Emerson Rosemount 248 Reference Manual page 30

Wireless temperature transmitter
Hide thumbs Also See for Rosemount 248:
Table of Contents

Advertisement

Installation
April 2024
5. Pull the wiring through the threaded conduit entry of the transmitter.
6. Attach the sensor wiring to the terminals as indicated on
7. Reattach and secure LCD display adapter plate to 5 in.-lb. of torque (if applicable).
8. Reattach the LCD display (if applicable).
9. Reattach and tighten the transmitter enclosure cover.
10. Remove the power module cover.
11. Connect the Green Power Module.
12. Reattach and tighten the power module cover.
13. Always ensure a proper seal by installing the electronics housing cover(s) so that
14.  Provide 1.75 in. (45 mm) of clearance for units without an LCD display. Provide 3 in.
30
Figure 3-5: Exploded View of LCD Display Assembly
A. Housing cover
B. LCD meter
C. Terminal screws
D. LCD adapter plate
E. Housing
F. Green power module
G. Power module cover
polymer contacts  polymer (i.e. no O-ring visible). Use Rosemount O-rings.
(76 mm) of  clearance for units with an LCD display for cover removal.
Reference Manual
00809-0100-4248
Figure
2-1.
Emerson.com/Rosemount