Invacare® SB®755 en Bed User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Page 2
All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
If a problem should arise in connection with the delivered product, manual. please contact your Invacare® dealer. An address list is shown on the back side at this manual. 1.3 Service life 1.1.1 Symbols in this manual...
2.3 Labels and symbols on the product the bed. – Make sure that no wires (mains or from other 2.3.1 Product label equipment) are jammed or damaged, when the bed is Invacare Rea AB EN 60601-2-52 used. Växjövägen 303 YYYY - MM 34375 Diö...
Invacare® SB®755 The product label is placed on the frame of the bed and contains the WEEE conform main product information, including technical data. Symbols This product complies with Directive 93/42/EEC concerning medical devices. Serial Number Abbreviations for technical data: Reference Number •...
Operating the bed C Leg section motor K Base Frame 3 Operating the bed D Thigh section L Brake - castor with brake E Thigh section motor M Control Unit 3.1 General Safety Information F Leg section N Mattress support, upper half WARNING! G Brake pedal - central O Mattress retainer...
Invacare® SB®755 3.4.2 Britt V and Line side rails 3.4.3 Verso II side rails Lock / Upper position Pull up the top wooden side rail bar, until the locking pin locks with an audible click Release Lift the top wooden side rail bar...
Castors might make marks on different types of absorbing floor surfaces - such as untreated or badly treated floors. To prevent marks, Invacare® recommends to place a suitable Steerable castor (optional) kind of protection between the castors and the floor.
Invacare® SB®755 3.8 Lifting pole 3.8.2 Changing the handle height WARNING! Risk of injury The bed can tip if the handle is used, while the lifting pole is turned away from the bed. – The lifting pole always has to be positioned with the handle hanging over the bed area.
Page 11
Operating the bed Both persons must hold the mattress section. One of them pulls out the cotter pin (from the motor in question). Both persons slowly lower the mattress section until it is completely down. 1567446-E...
Invacare® SB®755 Electrical components 4 Maintenance Method: Wipe off with a wet cloth or soft brush. Max temp: 40 °C 4.1 Cleaning and disinfection Solvent/ Water chemicals: IMPORTANT! Wrong fluids or methods can harm or damage you Metal components product.
After Use 5 After Use IMPORTANT! Accumulator back-up – Old accumulators must be returned to Invacare® or recycled as car batteries 5.1 Waste disposal Waste disposal/recycle must comply with the laws and 5.2 Reuse regulations for waste handling in each country.
Invacare® SB®755 6 Troubleshooting 6.1 Troubleshooting Symptom Possible cause Remedy Bed section does not move End of stroke reached Operate opposite button Bed not plugged in Plug in mains cable Ensure hand control correctly connected to Hand control not connected control unit.
With smaller or bigger castors, the heights differ accordingly. Scala Medium 3/4 length 165 cm x 46 Mounted on collapsible the side tube. Invacare® reserves the right to change the stated measurements and steel Release angles without warning. side rail button in the head section.
Invacare® SB®755 7.4 Environmental conditions Type B Applied Part complying with the specified Applied requirements for protection against electrical shock Storage and Operation Part according to IEC60601-1. transportation Temperature -10°C to +50°C +5°C to +40°C Sound level 55 dB (A)
Page 17
Technical data Immunity test IEC 60601–1–2 test level Compliance level Electromagnetic environment – guidance < 5% U (>95% dip in U < 5% U (>95% dip in U Mains power quality should be that of a typical for 0,5 cycle for 0,5 cycle Voltage dips, short commercial or hospital environment.
Page 18
Invacare® SB®755 0.01 0.12 0.12 0.23 0.37 0.37 0.74 1.17 1.17 2.33 3.69 3.69 7.38 11.67 11.67 23.33 For transmitters rated at a maximum output power not listed above the recommended separation, distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.