6 Maintenance and Service

6.2.1.

Basic Cleaning

If the response times are longer than specified, this may be due to contamination. In this case clean the
ventilation slots by blowing from inside out.
6.2.2.

Error Messages in PCS10

The PCS10 information section includes error messages from HTS201, if there are any errors present. The
following table shows the error codes and their meaning. Please note that this table directly corresponds to
Tab. 7 Device status indication register in chapter 5.2.7.
Error code
Description
7.x
Temperature measurement incorrect
9.x
Humidity measurement incorrect
19.x
Sensing element not accessible
21.x
Display communication faulty
Tab. 10
Possible error messages for HTS201
6.2.3.

Repairs

PLEASE NOTE
Repairs may be carried out by the manufacturer only. The attempt of unauthorized repair excludes any
warranty claims.

7 Accessories

Description
E+E Product Configuration Software
(Free download from www.epluse.com/pcs10)
USB Configuration Adapter for HTS201 digital
Recommended action
1. Check the actual temperature with the
temperature output scaling of the ordered
sensor, check the wiring
2. Service recommended, check the repair
possibilities with the E+E service
1. Check the actual temperature with the
temperature output scaling of the ordered
sensor, check the wiring
2. Service recommended, check the repair
possibilities with the E+E service
Service recommended, check the repair
possibilities with the E+E service
Service recommended, check the repair
possibilities with the E+E service
Code
PCS10
HA011066
User Manual HTS201
Humidity and Temperature Room Sensor | 18