6.5 4-Pole Earth/Ground Measurements
Pressing the Hz/OPTIONS button allows the following parameters to be changed
using the ► button:
• 128Hz blinks
• Test voltage blinks Switch between 16 and 32V
6.6 Soil Resistivity Measurements
Pressing the Hz/OPTIONS button allows the following parameters to be changed
using the ► button:
• ρ
• 128Hz blinks
• Test voltage blinks
6.7 Measurements with Two Clamps
Pressing the Hz/OPTIONS button allows the following parameters to be changed
using the ► button:
•
1611Hz blinks
6.8 Smoothing
In the manual mode, you can activate or deactivate the smoothing of the measure-
ment results by pressing the 2nd + DISPLAY (SMOOTH) buttons. This smoothing
consists in displaying an exponential mean value, a significant help with highly
fluctuating values.
48
Find Quality Products Online at:
Change the test frequency
blinks (Wenner method) Switch to ρ
W
Change the test frequency
GlobalTestSupply
www.
(Schlumberger method)
S
Change the test frequency
Switch between 16 and 32V
Ground Resistance Tester Model 6471
.com
sales@GlobalTestSupply.com