be restored by a factory reset. 6.3.1 Manual procedure ·...
Page 47
Bronkhorst® Symptom Possible cause Action pressure to valve and slowly increase voltage. If valve does not open, clean parts and re-adjust valve Sensor failure Return equipment to factory · · Measurement disturbed by Control behavior unstable If possible, avoid installation in close ·...
Page 48
Bronkhorst® Symptom Possible cause Action components (e.g. adapters, tubing, valves) Continuous maximum measured Inlet pressure too high Check inlet pressure value or output signal · Valve fully open Close valve · Check if valve is in safe state (normally open valves);...
If you have a question about a product or if you find the product does not meet the specifications as ordered, do not hesitate to contact your Bronkhorst representative. To enable us to help you quickly and effectively, make sure to have the serial number (SN) ready whenever seeking contact with your Bronkhorst representative about a specific item.
All instruments must be dispatched with a completely filled in 'Declaration on decontamination'. Instruments without this declaration will not be accepted. A safety information document containing a 'Declaration on decontamination' form (document no 9.17.032) can be downloaded from the Service & Support section of the Bronkhorst website (www.bronkhorst.com). Instruction Manual mini CORI-FLOW™ MI-series MKII 9.17.206A...
Bronkhorst® Parameters - Master/Slave Parameter index Master Node Slave Factor Parameters Parameters - Measurement and control Parameters - Alarms Analog Input Alarm Delay Time Density Actual Alarm Info Fmeasure Alarm Maximum Limit Fsetpoint Alarm Minimum Limit Measure Alarm Mode Sensor type...
Need help?
Do you have a question about the miniCORI-FLOW MI MKII Series and is the answer not in the manual?
Questions and answers