Specification—2246 Operators
Table 1-3
Environmental Characteristics
Characteristics
|
Description
STANDARD
INSTRUMENT
DE
NEE
Environmental Requirements
| Instrument meets or exceeds the environmental requirements of
MIL-T-28800D for Type Ill, Class 3, Style D equipment.
Temperature
Operating
— 10°C to +55°C (+14°F to +131°F).
-51°C
to +71°C
(—60 to 160°F). Tested
to MIL-T-28800D
paragraphs 4.5.5.1.3 and 4.5.5.1.4, except in 4.5.5.1.3, steps 4
and 5 (-10°C operating test) are performed ahead of step 2
(-51°C
non-operating
test). Equipment
shall remain off upon
return to room ambient during step 6. Excessive condensation -
shall be removed before operating during step 7.
Non-operating
Altitude
Operating
To
4,570
m
(15,000
ft),
Maximum
operating
temperature
decreases 1°C/1000 ft above 5000 ft.
Non-operating
To 15,240 m (50,000 ft).
Humidity (Operating and Non-operating)
Five cycles (120 hours) referenced to MIL-T-28800D paragraph
4.5.5.1.2.2, for type lil, class 3 instruments.
Non-operating and operating at 95%, 0% to +2% relative humid-
ity. Operating at +30°C and +55°C for all modes of operation.
Non-operating at +30°C to +60°C.
Radiated
and Conducted
Emission
Required
per
Meets Category B.
VDE 0871
15 minutes along each of 3 major axis at a total displacement of
0.025 inch p-p (4 g at 55 Hz) with frequency varied from 10 Hz to
55 Hz in 1-minute sweeps. Hold for 10 minutes at 55 Hz in each
of the three major axis. All major resonances
must be above
55 Hz.