Commissioning
April 2024
Note
Emerson provides 4-wire sensors for all single element RTDs. Use these RTDs in 3- or 2-wire
configurations by leaving the unneeded leads disconnected and insulated with electrical
tape.
Table 4-2:
Function
Device Information
Guided Setup
Manual Setup
Wireless Configuration
Sensor Calibration
Figure 4-5: RTD Lead Wire Configuration per IEC 60751
A. Single element, 3-wire
B. Single element, 4-wire
C. Dual element, 3-wire
D. Red
E. Black
F. White
G. Yellow
Note
To configure a single element, 4-wire RTD as a 3-wire system, connect only one white lead.
Insulate or terminate the unused white lead in a manner that prevents shorting to the
ground.
To configure a single element, 4-wire RTD as a 2-wire system, connect matching colored
wires first and then connect the paired wires to the terminal.
36
Wireless HART
Fast Key Sequences
®
Key sequence
1, 7
2, 1
2, 2
2, 2, 1
3, 5, 2
Reference Manual
00809-0100-4248
Menu items
Identification, Revisions, Radio, Security
Join Device to Network, Configure Update
Rate, Configure Sensor, Calibrate Sensor
Wireless, Process Sensor, Percent of Range,
Device Temperatures, Device Information,
Other
Network ID, Join to Network, Broadcast Info
Sensor Value, Sensor Status, Current Lower
Trim, Current Upper Trim, Lower Sensor
Trim, Upper Sensor Trim, Recall Factory
Trim