Adjustment Of The Ee23 Model T5 Via The Usb Service Interface; Maintenance; Troubleshooting; Fuse Replacement For Option Am3 - E+E Elektronik EE23 Operation Manual

Humidity/temperature sensor
Hide thumbs Also See for EE23:
Table of Contents

Advertisement

5.3

Adjustment of the EE23 Model T5 via the USB Service Interface

1. Download and install the EE-PCS Product Configuration Software from www.epluse.com/configurator
2. Connect the USB service interface of EE23-T5 to the PC.
3. Start the EE-PCS software.
4. Select the desired adjustment mode and follow the instructions of the EE-PCS.
6

Maintenance

When employed in dusty, polluted environment:
• The filter cap shall be replaced as needed with an E+E original one. A polluted filter cap causes
longer response time of the device.
• For cleaning of the sensing head please see "Cleaning Instructions" at www.epluse.com/ee23.
7

Troubleshooting

See status LEDs at „3 Operating Components" on Page 6 and error messages on the display at „3.4
Display for Model T5" on Page 9.
7.1

Fuse Replacement for Option AM3

If the green LED on the PCB is not flashing with the supply voltage switched on check the fuse and
replace if required.
Fuse secondary:
Recommended replacement types:
Series: MSTU 250
Series: 374
Fig. 11
Fuse replacement
16
250mA / T UL248-14
Manufacturer: Schurter
Manufacturer: Littelfuse
Order No.: 0034.7109
Order No.: 374 0250