5    Troubleshooting
5.1    Troubleshooting electrical system
Symptom
Mains indicator does not light up
Mains indicator lights up, but the motor is not 
running
Control box is in order except for one direction 
on one channel
Motor is running, but the piston rod does not 
move
The motor cannot lift full load
Motor noise, but no movement of piston rod
Piston rod operates inwards and not outwards
         
* Service and maintenance of the bed must only be performed by qualified technicians who have received the necessary 
instruction or training. 
WARNING!
Risk of personal injury and damage to the product                  
 — The bed must be unplugged from the main power source before opening or repairing electrical parts.
1577126-J
Possible cause
Mains are not connected
Fuse in the control box is blown
Control box is defective
Motor plug is not fully inserted into the 
control box.
The motor is defective
Motor cable is damaged
Control box is defective
Hand control is defective
Control box is defective
Hand control is defective
Motor is damaged *
Remedy
Connect mains
* Replace the control box
* Replace the control box
Insert the motor plug properly into the 
control box
* Replace the motor
* Replace the cable
* Replace the control box
* Replace the hand control
* Replace the control box
* Replace the hand control
Replace the motor