Alternatively, you can remove the charging wand from
your implant site, which will cause your Vesta Charger to
time out and shut off automatically.
Note: If you wish to resume charging your OPTIMIZER
Lite IPG after you terminate a charging session, please
wait for approximately 10 minutes before initiating a new
charging session to allow the temperature of your
OPTIMIZER Lite IPG to return to its baseline temperature.
7.8.2
Termination of Charging Session Due to IPG
Temperature
To ensure your safety while charging your OPTIMIZER
Lite IPG, the temperature of your IPG is monitored during
the charging process. If the reported temperature of your
OPTIMIZER Lite IPG at the beginning of the charging
session is outside the acceptable temperature range or if
the temperature of your implanted OPTIMIZER Lite IPG
remaining consistently high for more than 10 minutes
while it is being charged, then your Vesta Charger will
emit 3 long beeping tones and display the Charging IPG
Temperature Error screen, indicated by a thermometer
icon in the center of the screen (see Figure 19). Your
Charger will then shut off. If this should occur, please wait
for approximately 20 minutes before initiating a new
charging session.
Figure 19: Charging IPG Temperature Error Screen
23