Installation
April 2024
8. Always ensure a proper seal by installing the covers so that polymer contacts
32
polymer (i.e. no O-ring visible). Emerson recommends using Rosemount O-rings.
To enable and configure the LCD display, refer to
Note the following LCD display temperature limits:
Operating: –40 to 185 °F (–40 to 85 °C)
Storage: –40 to 185 °F (–40 to 85 °C)
Note
Only use Rosemount wireless LCD display part number: 00753-3203-0001.
An LCD display from a wired device will not function in a wireless device.
Figure 3-7: Optional LCD Display
Reference Manual
00809-0100-4248
Configuring the LCD
display.