BRUEL & KJAER 2116 Instruction Manual page 32

Audio test station
Table of Contents

Advertisement

26
The procedure for setting up and calibrating the instrumentation is as follows. For de-
tailed information about setting up the Artificial Mastoid please refer to its Instruction
Manual
.
1
.
Connect the output of the Force Transducer of the Artificial Mastoid to the preampli-
fier Type 2642 using the cable supplied with the Mastoid
.
An adaptor
JJ
2615 is re-
quired to connect the mini plug of the transducer cable to the half-inch preamplifier
.
2.
Connect the preamplifier to the INPUT socket on the rear panel of the Audio Test Sta-
tion.
3
.
Connect the bone conductor to the OUTPUT socket on the rear panel of the 2116
.
The pin connection diagram is shown in Fig.3.2.
4
.
Place the bone conductor on the calibration surface of the Artificial Mastoid and ad-
just the preload to the required value as described in detail in the manual for the Ar-
tificial Mastoid
.
5.
Ensure that the COMPRESSION switch on the rear panel of the 2116 is in the
"Out" position so that a constant voltage output is obtained.
6.
Select a suitable output level using the SPL UP and DOWN pushkeys
.
7
.
Select the required measurement parameter and press RUN to start the measure
-
ment.