Emerson Rosemount 248 Reference Manual page 21

Wireless temperature transmitter
Hide thumbs Also See for Rosemount 248:
Table of Contents

Advertisement

Reference Manual
00809-0100-4248
% of Range
Supply Voltage
Related information
LCD display screen messages
Enabling and configuring LCD display with a Field
Communicator
From the Home screen, enter the Fast Key sequence: 2, 1, 5
Transmitters ordered with the LCD display will be shipped with the display enabled and
installed.
If the transmitter was ordered without the LCD display (and the LCD display was ordered
as a spare part) or if the LCD display was disabled, follow these steps to enable the LCD
display on the transmitter:
Procedure
1. From the Home screen, select 2: Configure.
2. Select 1: Guided Setup.
3. Select 5: Configure Device Display.
4. Select the option Periodic.
5. Select desired display options and click Enter.
Configuring LCD display with AMS Device Manager
Procedure
1. Right-click on the device and select Configure.
2. Select Configure Device Display button under Optional Setup.
3. Select desired display options and select Enter.