Table 2: Measurement characteristics (cont.)
Characteristics
Resolution (by range)
300 W
3 kW
30 kW
300 kW
3 MW
30 MW
Continuity check threshold
Diode test
Maximum test current
Test voltage (open circuit)
Resolution
1
AC conversions are AC-coupled, true RMS responding, and calibrated to the RMS value of a sine wave input. The basic
accuracy is for sine wave at full scale and non-sine wave below half scale (3 V range only for sine wave measurement).
For non-sine wave accuracy, refer to the following crest factor guide:
1.4 to 2.0, add 0.5% to accuracy
2.0 to 2.5, add 2% to accuracy
2.5 to 3.0, add 4% to accuracy
Table 3: Certifications and compliances
EC Declaration of Conformity
Certifications
Overvoltage Category
Pollution Degree 2
Handheld Instruments Basic Service
Description
0.1 W
1 W
10 W
100 W
1 kW
10 kW
Approximately 50 W — tone will sound
1.5 mA
3.3 V maximum
1 mV
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility and Low
Voltage Directive 73/23/EEC for Product Safety. Compliance was demonstrated to the
following specifications as listed in the Official Journal of the European Communities:
EMC Directive 89/336/EEC:
EN 55011
Class B Radiated and Conducted Emissions
EN 50082-1 Immunity:
IEC 801-2
Electrostatic Discharge Immunity
IEC 801-3
RF Electromagnetic Field Immunity
Low Voltage Directive 73/23/EEC as amended by 93/68/EEC:
EN 61010-1/A2
Safety requirements for electrical equipment for
measurement, control, and laboratory use
Listed UL3111-1 and CAN/CSA C22.2 No. 1010.1.
Category:
Examples of Products in this Category:
CAT III
Distribution-level mains, fixed installation
CAT II
Local-level mains, appliances, portable equipment
CAT I
Signal levels in special equipment or parts of equipment,
telecommunications, electronics
Do not operate in environments where conductive pollutants may be present.