Reference Manual
00809-0100-4248
3.4.3 
LCD display
Transmitters ordered with the LCD display will be shipped with the display installed.
The optional LCD display can be rotated in 90-degree increments by squeezing the two
tabs, pulling out, rotating and snapping back into place.
Use the following procedure and
Procedure
1. Remove the power module cover and Green Power Module.
2. Remove the transmitter enclosure cover.
3. Tighten the captive screws and secure LCD adapter plate to 5 in-lb of torque.
4. Attach the LCD display on the LCD adapter plate, rotate to the desired position, and
5. Reattach and tighten the transmitter enclosure cover.
6. Connect the Green Power Module.
7. Reattach and tighten the power module cover.
Rosemount 248
Figure 3-6: Remote Mount
Note
Dimensions are in inches (millimeters).
Wireless devices must be powered up in order of proximity from the Wireless
Gateway, beginning with the closest device to the Gateway. This will result in a
simpler and faster network installation.
 CAUTION
When the circuit is live, do not remove the instrument covers in explosive
environments.
snap into place.
Figure 3-7
to install the LCD display: